Severity 3 (High) recalls have a known hazard but no confirmed illness, injury, or property damage in the source text. The risk is theoretical-but-real: a pathogen pre-emptively pulled from shelves before anyone got sick, an allergen mislabeling on an uncommon allergen, a precautionary correction on a device that hasn't yet malfunctioned in the field. Most allergen recalls land here, as do most preventive food recalls based on routine sampling. You should still stop using the product, but you don't need to panic.

The full rubric is on the methodology page. Severity is computed deterministically from the agency’s own classification plus the source text — no editorial vibes.

14926–14950 of 53092

  • HighFDA (Devices)·Z-0722-2023·2022-12-28

    GE Centricity Universal Viewer 6.0 Medical Imaging Software May Hide Studies

    GE Medical Systems' Centricity Universal Viewer 6.0 image display software may hide medical studies from users after IAS tool installation, requiring GE assistance to access hidden images and causing delays in reading diagnostic studies.

    Product
    Centricity Universal Viewer 6.0. Used to display medical images (Including mammograms) and data from various imaging sources.
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-0757-2023·2022-12-28

    GE Vivid q Ultrasound Systems: Battery Fire Risk Recall

    GE is recalling Vivid q ultrasound systems because batteries not replaced every 2 years as recommended may fail and emit smoke or catch fire.

    Product
    Vivid q ultrasound. Used in ultrasound imaging and analysis in Fetal/Obstetrics; Abdominal/GYN; Pediatric; Small Organ (breast, testes, thyroid); Neonatal Cephalic; Adult Cephalic; Cardiac (adult and pediatric); Peripheral Vascular; Musculo-skeletal Conventional and…
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-0766-2023·2022-12-28

    Lamaze Chill Teether Recalled for Potential Microbial Growth in Fluid

    Lamaze Chill Teether teething rings are being recalled due to potential microbial growth in the internal fluid. This poses an infection risk, particularly for children with weakened immune systems.

    Product
    Lamaze Chill Teether, Fluid-Filled Teething Ring, Model No. Y5288L
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-0765-2023·2022-12-28

    Medical Action Industries Littauer Scissors Recalled for Debris Contamination

    Medical Action Industries is recalling Littauer surgical scissors because the sterile instruments may contain debris that could compromise their safety during clinical use.

    Product
    Medical Action Industries Inc. Disposable Instrument Pack, Littauer Scissor, REF 56247, packaged 1/package, 20 packages/box, 12 boxes/case, sterile.
    Category
    Medical Device
    Distribution
    8 states
  • HighFDA (Food)·F-0206-2023·2022-12-28

    Red Button French Silk Pies Recalled for Undeclared Almonds

    Rocky Mountain Pies is recalling Red Button Vintage Creamery French Silk Pies because some contain Turtle Cream Pies with almonds not declared on the label. Consumers with almond allergies are at risk.

    Product
    Red Button Vintage Creamery French Silk Pie, Net Wt. 31.88 oz. (904g), in a clear plastic dome and black plastic base, UPC 0 41172 81290 9, Distributed By: Associated Food Stores, Salt Lake City, Utah 84119 RMYA0421" printed on sleeve. Keep refrigerated.
    Category
    Food
    Distribution
    9 states
  • HighFDA (Devices)·Z-0723-2023·2022-12-28

    ConvaTec AQUACEL Foam Ag Adhesive wound dressing recalled for visual contamination

    ConvaTec is voluntarily recalling AQUACEL Foam Ag Adhesive wound dressings due to visual contamination (brown spots) found on several dressings. Approximately 5,300 units distributed nationwide are affected.

    Product
    ConvaTec AQUACEL Foam Ag Adhesive, wound dressing advesive
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-0734-2023·2022-12-28

    Fresenius 2008T Hemodialysis Machines Recalled for False USB Device Alarm

    Fresenius Medical Care is recalling 2008T and 2008T Bluestar Hemodialysis Machines with CDX due to a software and hardware sensitivity to electromagnetic interference that triggers false alarms.

    Product
    2008T and 2008TBluestar Hemodialysis Machines with CDX-Indicated for acute and chronic dialysis therapy in a healthcare facility Models: (1) 190713 2008T HEMODIALYSIS SYS., with CDX; (2) 190766 2008T HEMODIALYSIS SYSTEM W/BIBAG W/CDX ; (3) 191124…
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-0759-2023·2022-12-28

    Ultrasound System Batteries May Fail, Emit Smoke, or Catch Fire

    GE Medical Systems is recalling Vivid q N ultrasound systems due to potential battery failures that could emit smoke or catch fire if batteries are not replaced per the 2-year service manual recommendation.

    Product
    Vivid q N ultrasound. Not marketed in the US. Used in ultrasound imaging and analysis in Fetal/Obstetrics; Abdominal/GYN; Pediatric; Small Organ (breast, testes, thyroid); Neonatal Cephalic; Adult Cephalic; Cardiac (adult and pediatric); Peripheral Vascular…
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-0756-2023·2022-12-28

    GE Vivid i Ultrasound Systems Battery Failure Recall

    GE Medical Systems is recalling Vivid i ultrasound systems because batteries not replaced every two years may fail, emit smoke, or catch fire. About 7,931 units were distributed worldwide.

    Product
    Vivid i ultrasound. Marketed under the following 510(k) numbers: K033139, K061525, K082374, K092140, K102388, K121062). Used in ultrasound imaging and analysis in Fetal/Obstetrics; Abdominal/GYN; Pediatric; Small Organ (breast, testes, thyroid); Neonatal Cephalic; Adult…
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-0749-2023·2022-12-28

    HeartWare HVAD ventricular assist pump driveline cover hardening recall

    HeartWare HVAD ventricular assist pump kits may develop hardened driveline covers that impair access to critical connectors. If urgent servicing is needed, delayed access could result in patient harm from prolonged pump stoppage.

    Product
    HeartWare HVAD Pump Kit, REF 1205
    Category
    Medical Device
    Distribution
    0 states
  • HighFDA (Devices)·Z-0764-2023·2022-12-28

    Baxter Revaclear 400 Dialyzers Recalled for Polyurethane Displacement

    Baxter Healthcare is recalling Revaclear 400 dialyzers because polyurethane displacement may cause blood leaks during dialysis treatment. Approximately 1.14 million units are affected worldwide.

    Product
    Revaclear 400 dialyzers. Indicated for treatment of chronic and acute renal failure by hemodialysis.
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-0761-2023·2022-12-28

    Revaclear 400 Dialyzer recall due to potential internal blood leaks

    Baxter Healthcare recalls Revaclear 400 Dialyzer units due to potential internal blood leaks caused by twisted gaskets. The defect may affect dialysis patients worldwide.

    Product
    Revaclear 400 Dialyzer. Product Code: 114749M. Not distributed in the USA. Used to treat chronic and acute renal failure by hemodialysis.
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-0728-2023·2022-12-28

    VADER Pedicle System Torque Wrench Recalled for Out-of-Specification Calibration

    Icotec Ag is recalling six VADER Pedicle System Torque Wrench devices due to a calibration error where the torque limiting function was found to be out of specification, delivering 4 Nm instead of the specified 12 Nm.

    Product
    VADER Pedicle System Torque Wrench, Hudson Coupling, 12 Nm, REF 42-702
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-0748-2023·2022-12-28

    HeartWare HVAD Pump Kit driveline cover hardening may impair urgent access

    HeartWare HVAD Pump Kit driveline covers may harden over time, making them difficult to remove for urgent servicing. In an emergency scenario requiring driveline access, delays could harm patients.

    Product
    HeartWare HVAD Pump Kit, REF 1104JP
    Category
    Medical Device
    Distribution
    0 states
  • HighFDA (Devices)·Z-0721-2023·2022-12-28

    Medical Imaging Software Update May Hide Patient Studies

    After installing the IAS tool, Centricity PACS-IW Universal Viewer version 5.0 may hide some medical imaging studies without notification. Accessing hidden studies requires GE Healthcare assistance and may cause diagnostic delays.

    Product
    Centricity PACS-IW with Universal Viewer version 5.0. Used to display medical images (Including mammograms) and data from various imaging sources.
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-0733-2023·2022-12-28

    GE B125M Patient Monitor CO2 Measurement Inaccuracy at High Altitude

    GE Healthcare is recalling B125M patient monitors because the CO2 measurement display can be inaccurate at locations above sea level when configured for mmHg or kPa units.

    Product
    B125M Patient Monitor, REF 6160000-005-XXXXXXX
    Category
    Medical Device
    Distribution
    0 states
  • HighFDA (Devices)·Z-0752-2023·2022-12-28

    GE Vivid S5 Ultrasound Systems Battery Failure Fire Risk

    GE Medical Systems is recalling Vivid S5 ultrasound systems because batteries may fail if not replaced at the recommended 2-year intervals, potentially causing smoke or fire.

    Product
    Vivid S5 Ultrasound. Used in ultrasound imaging and analysis in Fetal/Obstetrics; Abdominal/GYN; Pediatric; Small Organ (breast, testes, thyroid); Neonatal Cephalic; Adult Cephalic; Cardiac (adult and pediatric); Peripheral Vascular; Musculo-skeletal Conventional and…
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-0740-2023·2022-12-28

    Abbott Alinity m Resp-4-Plex Amp Kit Recalled for False Positive Results

    Abbott Molecular is recalling approximately 12,116 units of the Abbott Alinity m Resp-4-Plex Amp Kit due to reports of false positive results and reactive negative controls. No illnesses or injuries have been reported.

    Product
    Abbott Alinity m Resp-4-Plex Amp Kit, Part Number 09N79-090
    Category
    Medical Device
    Distribution
    0 states
  • HighFDA (Food)·F-0198-2023·2022-12-28

    Circle K frozen pizza dough recalled for foreign object contamination

    Circle K 22 oz Blonde Pizza Dough balls are being recalled due to potential contamination with foreign objects from packaging. Consumers with affected lots should not consume the product.

    Product
    Circle K 22 oz Blonde Pizza Dough balls, 20 ct/case; frozen pizza dough
    Category
    Food
    Distribution
    5 states
  • HighFDA (Devices)·Z-0735-2023·2022-12-28

    Natus Cranial Access Kit Recalled Due to Defective Sterile Surgical Drapes

    Natus Medical is recalling 249 Cranial Access Kits distributed nationwide. The sterile surgical drape liners are difficult to remove without damage, potentially rendering the kits unusable for neurosurgical procedures.

    Product
    The Natus Cranial Access Kit is designed to be a complete, single use and disposable accessory and instrumentation set that is used to create an intracranial burr-hole typically required for procedures such as ventriculostomy or intracranial pressure (ICP) monitoring. The…
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Drugs)·D-0088-2023·2022-12-28

    Prescription Drug Quinapril Recalled for Chemical Impurity Exceeding Safety Limits

    Lupin Pharmaceuticals is recalling Quinapril 40mg tablets due to detection of N-Nitroso-quinapril impurity above acceptable daily intake limits. The recall affects approximately 30,612 bottles distributed nationwide.

    Product
    QUINAPRIL — QUINAPRIL (QUINAPRIL)
    Category
    Drug
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-0747-2023·2022-12-28

    HeartWare HVAD Pump Kit driveline cover hardening may prevent repairs

    A hardened driveline cover on the HeartWare HVAD Pump Kit may prevent or delay access to the driveline to controller connector. If urgent servicing is needed, the delay could result in patient harm from pump malfunction.

    Product
    HeartWare HVAD Pump Kit, REF 1104CA-CLIN
    Category
    Medical Device
    Distribution
    0 states
  • HighFDA (Devices)·Z-0758-2023·2022-12-28

    Vivid i N ultrasound device batteries may fail, smoke, or catch fire

    GE Medical Systems recalls Vivid i N ultrasound machines worldwide because unreplaced batteries may fail, emit smoke, or catch fire. Users should replace batteries every 2 years per the service manual.

    Product
    Vivid i N ultrasound. Not marketed in the US. Used in ultrasound imaging and analysis in Fetal/Obstetrics; Abdominal/GYN; Pediatric; Small Organ (breast, testes, thyroid); Neonatal Cephalic; Adult Cephalic; Cardiac (adult and pediatric); Peripheral Vascular…
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-0762-2023·2022-12-28

    Baxter Revaclear 300 Dialyzers Recalled for Polyurethane Displacement

    Baxter is recalling approximately 14 million units of Revaclear 300 dialyzers worldwide due to polyurethane displacement that could cause blood leaks. No injuries have been reported.

    Product
    Revaclear 300 dialyzers. Indicated for treatment of chronic and acute renal failure by hemodialysis.
    Category
    Medical Device
    Distribution
    Distributed nationwide