The Recall Desk

Recalls

All US Product Recalls

Searchable archive of product recalls from the FDA, USDA FSIS, CPSC, NHTSA, and EPA — translated into plain English, severity-scored from objective criteria, and always linked back to the original government notice.

How to read this listing

Every recall on this list has been scored 1–5 against an objective rubric from the agency’s own classification, presence of reported injuries or deaths, distribution scope, and hazard type. The colored severity badge on each card is the score — 5 (Critical) means the agency’s own highest classification or reported deaths or serious injuries, 4 (Severe) means significant injury or hospitalization risk, and 3 (High) means a known hazard with no confirmed harm yet. Filter to 3+ (High and above) if you want to skip routine Class-III notices.

The agency chips above control which streams are visible. FDA spans food, drugs, and medical devices. USDA FSIS is meat, poultry, and processed egg products. CPSC is household consumer products. NHTSA is vehicles, tires, and child seats. Nationwide on a card means the recall affects every US state — most do — while a state list means distribution was limited.

Click any recall to see the full summary, affected lot codes and model numbers, the agency’s own classification, and a direct link back to the original government notice for verification.

Filters

Agency

Category

Minimum severity

Clear filters

3126–3150 of 90254

  • HighFDA (Devices)·Z-1762-2026·2026-04-08

    Burlington Medical half aprons may lose radiation attenuation over time

    Burlington Medical is recalling half aprons because of the potential for attenuation degradation over time, decreasing the lifespan of the product.

    Product
    Burlington Medical, Half Aprons.
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-1660-2026·2026-04-08

    Philips Allura Xper FD10F X-ray may not start from foot switch

    Philips is recalling the Allura Xper FD10F because X-ray imaging may not be initiated, or may occur intermittently, when using the wired foot switch.

    Product
    ALLURA Xper FD10F; System Code: 722002;
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Drugs)·D-0414-2026·2026-04-08

    GNP redness and dry eye relief drops recalled over sterility assurance

    GNP Eye Drops Redness and Dry Eye Relief, containing glycerin and naphazoline hydrochloride, are being recalled for lack of assurance of sterility.

    Product
    LEADER EYE DROPS REDNESS RELIEF — LEADER EYE DROPS REDNESS RELIEF (GLYCERIN, NAPHAZOLINE HYDROCHLORIDE)
    Category
    Drug
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-1752-2026·2026-04-08

    Paradigm REAL-Time Veo pump delivery varies with pump height changes

    Medtronic is recalling Paradigm REAL-Time Veo insulin pumps because raising or lowering the pump relative to the infusion site can cause unintended over- or under-delivery of insulin.

    Product
    Paradigm REAL-Time Veo Insulin Pump (MMT-554, MMT-754)
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-1662-2026·2026-04-08

    Philips Allura Xper FD10/10 X-ray may not start from foot switch

    Philips is recalling the Allura Xper FD10/10 because X-ray imaging may not be initiated, or may occur intermittently, when using the wired foot switch.

    Product
    ALLURA Xper FD10/10; System Code: (1) 722005, (2)722011, (3)722027;
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-1699-2026·2026-04-08

    da Vinci grasping retractor instruments recalled over broken grip cables

    Intuitive Surgical is recalling da Vinci S and Si Grasping Retractor Instruments following increased complaints of broken or frayed grip cables in reusable instruments.

    Product
    Brand Name: da Vinci S, Si Grasping Retractor Instrument Product Name: da Vinci S, Si Grasping Retractor Instrument Model/Catalog Number: 420278 Software Version: NA Product Description: EndoWrist Instruments or Da Vinci S/Si Reusable instruments, including blunt and sharp…
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-1706-2026·2026-04-08

    Medline bronchoscopy packs contain recalled Olympus biopsy valves shedding fragments

    Medline is recalling three bronchoscopy convenience kits because they contain recalled Olympus biopsy valves, where rubber fragments may detach from the valve slit during use.

    Product
    Medline convenience kits containing Olympus biopsy valves: 1. BRONCH PACK, DYKE2096 2. NON MB BRONCHOPSCOPY CCH LH PK, DYNJT6182 3. PULMONARY BRONCH PACK, DYNDA2861A
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-1689-2026·2026-04-08

    Jiffy composite polisher cups may crumble and break apart easily

    Ultradent Products is recalling one lot of Jiffy Original Composite Polisher Cups because the cups may crumble and break apart easily, which could delay patient treatment.

    Product
    Jiffy Original Composite Polisher Cups (Coarse) 12pk, REF 7011.
    Category
    Medical Device
    Distribution
    31 states
  • HighFDA (Devices)·Z-1751-2026·2026-04-08

    MiniMed 530G insulin pump delivery varies with pump height changes

    Medtronic is recalling MiniMed 530G insulin pumps because raising or lowering the pump relative to the infusion site can cause unintended over- or under-delivery of insulin.

    Product
    MiniMed 530G Insulin Pump (MMT-551, MMT-751)
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-1741-2026·2026-04-08

    MiniMed 780G insulin pump delivery varies with pump height changes

    Medtronic is recalling MiniMed 780G insulin pumps because raising or lowering the pump relative to the infusion site can cause unintended over- or under-delivery of insulin.

    Product
    MiniMed 780G Insulin Pump (MMT-1884, MMT-1885, MMT-1886, MMT-1894, MMT-1895, MMT-1896)
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-1748-2026·2026-04-08

    Paradigm insulin pump delivery varies with pump height changes

    Medtronic is recalling Paradigm insulin pumps models MMT-712 and MMT-715 because raising or lowering the pump relative to the infusion site can cause unintended over- or under-delivery of insulin.

    Product
    Paradigm Insulin Pump (MMT-712, MMT-715)
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-1711-2026·2026-04-08

    Medline anterior hip pack contains recalled Olympus biopsy valves

    Medline is recalling the Anterior Hip Pack because it contains recalled Olympus biopsy valves, where rubber fragments may detach from the valve slit during use.

    Product
    MEDLINE ANTERIOR HIP PACK DYNJ64672B
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-1664-2026·2026-04-08

    Philips Allura Xper FD20 Biplane X-ray may not start reliably

    Philips is recalling the Allura Xper FD20 Biplane because X-ray imaging may not be initiated, or may occur intermittently, when using the wired foot switch.

    Product
    ALLURA Xper FD20 Biplane; System Code: (1) 722008, (2)722013;
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-1764-2026·2026-04-08

    Burlington Medical blockers may lose radiation attenuation over time

    Burlington Medical is recalling blockers because of the potential for attenuation degradation over time, decreasing the lifespan of the product.

    Product
    Burlington Medical, Blockers.
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-1731-2026·2026-04-08

    Civco eTRAX 12G starter kit may misreport needle tip position

    Civco is recalling the eTRAX Needle System Starter Kit 12G for Aurora trackers because an inspection and programming error may cause the needle tip position to be incorrectly identified.

    Product
    eTRAX Needle System Starter Kit 12G (for Aurora Trackers), Part Number 667-149
    Category
    Medical Device
    Distribution
    3 states
  • HighFDA (Devices)·Z-1707-2026·2026-04-08

    Medline thoracic robot kits contain recalled Olympus biopsy valves

    Medline is recalling Thoracic Robots kits, model DYNJ908777B, because they contain recalled Olympus biopsy valves where rubber fragments may detach from the valve slit.

    Product
    THORACIC ROBOTS, DYNJ908777B
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Drugs)·D-0421-2026·2026-04-08

    Prazosin hydrochloride further lots recalled over nitrosamine impurity levels

    A further quantity of Prazosin Hydrochloride tablets is being recalled after detection of the nitrosamine impurity N-nitroso-prazosin impurity C above acceptable limits.

    Product
    PRAZOSIN HYDROCHLORIDE — PRAZOSIN HYDROCHLORIDE (PRAZOSIN HYDROCHLORIDE)
    Category
    Drug
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-1673-2026·2026-04-08

    Philips Azurion 3 M12 X-ray may not start from foot switch

    Philips is recalling the Azurion 3 M12 because X-ray imaging may not be initiated, or may occur intermittently, when using the wired foot switch.

    Product
    Azurion 3 M12; System Code: (1)722063, (2)722221, (3)722229;
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Drugs)·D-0408-2026·2026-04-08

    Amneal tramadol tablets failed nitrosamine impurity specification at stability

    Amneal Pharmaceuticals is recalling two lots of Tramadol Hydrochloride Tablets 50 mg after a nitrosamine impurity was found out of specification at the 24-month stability interval.

    Product
    TRAMADOL HYDROCHLORIDE — TRAMADOL HYDROCHLORIDE (TRAMADOL HYDROCHLORIDE)
    Category
    Drug
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-1733-2026·2026-04-08

    Civco eTRAX needle starter kit may misreport needle tip position

    Civco is recalling the eTRAX Needle System Starter Kit 16G for Aurora trackers because an inspection and programming error may cause the needle tip position to be incorrectly identified.

    Product
    eTRAX Needle System Starter Kit 16G(for Aurora Trackers), Part Number 667-151
    Category
    Medical Device
    Distribution
    3 states
  • HighFDA (Devices)·Z-1676-2026·2026-04-08

    Philips Azurion 7 B20 X-ray may not start from foot switch

    Philips is recalling the Azurion 7 B20 because X-ray imaging may not be initiated, or may occur intermittently, when using the wired foot switch.

    Product
    Azurion 7 B20; System Code: (1)722068, (2)722226, (3)722236;
    Category
    Medical Device
    Distribution
    Distributed nationwide