The Recall Desk

Recalls

All US Product Recalls

Searchable archive of product recalls from the FDA, USDA FSIS, CPSC, NHTSA, and EPA — translated into plain English, severity-scored from objective criteria, and always linked back to the original government notice.

How to read this listing

Every recall on this list has been scored 1–5 against an objective rubric from the agency’s own classification, presence of reported injuries or deaths, distribution scope, and hazard type. The colored severity badge on each card is the score — 5 (Critical) means deaths or serious injuries are reported, 4 (Severe) means significant injury or hospitalization risk, and 3 (High) means a known hazard with no confirmed harm yet. Filter to 3+ (High and above) if you want to skip routine Class-III notices.

The agency chips above control which streams are visible. FDA spans food, drugs, and medical devices. USDA FSIS is meat, poultry, and processed egg products. CPSC is household consumer products. NHTSA is vehicles, tires, and child seats. Nationwide on a card means the recall affects every US state — most do — while a state list means distribution was limited.

Click any recall to see the full summary, affected lot codes and model numbers, the agency’s own classification, and a direct link back to the original government notice for verification.

Filters

Agency

Category

Minimum severity

Clear filters

2426–2450 of 39987

  • HighFDA (Devices)·Z-0657-2026·2025-12-03

    IV set 490427 recalled over piggyback backflow and priming occlusion

    B. Braun Medical is recalling 224,856 units iv administration sets over the potential for medication backflow from secondary piggyback containers and an inability to prime.

    Product
    IV Administration Set utilized in gravity IV administration sets and pump administration sets utilized with BBMI's Infusomat Space Large Volume Pump, Outlook Pump, and Vista Basic Pump. Labeled as the following with corresponding catalog number: 1. IV ADMIN SET W/ 15 DROPS/ML
    Category
    Medical Device
    Distribution
    0 states
  • HighFDA (Devices)·Z-0652-2026·2025-12-03

    IV set 490241 recalled over backflow and inability to prime

    B. Braun Medical is recalling 66,792 IV administration sets over the potential for medication backflow from secondary piggyback containers and an inability to prime.

    Product
    IV Administration Set utilized in gravity IV administration sets and pump administration sets utilized with BBMI's Infusomat Space Large Volume Pump, Outlook Pump, and Vista Basic Pump. Labeled as the following with corresponding catalog number: 1. IV ADMIN SET HF STOPCOCK,10
    Category
    Medical Device
    Distribution
    0 states
  • HighFDA (Devices)·Z-0600-2026·2025-12-03

    Illumination packs recalled because packaging tape contains undisclosed latex

    Stryker is recalling 39,148 NICO Myriad Illumination Packs because tape used to secure components in the packaging contains latex despite the products being labelled latex-free.

    Product
    NICO Myriad Illumination Pack Product No. NN-1000 11g Illumination Sleeve/Fiber; Product No. NN-1001 13g Illumination Sleeve/Fiber
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-0639-2026·2025-12-03

    Extension set V5405 recalled over backflow and inability to prime

    B. Braun Medical is recalling 300 units extension sets over the potential for medication backflow from secondary piggyback containers and an inability to prime.

    Product
    Extension Set utilized in gravity IV administration sets and pump administration sets utilized with BBMI's Infusomat Space Large Volume Pump, Outlook Pump, and Vista Basic Pump. Labeled as the following with corresponding catalog number: 1. EXT/2INJ/FLBL/FML LRLK/CV 34; Catal
    Category
    Medical Device
    Distribution
    0 states
  • HighFDA (Devices)·Z-0636-2026·2025-12-03

    Extension set 490238 recalled over piggyback backflow and occlusion risk

    B. Braun Medical is recalling 11,800 units extension sets over the potential for medication backflow from secondary piggyback containers and an inability to prime.

    Product
    Extension Set utilized in gravity IV administration sets and pump administration sets utilized with BBMI's Infusomat Space Large Volume Pump, Outlook Pump, and Vista Basic Pump. Labeled as the following with corresponding catalog number: 1. EXT SET W/ 6 B/C VALVES, 17 IN; Cat
    Category
    Medical Device
    Distribution
    0 states
  • HighFDA (Devices)·Z-0637-2026·2025-12-03

    Extension set 456521 recalled over backflow and priming occlusion

    B. Braun Medical is recalling 61,423 units extension sets over the potential for medication backflow from secondary piggyback containers and an inability to prime.

    Product
    Extension Set utilized in gravity IV administration sets and pump administration sets utilized with BBMI's Infusomat Space Large Volume Pump, Outlook Pump, and Vista Basic Pump. Labeled as the following with corresponding catalog numbers: 1. ANESTHESIA SET W/ SAFEPORT MANIFOL
    Category
    Medical Device
    Distribution
    0 states
  • HighFDA (Devices)·Z-0662-2026·2025-12-03

    IV set 490367 recalled over medication backflow and priming failure

    B. Braun Medical is recalling 6,768 IV administration sets over the potential for medication backflow from secondary piggyback containers and an inability to prime.

    Product
    IV Administration Set utilized in gravity IV administration sets and pump administration sets utilized with BBMI's Infusomat Space Large Volume Pump, Outlook Pump, and Vista Basic Pump. Labeled as the following with corresponding catalog number: 1. IV ADMIN SET W/60 DR/ML, CA
    Category
    Medical Device
    Distribution
    0 states
  • HighFDA (Devices)·Z-0646-2026·2025-12-03

    IV set 457506 recalled over piggyback backflow and occlusion risk

    B. Braun Medical is recalling 1,019 IV administration sets over the potential for medication backflow from secondary piggyback containers and an inability to prime.

    Product
    IV Administration Set utilized in gravity IV administration sets and pump administration sets utilized with BBMI's Infusomat Space Large Volume Pump, Outlook Pump, and Vista Basic Pump. Labeled as the following with corresponding catalog number: 1. IV SET W/2 CARESITE & 1 ULT
    Category
    Medical Device
    Distribution
    0 states
  • HighFDA (Devices)·Z-0638-2026·2025-12-03

    Caresite extension sets recalled over piggyback backflow and occlusion

    B. Braun Medical is recalling 32,500 Caresite extension sets over the potential for medication backflow from secondary piggyback containers and an inability to prime.

    Product
    Extension Set utilized in gravity IV administration sets and pump administration sets utilized with BBMI's Infusomat Space Large Volume Pump, Outlook Pump, and Vista Basic Pump. Labeled as the following with corresponding catalog number: 1. Caresite Extension Set_Export; Cata
    Category
    Medical Device
    Distribution
    0 states
  • HighFDA (Devices)·Z-0583-2026·2025-12-03

    Skin cancer aid recalled because it may not meet specifications

    DermaSensor is recalling nine affected units of its software-aided diagnostic device because it may not meet specifications, risking incorrect results or a delayed referral.

    Product
    DermaSensor. Model/Catalog Number: 10101. Software Version: 3.0.2. Software-aided adjunctive diagnostic device for use by physicians on lesions suspicious for skin cancer
    Category
    Medical Device
    Distribution
    8 states
  • HighFDA (Devices)·Z-0708-2026·2025-12-03

    Outlook pump sets recalled over backflow into the primary container

    B. Braun Medical is recalling 2,352 Outlook pump sets over the potential for medication backflow from secondary piggyback containers and an inability to prime.

    Product
    Outlook set utilized in gravity IV administration sets and pump administration sets utilized with BBMI's Infusomat Space Large Volume Pump, Outlook Pump, and Vista Basic Pump. Labeled as the following with corresponding catalog number: 1. OUTLOOK PUMP SET 3 CARESITE LADS 133
    Category
    Medical Device
    Distribution
    0 states
  • HighFDA (Devices)·Z-0659-2026·2025-12-03

    IV set 490242 recalled over backflow into the primary bag

    B. Braun Medical is recalling 129,480 IV administration sets over the potential for medication backflow from secondary piggyback containers and an inability to prime.

    Product
    IV Administration Set utilized in gravity IV administration sets and pump administration sets utilized with BBMI's Infusomat Space Large Volume Pump, Outlook Pump, and Vista Basic Pump. Labeled as the following with corresponding catalog number: 1. IV ADMIN SET 60 DROP/ML, 10
    Category
    Medical Device
    Distribution
    0 states
  • HighFDA (Devices)·Z-0690-2026·2025-12-03

    IV set 490303 recalled over piggyback backflow and occlusion risk

    B. Braun Medical is recalling 32,304 IV administration sets over the potential for medication backflow from secondary piggyback containers and an inability to prime.

    Product
    IV Administration Set utilized in gravity IV administration sets and pump administration sets utilized with BBMI's Infusomat Space Large Volume Pump, Outlook Pump, and Vista Basic Pump. Labeled as the following with corresponding catalog number: 1. IV ADMIN SET,3 CARESITE LAD
    Category
    Medical Device
    Distribution
    0 states
  • HighFDA (Devices)·Z-0618-2026·2025-12-03

    Blood set 490530 recalled over backflow and inability to prime

    B. Braun Medical is recalling 33,528 blood administration sets over the potential for medication backflow from secondary piggyback containers and an inability to prime.

    Product
    Blood Administration Set utilized in gravity IV administration sets and pump administration sets utilized with BBMI's Infusomat Space Large Volume Pump, Outlook Pump, and Vista Basic Pump. Labeled as the following with corresponding catalog number: 1. Y-TYPE BLOOD SET, 3 CARS
    Category
    Medical Device
    Distribution
    0 states
  • HighFDA (Devices)·Z-0599-2026·2025-12-03

    Myriad handpieces recalled because packaging tape contains undisclosed latex

    Stryker is recalling 39,148 NICO Myriad Handpieces because tape used to secure components in the packaging contains latex despite the products being labelled latex-free.

    Product
    NICO Myriad Handpiece Product Number/Product Name: NN-8000 15x10 MHP; NN-8001 15x 13 MHP; NN-8002 15x 25 MHP - MINOP; NN-8003 13x10 MHP; NN-8004 13x13 MHP; NN-8005 15x25 MHP GAAB; NN-8006 11X10 MHP; NN-8007 11x13 MHP; NN-8008 19x28 MHP OI; NN-8010 17x31.
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-0693-2026·2025-12-03

    Standard IV sets recalled over medication backflow from piggyback containers

    B. Braun Medical is recalling 41,976 IV administration sets over the potential for medication backflow from secondary piggyback containers and an inability to prime.

    Product
    IV Administration Set utilized in gravity IV administration sets and pump administration sets utilized with BBMI's Infusomat Space Large Volume Pump, Outlook Pump, and Vista Basic Pump. Labeled as the following with corresponding catalog number: 1. IV ADMIN SET W/15 DROPS/ML
    Category
    Medical Device
    Distribution
    0 states
  • HighFDA (Devices)·Z-0679-2026·2025-12-03

    IV set 490025 recalled over backflow into the primary container

    B. Braun Medical is recalling 15,250 units iv administration sets over the potential for medication backflow from secondary piggyback containers and an inability to prime.

    Product
    IV Administration Set utilized in gravity IV administration sets and pump administration sets utilized with BBMI's Infusomat Space Large Volume Pump, Outlook Pump, and Vista Basic Pump. Labeled as the following with corresponding catalog number: 1. PB1500 15 DROP IV SET W/2 S
    Category
    Medical Device
    Distribution
    0 states
  • HighFDA (Devices)·Z-0670-2026·2025-12-03

    IV set 352636 recalled over piggyback backflow and priming occlusion

    B. Braun Medical is recalling 20,694,734 units iv administration sets over the potential for medication backflow from secondary piggyback containers and an inability to prime.

    Product
    IV Administration Set utilized in gravity IV administration sets and pump administration sets utilized with BBMI's Infusomat Space Large Volume Pump, Outlook Pump, and Vista Basic Pump. Labeled as the following with corresponding catalog number: 1. IV ADMIN SET 15 DR W/2 SAFE
    Category
    Medical Device
    Distribution
    0 states
  • HighFDA (Devices)·Z-0608-2026·2025-12-03

    Anesthesia IV sets recalled over backflow and inability to prime

    B. Braun Medical is recalling 14,712 anesthesia IV sets over the potential for medication backflow from secondary piggyback containers and an inability to prime.

    Product
    Anesthesia IV Set utilized in gravity IV administration sets and pump administration sets utilized with BBMI's Infusomat Space Large Volume Pump, Outlook Pump, and Vista Basic Pump. Labeled as the following with corresponding catalog number: 1. IV AD. SET,15 DROPS/ML, W/EXT S
    Category
    Medical Device
    Distribution
    0 states
  • HighFDA (Devices)·Z-0706-2026·2025-12-03

    Outlook infusion sets recalled over backflow and inability to prime

    B. Braun Medical is recalling 6,672 Outlook sets over the potential for medication backflow from secondary piggyback containers and an inability to prime.

    Product
    Outlook set utilized in gravity IV administration sets and pump administration sets utilized with BBMI's Infusomat Space Large Volume Pump, Outlook Pump, and Vista Basic Pump. Labeled as the following with corresponding catalog number: 1. OUTLOOK PUMP SET,3 CARESITE LADS,115
    Category
    Medical Device
    Distribution
    0 states
  • HighFDA (Devices)·Z-0594-2026·2025-12-03

    Diabetes clinic software recalled over incorrect sensor glucose graph data

    Medtronic MiniMed is recalling 12,126 CareLink Clinic installations because a software error displays incorrect data on the 24-hour sensor glucose overview graph.

    Product
    CareLink Clinic, REF: MMT-7350
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-0721-2026·2025-12-03

    Isoflurane vaporizers recalled over the potential for anesthesia leakage

    Mindray is recalling 54 Isoflurane V90 Electronic Vaporizers used with the A9 Anesthesia System over the potential for anesthesia leakage.

    Product
    Isoflurane (ISO) V90 Electronic Vaporizer with the A9 Anesthesia System. Labeled as the following: 1. Iso E_Vap. (Safety Filling adapter), Part Number 115-066758-00. 2. Exchange Iso E_Vap. (Safety Filling adapter), Part Number E115-066758-00. Product Description: The
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-0695-2026·2025-12-03

    IV set 490193 recalled over backflow and inability to prime

    B. Braun Medical is recalling 111,312 units iv administration sets over the potential for medication backflow from secondary piggyback containers and an inability to prime.

    Product
    IV Administration Set utilized in gravity IV administration sets and pump administration sets utilized with BBMI's Infusomat Space Large Volume Pump, Outlook Pump, and Vista Basic Pump. Labeled as the following with corresponding catalog number: 1. IV ADMIN SET 15DR HF 4-WAY
    Category
    Medical Device
    Distribution
    0 states
  • HighFDA (Devices)·Z-0711-2026·2025-12-03

    SafeDAY set 352635 recalled over piggyback backflow and inability to prime

    B. Braun Medical is recalling 200,917 SafeDAY sets over the potential for medication backflow from secondary piggyback containers and an inability to prime.

    Product
    SafeDAY set utilized in gravity IV administration sets and pump administration sets utilized with BBMI's Infusomat Space Large Volume Pump, Outlook Pump, and Vista Basic Pump. Labeled as the following with corresponding catalog number: 1. IV ADMIN SET 60 DROP W/2 SAFEDAY; Cat
    Category
    Medical Device
    Distribution
    0 states
  • HighFDA (Devices)·Z-0682-2026·2025-12-03

    IV set 490488 recalled over backflow between primary and secondary containers

    B. Braun Medical is recalling 187,464 IV administration sets over the potential for medication backflow from secondary piggyback containers and an inability to prime.

    Product
    IV Administration Set utilized in gravity IV administration sets and pump administration sets utilized with BBMI's Infusomat Space Large Volume Pump, Outlook Pump, and Vista Basic Pump. Labeled as the following with corresponding catalog number: 1. IV ADMIN SET W/15 DROPS/ML,
    Category
    Medical Device
    Distribution
    0 states