  • HighFDA (Devices)·Z-0612-2022·2022-02-16

    GE Healthcare MRI superconducting magnets recalled for cryogen ventilation defect

    GE Healthcare is recalling 17,228 MRI superconducting magnets that may have improper cryogen ventilation systems. The defect could fail to properly vent cryogenic materials in the SIGNA 3.0T MR System.

    Product
    GE Healthcare MR superconducting magnets, a component of GE Healthcare SIGNA 3.0T MR System, nuclear magnetic resonance imaging system.
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-0589-2022·2022-02-16

    QC Blood Culture Identification Panel Recall for Incorrect Antibiotic Susceptibility Test Results

    Microbiologics recalls KWIK-STIK blood culture identification quality control panels due to incorrect susceptibility test results. High-level Gentamicin Synergy values were out of specification, risking inaccurate diagnostic guidance.

    Product
    QC Sets and Panels, KWIK-STIK Blood Culture Identification (BCID) Verification Panel (Live Culture). KWIK-STIK kit that contains Catalog Number 0959P. In Vitro Diagnostic Control microorganisms
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-0585-2022·2022-02-16

    Quality Control Kit for Antibiotic Susceptibility Testing Returning Incorrect Results

    Microbiologics Inc is recalling KWIK-STIK quality control kits used by laboratories because certain lots return incorrect test results that could affect laboratory testing accuracy.

    Product
    KWIK-STIK, Quality control kit for culture media, sold as: a. KWIK-STIK 2-pack Enterococcus faecalis derived from ATCC 51299 b. KWIK-STIK 6-pack Enterococcus faecalis derived from ATCC 51299
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighNHTSA·22V087000·2022-02-16

    Ford F-250 and F-350 Driveshaft Fracture Risk From Loose Insulators

    Ford is recalling 2017–2022 F-250 and F-350 Super Duty trucks. Underbody insulators may loosen and damage the aluminum driveshaft, potentially causing loss of drive power or vehicle control.

    Product
    FORD — FORD F-250 SD, F-350 SD
    Category
    Vehicle
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-0615-2022·2022-02-16

    GE Healthcare MRI Superconducting Magnet Cryogen Ventilation System Defect

    GE Healthcare MR superconducting magnets in SIGNA MRI systems may have cryogen ventilation systems that do not meet safety venting requirements. Approximately 17,228 systems are affected.

    Product
    GE Healthcare MR superconducting magnets, a component of GE SIGNA Excite 1.5T MR System, GE SIGNA Excite 3.0T MR System, nuclear magnetic resonance imaging system.
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-0631-2022·2022-02-16

    Knee Replacement Component Recalled for Adhesive Residue Risk

    DePuy Orthopaedics is recalling ATTUNE Revision knee implant components due to possible adhesive residue on the surface. If undetected before surgery, the residue may cause soft tissue irritation or infection.

    Product
    ATTUNE REVISION POSTERIOR FEMORAL AUGMENT SZ 9 12MM-intended as a Component in total knee arthroplasty (TKA) revision surgeries Product Number: 154909003
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-0563-2022·2022-02-16

    Arterial Line Kits recalled due to guidewire-needle incompatibility

    Argon Medical Devices is recalling Arterial Line Kits because the included guidewires don't fit through the kit's needles. This could cause procedure delays and minor blood loss.

    Product
    1. Arterial Line Kit 20ga x 6" 2. Arterial Line Kit /1 20ga x 6"
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-0597-2022·2022-02-16

    GE Healthcare MRI Magnets Recalled for Inadequate Cryogen Ventilation

    GE Healthcare is recalling MRI magnets used in SIGNA MR355 and MR360 systems that could have cryogen ventilation systems not meeting venting requirements. These systems were distributed nationwide and internationally.

    Product
    GE Healthcare MR superconducting magnets, a component of GE Healthcare SIGNA MR355 and SIGNA MR360 System, nuclear magnetic resonance imaging system.
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighCPSC·22082·2022-02-16

    Viking 5 Series Freestanding Gas Ranges Recalled for Gas Leak and Fire Risk

    Viking Range is recalling about 3,050 Viking 5 Series freestanding gas ranges because the rigid gas tubing can separate at burner joints, creating a risk of gas leak and fire. Consumers should stop using the bake, broil, and griddle functions immediately and contact Viking for a free repair.

    Product
    Viking 5 Series Freestanding Gas Ranges
    Category
    Consumer Product
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-0588-2022·2022-02-16

    Diagnostic Quality Control Microorganism Set Fails Gentamicin Susceptibility Specifications

    Microbiologics Inc is recalling 139 units of KWIK-STIK AST-GP QC microorganism sets due to incorrect susceptibility test results. The products failed gentamicin synergy specifications and could affect the accuracy of laboratory diagnostic testing.

    Product
    QC Sets and Panels, KWIK-STIK AST-GP (6 Strains) QC Set. KWIK-STIK kit that contains Catalog Number 0959P. In Vitro Diagnostic Control microorganisms
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-0583-2022·2022-02-16

    LYFO DISK Streptococcus Pneumoniae Culture Kit Recalled for Bacterial Contamination

    Microbiologics Inc is recalling LYFO DISK Streptococcus pneumoniae microorganism culture kits (lot 947-126-1) distributed worldwide due to potential contamination with E. coli, Staphylococcus epidermidis, and S. warneri.

    Product
    LYFO DISK(TM), LYFO DISK(TM) Streptococcus pneumoniae derived from ATCC(R) 49619(TM). LYFO DISK microorganisms are packaged in a resealable vial that contains 6 lyophilized microorganism pellets and a desiccant to prevent adverse moisture accumulation. The LYFO DISK…
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-0611-2022·2022-02-16

    GE Healthcare MRI Systems Recalled for Inadequate Cryogenic Ventilation

    GE Healthcare has recalled 17,228 MRI systems because the cryogenic ventilation system does not meet safety requirements. Systems were distributed nationwide and internationally.

    Product
    GE Healthcare MR superconducting magnets, a component of GE Healthcare Brivo MR355 System, nuclear magnetic resonance imaging system.
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-0598-2022·2022-02-16

    GE Healthcare MRI Systems with Potentially Non-Compliant Cryogen Ventilation Systems

    GE Healthcare has recalled approximately 17,228 MRI systems due to cryogen ventilation systems that may not meet required venting standards. The issue affects systems distributed nationwide and internationally.

    Product
    GE Healthcare MR superconducting magnets, a component of GE Healthcare SIGNA MR380 System, nuclear magnetic resonance system.
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-0617-2022·2022-02-16

    GE Healthcare MRI systems recalled for cryogenic ventilation defect

    GE Healthcare MRI systems may have cryogenic ventilation systems that do not meet venting requirements. The recall affects 17,228 devices distributed nationwide and globally. Healthcare facilities should contact GE Healthcare for corrective actions.

    Product
    GE Healthcare MR superconducting magnets, a component of GE Healthcare SIGNA Horizon Cx System (sold under SIGNA Horizon Cx, SIGNA Horizon Lx, SIGNA Infinity, SIGNA MR/I, SmartSpeed, EchoSpeed Plus, and HighSpeed Plus), nuclear magnetic resonance imaging system.
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-0606-2022·2022-02-16

    GE Healthcare MRI Systems Recalled for Inadequate Cryogen Ventilation

    GE Healthcare SIGNA HFO/I MR Systems may have cryogen ventilation systems that do not meet venting requirements. Approximately 17,228 systems are affected nationwide and internationally.

    Product
    GE Healthcare MR superconducting magnets, a component of GE Healthcare SIGNA HFO/I MR System, nuclear magnetic resonance imaging system.
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-0630-2022·2022-02-16

    BD Veritor Plus Analyzer may overheat when using unauthorized power adapter

    BD Veritor Plus Analyzer devices may overheat or catch fire if powered with an unauthorized AC adapter instead of the manufacturer-provided one. The recall affects approximately 129,848 analyzers distributed worldwide.

    Product
    BD Veritor Plus Analyzer - intended to provide rapid test results in near-patient settings. Catalog Number: 256066
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-0608-2022·2022-02-16

    GE Healthcare MRI systems recalled over improper cryogen ventilation

    GE Healthcare is recalling MRI systems with potentially non-compliant cryogen ventilation. Over 17,228 superconducting magnetic resonance imaging devices may have cryogen venting systems that fail to meet FDA requirements.

    Product
    GE Healthcare MR superconducting magnets, a component of GE 0.7T SIGNA OpenSpeed with Excite MR System, nuclear magnetic resonance imaging system.
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-0569-2022·2022-02-16

    R82 by ETAC Push Brace for High-Low recalled due to cracking hazard

    The R82 by ETAC Push Brace for High-Low is being recalled because the chrome surface can crack during use, potentially cutting users. Approximately 396 units manufactured between August and October 2021 are affected.

    Product
    R82 by ETAC Push Brace for High-Low
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighNHTSA·22E009000·2022-02-16

    Fuel Helmets Recalls SH-WS001 Motorcycle Helmets for Impact Protection Defect

    Fuel Helmets is recalling specific SH-WS001 motorcycle helmets that may not adequately protect the wearer in a crash. Owners will receive a credit to replace or reimburse the helmet.

    Product
    EQUIPMENT:MOTORCYCLE:HELMETS recall
    Category
    Consumer Product
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-0610-2022·2022-02-16

    GE Healthcare MRI Superconducting Magnets: Cryogen Ventilation System Defect

    GE Healthcare is recalling 17,228 MRI systems because the cryogen ventilation system may not meet venting requirements. The company is providing field corrections to ensure proper ventilation.

    Product
    GE Healthcare MR superconducting magnets, a component of GE Healthcare Brivo MR355/Optima MR360 System, nuclear magnetic resonance imaging system.
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-0591-2022·2022-02-16

    GE Healthcare MRI magnets recalled for cryogen ventilation defect

    GE Healthcare is recalling 17,228 MRI system magnets with potential cryogen ventilation defects. No injuries have been reported.

    Product
    GE Healthcare MR superconducting magnets, a component of GE Healthcare SIGNA Architect system, nuclear magnetic resonance imaging system.
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighNHTSA·22V088000·2022-02-16

    2022 Lincoln Aviator Fuel Filler Tube May Detach During Crash

    Ford is recalling certain 2022 Lincoln Aviator vehicles because the fuel filler tube may detach during a crash, creating a fuel leak and fire risk. Free repairs are available from Ford dealers.

    Product
    FORD — FORD, LINCOLN EXPLORER, AVIATOR
    Category
    Vehicle
    Distribution
    Distributed nationwide
  • HighCPSC·22080·2022-02-16

    Esme Children's Sleepwear Garments Fail Flammability Standards

    Esme is recalling about 3,600 children's sleepwear garments because they fail federal flammability standards and pose a burn risk. Consumers should stop using them immediately and contact Esme for a free replacement or refund.

    Product
    Children's Sleepwear Garments
    Category
    Consumer Product
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-0599-2022·2022-02-16

    GE Healthcare MRI superconducting magnets recalled for cryogen ventilation defects

    GE Healthcare is recalling MR superconducting magnets used in SIGNA Voyager MRI systems due to cryogen ventilation systems that may not meet venting requirements. The recall affects 17,228 devices.

    Product
    GE Healthcare MR superconducting magnets, a component of GE Healthcare SIGNA Voyager and SIGNA Voyager Premier Edition system, nuclear magnetic resonance imaging system.
    Category
    Medical Device
    Distribution
    Distributed nationwide