ADVIA Chemistry XPT System Software May Report Incorrect Patient Test Results
Siemens Healthcare is recalling certain ADVIA Chemistry XPT systems with software version 1.4 or lower due to a default mixer failure detection setting that may allow incorrect patient lab results to be reported without alerting clinicians.
- Product
- The ADVIA Chemistry XPT Chemistry System, Siemens Material Number (SMN) 10723034, is an automated, clinical chemistry analyzer that runs tests on serum, plasma, urine, or cerebral spinal fluid in random access and batch modes. Tests performed using this system are intended for…
- Category
- Medical Device
- Distribution
- Distributed nationwide