The Recall Desk

Recalls

All US Product Recalls

Searchable archive of product recalls from the FDA, USDA FSIS, CPSC, NHTSA, and EPA — translated into plain English, severity-scored from objective criteria, and always linked back to the original government notice.

How to read this listing

Every recall on this list has been scored 1–5 against an objective rubric from the agency’s own classification, presence of reported injuries or deaths, distribution scope, and hazard type. The colored severity badge on each card is the score — 5 (Critical) means the agency’s own highest classification or reported deaths or serious injuries, 4 (Severe) means significant injury or hospitalization risk, and 3 (High) means a known hazard with no confirmed harm yet. Filter to 3+ (High and above) if you want to skip routine Class-III notices.

The agency chips above control which streams are visible. FDA spans food, drugs, and medical devices. USDA FSIS is meat, poultry, and processed egg products. CPSC is household consumer products. NHTSA is vehicles, tires, and child seats. Nationwide on a card means the recall affects every US state — most do — while a state list means distribution was limited.

Click any recall to see the full summary, affected lot codes and model numbers, the agency’s own classification, and a direct link back to the original government notice for verification.

Filters

Agency

Category

Minimum severity

Clear filters

21451–21475 of 40339

  • HighFDA (Devices)·Z-0781-2019·2019-02-06

    Siemens Atellica Analyzer Software Recall for Multiple System Issues

    Siemens Healthcare Diagnostics is recalling Atellica analyzer software versions 1.14.2 and lower due to multiple identified issues affecting in vitro diagnostic testing systems.

    Product
    Atellica IM 1600 Analyzer: In vitro diagnostic testing of clinical specimens - Product Usage: The Atellica Solution is a multi-component system for in vitro diagnostic testing of clinical specimens. The system is intended for the qualitative and quantitative analysis of various…
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • ModerateFDA (Devices)·Z-0774-2019·2019-02-06

    Wilson-Cook Medical Recalls Mislabelled Captura Serrated Biopsy Forceps

    Wilson-Cook Medical is recalling 8 units of Captura Serrated Biopsy Forceps in France, Ireland, and Japan due to mislabeling.

    Product
    COOK MEDICAL Captura SERRATED BIOPSY FORCEPS WITH SPIKE,2.4 mm, 230 cm, with Spike, DBF-2.4SN-230SP-20-S
    Category
    Medical Device
    Distribution
    0 states
  • ModerateFDA (Devices)·Z-0783-2019·2019-02-06

    Siemens Recalls Atellica Sample Handler Prime Due to Software Issues

    Siemens Healthcare Diagnostics is recalling 410 Atellica Sample Handler Prime units because software version 1.14.2 and lower contain multiple issues.

    Product
    Atellica Sample Handler Prime. In vitro diagnostic testing of clinical specimens - Siemens Material Number (SMN): 11069001 - Product Usage: The Atellica Solution is a multi-component system for in vitro diagnostic testing of clinical specimens. The system is intended for the…
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • ModerateFDA (Devices)·Z-0789-2019·2019-02-06

    Abbott Recalls Alinity i Processing Modules Due to Gear Pump Assembly Issue

    Abbott is recalling Alinity i Processing Modules because a gear pump assembly issue causes foaming and dried residue in the wash buffer reservoir.

    Product
    Alinity i Processing Module, (a) Part Number 03R65-01 (b) Part Number A-30108552-01 (c) Part Number A-30109694-01 Product Usage: The Alinity i processing module is a fully automated immunoassay analyzer allowing random and continuous access, as well as priority and…
    Category
    Medical Device
    Distribution
    0 states
  • ModerateFDA (Devices)·Z-0775-2019·2019-02-06

    Wilson-Cook Medical Recalls Mislabeled Captura Serrated Forceps

    Wilson-Cook Medical is recalling 6 units of Captura Serrated Forceps due to mislabeling that swapped spike and no-spike models.

    Product
    COOK MEDICAL Captura SERRATED LARGE FORCEP-NO SPIKE, 2.4 mm, 230 cm, DBF-2.4SL- 230-20-S
    Category
    Medical Device
    Distribution
    0 states
  • LowFDA (Devices)·Z-0777-2019·2019-02-06

    Riverpoint Medical Recalls 18G Brachy Grids for Incorrect CE Labeling

    Riverpoint Medical is recalling 443 units of 18G Brachy Grids because they are labeled with a CE mark despite not being approved in the European Union.

    Product
    RP RIVERPOINT MEDICAL 18G Brachy Grid B&K(TM) style, REF RPG-18 Product Usage: The Riverpoint Brachytherapy Grid is used during Brachytherapy procedures to assist with the precise insertion of the needle or other probes. The Riverpoint Brachytherapy Grid is placed against…
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • CriticalFDA (Devices)·Z-0737-2019·2019-01-30

    Draeger Medical Recalls VentStar Basic 250 Breathing Circuits Due to Assembly Error

    Draeger Medical is recalling 170 units of VentStar Basic 250 breathing circuits because they were incorrectly pre-assembled, creating a risk of incorrect connection or short circuiting.

    Product
    VentStar Basic 250, Catalog Number MP00350 Breathing Circuits/Anesthesia Sets without a Water trap: Breathing circuit for conduction of respiratory gases between anesthesia machine or ventilator and patient. Intended for single use only. Breathing Circuits/Anesthesia Sets…
    Category
    Medical Device
    Distribution
    12 states
  • SevereFDA (Devices)·Z-0736-2019·2019-01-30

    Draeger Medical Recalls VentStar Basic 180 Breathing Circuits

    Draeger Medical is recalling 325 units of VentStar Basic 180 breathing circuits due to incorrect pre-assembly that may cause incorrect connections.

    Product
    VentStar Basic 180, Catalog Number MP00349 Breathing Circuits/Anesthesia Sets without a Water trap: Breathing circuit for conduction of respiratory gases between anesthesia machine or ventilator and patient. Intended for single use only. Breathing Circuits/Anesthesia Sets…
    Category
    Medical Device
    Distribution
    12 states
  • SevereFDA (Devices)·Z-0738-2019·2019-01-30

    Draeger Medical Recalls VentStar Basic Breathing Circuits Due to Assembly Error

    Draeger Medical is recalling 525 units of VentStar Basic (P) 180 breathing circuits because they were incorrectly pre-assembled, posing a risk of incorrect connection or short circuiting.

    Product
    VentStar Basic (P) 180, Catalog Number MP00351 Breathing Circuits/Anesthesia Sets without a Water trap: Breathing circuit for conduction of respiratory gases between anesthesia machine or ventilator and patient. Intended for single use only. Breathing Circuits/Anesthesia…
    Category
    Medical Device
    Distribution
    12 states
  • SevereFDA (Drugs)·D-0383-2019·2019-01-30

    Bound Tree Medical Recalls Epi Safe Kits Due to Incorrect Instructions

    Bound Tree Medical is recalling Epi Safe Administration and Training Kits because the instructions recommend an unapproved epinephrine dosage for young children.

    Product
    curaplex Epi Safe Administration and Training Kits, #8600-01100, Contains: 2 Epi Safe Administration Kit (8600-01101), 1 Epi Safe Training Kit (8600-01102), Rx Only, Distributed by Sarnova HC, LLC s family of companies: Bound Tree Medical, LLC, Cardio Partners, Inc., Emergency…
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • SevereFDA (Devices)·Z-0740-2019·2019-01-30

    Draeger Medical Recalls VentStar Watertrap Breathing Circuits Due to Assembly Error

    Draeger Medical is recalling 25 units of VentStar Watertrap (P) 180 breathing circuits because they were incorrectly pre-assembled, posing a risk of incorrect connection.

    Product
    VentStar Watertrap (P) 180, Catalog Number MP00361 Breathing Circuits/Anesthesia Sets without a Water trap: Breathing circuit for conduction of respiratory gases between anesthesia machine or ventilator and patient. Intended for single use only. Breathing Circuits/Anesthesia…
    Category
    Medical Device
    Distribution
    12 states
  • SevereFDA (Devices)·Z-0753-2019·2019-01-30

    Teleflex Percuvance Percutaneous Shaft Recall for Potential Tip Breakage

    Teleflex Medical is recalling Percuvance Percutaneous Shafts because the distal tip may deform or break off during use, potentially compromising the tool's security.

    Product
    Percuvance(R) Percutaneous Shaft 29cm, PCVSH3 The Percutaneous Surgical System with 5mm attachments is indicated for the means to penetrate Intended Use: soft tissue to access certain areas of the abdomen. The system is used to grasp, manipulate, cut, cauterize and deliver…
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • SevereFDA (Devices)·Z-0754-2019·2019-01-30

    Teleflex Percuvance Percutaneous Shaft Recall Due to Tip Breakage Risk

    Teleflex Medical is recalling 1,962 units of Percuvance Percutaneous Shafts because the distal tip may deform or break off during use, potentially requiring surgical intervention.

    Product
    Percuvance(R) Percutaneous Shaft 36cm, PCVSHL3 The Percutaneous Surgical System with 5mm attachments is indicated for the means to penetrate Intended Use: soft tissue to access certain areas of the abdomen. The system is used to grasp, manipulate, cut, cauterize and deliver…
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • SevereFDA (Devices)·Z-0742-2019·2019-01-30

    Draeger Medical Recalls ID Circuit Watertrap Breathing Circuits Due to Assembly Error

    Draeger Medical is recalling 25 units of ID Circuit Watertrap breathing circuits because they were incorrectly pre-assembled, creating a risk of incorrect connection or short circuit.

    Product
    ID Circuit Watertrap (P) 180, Catalog Number MP01341 Breathing Circuits/Anesthesia Sets without a Water trap: Breathing circuit for conduction of respiratory gases between anesthesia machine or ventilator and patient. Intended for single use only. Breathing…
    Category
    Medical Device
    Distribution
    12 states
  • SevereFDA (Devices)·Z-0744-2019·2019-01-30

    Draeger ID Circuit Basic 180 Breathing Circuits Recalled for Assembly Defect

    Draeger Medical is recalling 25 units of ID Circuit Basic 180 breathing circuits due to incorrect pre-assembly that may cause connection errors.

    Product
    ID Circuit Basic 180, Catalog Number MP01350 Breathing Circuits/Anesthesia Sets without a Water trap: Breathing circuit for conduction of respiratory gases between anesthesia machine or ventilator and patient. Intended for single use only. Breathing Circuits/Anesthesia Sets…
    Category
    Medical Device
    Distribution
    12 states
  • SevereFDA (Devices)·Z-0743-2019·2019-01-30

    Draeger Medical Recalls ID Circuit Basic 250 Breathing Circuits

    Draeger Medical is recalling 20 units of ID Circuit Basic 250 breathing circuits due to incorrect pre-assembly that may cause connection errors.

    Product
    ID Circuit Basic 250, Catalog Number MP01348 Breathing Circuits/Anesthesia Sets without a Water trap: Breathing circuit for conduction of respiratory gases between anesthesia machine or ventilator and patient. Intended for single use only. Breathing Circuits/Anesthesia Sets…
    Category
    Medical Device
    Distribution
    12 states
  • SevereFDA (Devices)·Z-0739-2019·2019-01-30

    Draeger Medical Recalls VentStar Basic Breathing Circuits Due to Assembly Error

    Draeger Medical is recalling 75 units of VentStar Basic (P) 250 breathing circuits because they were incorrectly pre-assembled, creating a risk of incorrect connection or short circuit.

    Product
    VentStar Basic (P) 250, Catalog Number MP00352 Breathing Circuits/Anesthesia Sets without a Water trap: Breathing circuit for conduction of respiratory gases between anesthesia machine or ventilator and patient. Intended for single use only. Breathing Circuits/Anesthesia…
    Category
    Medical Device
    Distribution
    12 states
  • SevereFDA (Devices)·Z-0741-2019·2019-01-30

    Draeger Medical Recalls 10 VentStar Anesthesia Breathing Circuits

    Draeger Medical is recalling 10 units of VentStar Anesthesia WT(P) 180 breathing circuits due to incorrect pre-assembly that could cause incorrect connections.

    Product
    VentStar Anesthesia WT(P) 180, Catalog Number MP00374 Breathing Circuits/Anesthesia Sets without a Water trap: Breathing circuit for conduction of respiratory gases between anesthesia machine or ventilator and patient. Intended for single use only. Breathing…
    Category
    Medical Device
    Distribution
    12 states
  • SevereFDA (Drugs)·D-0384-2019·2019-01-30

    Recall of Curaplex Epi Safe Kits Due to Incorrect Epinephrine Instructions

    Bound Tree Medical is recalling Curaplex Epi Safe Kits because the instructions recommend an unapproved epinephrine dosage for young children.

    Product
    curaplex Epi Safe Kit, 8600-01101, contains:" 1ml Vial of Epinephrine " 1 Epi-Safe Syringe " 1 Safety Needle " 2 Alcohol Prep Pads " 1 Adhesive Dressing " 1 Insert, Rx Only. Distributed by Sarnova HC, LLC s family of companies: Bound Tree Medical, LLC, Cardio Partners, Inc.…
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-0756-2019·2019-01-30

    Fetzer Medical Recalls Wrigley Obstetric Forceps Due to Functional Defect

    Fetzer Medical is recalling 10 units of Wrigley Obstetric Forceps because doctors reported being unable to pull the instruments apart.

    Product
    12002 / Boss Instruments Ltd. Wrigley Obstetric Forceps, 11" (28cm), Item No./Product Code 46-5004 Fetzer Medical Obstetrical Forceps is a hand-held instruments intended to assist the birth of a foetus during difficult vaginal births through grasping the foetal head in order…
    Category
    Medical Device
    Distribution
    2 states
  • HighFDA (Devices)·Z-0757-2019·2019-01-30

    Fetzer De Lee Obstetric Forceps Recalled Due to Functional Defect

    Fetzer Medical is recalling 10 units of De Lee Obstetric Forceps because doctors reported being unable to pull the instruments apart.

    Product
    16600 / Fetzer Surgical De Lee Obstetric Forceps, 11" (30cm), Item No./Product Code 46-5034 Fetzer Medical Obstetrical Forceps is a hand-held instruments intended to assist the birth of a foetus during difficult vaginal births through grasping the foetal head in order to…
    Category
    Medical Device
    Distribution
    2 states
  • HighFDA (Devices)·Z-0771-2019·2019-01-30

    Varian Eclipse Treatment Planning System Proton Algorithm Recall

    Varian Medical Systems is recalling the Eclipse Treatment Planning System due to a software error in the Proton Convolution Superposition algorithm that miscalculates dose ranges for non-square CT images.

    Product
    Eclipse(TM) Treatment Planning System with Proton Convolution Superposition algorithm license only The Eclipse Treatment Planning System (Eclipse TPS) is used to plan radiotherapy treatments for patients with malignant or benign diseases. Eclipse TPS is used to plan external…
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-0761-2019·2019-01-30

    Stihler PRISMAFLO II Blood Return Warmer Recalled for Burn Risk

    Stihler Electronic Gmbh is recalling PRISMAFLO II blood return warmers because damage to the heating profile may cause low-level skin burns.

    Product
    Heating Profile PRISMAFLO II blood return warmer for Prismaflex Product Code: PF2-WP2618 Product Usage: The PRISMAFLO IIS warmer is used for warming return blood flow.
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-0772-2019·2019-01-30

    Leica M220 F12 Surgical Microscope Recalled for Falling Optics Hazard

    Leica Microsystems is recalling 15 Leica M220 F12 surgical microscopes because the optics may drop into the surgical field, risking contact with the patient.

    Product
    Leica M220 F12 surgical microscope The Leica M220 F12 surgical microscope is an optical instrument for improving the visibility of objects through magnification and illumination. It can be applied for observation and documentation and for human and veterinary medical…
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-0759-2019·2019-01-30

    Stihler PRISMAFLO II S Blood Return Warmer Recalled for Burn Risk

    Stihler Electronic Gmbh is recalling PRISMAFLO II S blood return warmers because damage to the heating profile may cause low-level skin burns.

    Product
    Heating Profile PRISMAFLO II S blood return warmer for Prismaflex Product Code: PF2-WP33 Product Usage: The PRISMAFLO IIS warmer is used for warming return blood flow.
    Category
    Medical Device
    Distribution
    Distributed nationwide