Siemens Atellica CH β2-Microglobulin Assay Lot 211446 Recalled for Measurement Bias
Siemens Healthcare Diagnostics recalled lot 211446 of the Atellica CH β2-Microglobulin assay due to analytical bias in test results. The assay showed positive bias up to 20.9% at some concentrations and negative bias up to 10.5% at others.
- Product
- Atellica CH ¿2-Microglobulin (B2M) -M) Assay for in vitro diagnostic use in the quantitative determination of ¿2-microglobulin in human serum and plasma Siemens Material Number (SMN): 11097635
- Category
- Medical Device
- Distribution
- Distributed nationwide