The Recall Desk

Recalls

All US Product Recalls

Searchable archive of product recalls from the FDA, USDA FSIS, CPSC, NHTSA, and EPA — translated into plain English, severity-scored from objective criteria, and always linked back to the original government notice.

How to read this listing

Every recall on this list has been scored 1–5 against an objective rubric from the agency’s own classification, presence of reported injuries or deaths, distribution scope, and hazard type. The colored severity badge on each card is the score — 5 (Critical) means deaths or serious injuries are reported, 4 (Severe) means significant injury or hospitalization risk, and 3 (High) means a known hazard with no confirmed harm yet. Filter to 3+ (High and above) if you want to skip routine Class-III notices.

The agency chips above control which streams are visible. FDA spans food, drugs, and medical devices. USDA FSIS is meat, poultry, and processed egg products. CPSC is household consumer products. NHTSA is vehicles, tires, and child seats. Nationwide on a card means the recall affects every US state — most do — while a state list means distribution was limited.

Click any recall to see the full summary, affected lot codes and model numbers, the agency’s own classification, and a direct link back to the original government notice for verification.

Filters

Agency

Category

Minimum severity

Clear filters

951–975 of 27327

  • HighFDA (Devices)·Z-1672-2026·2026-04-08

    Philips Allura Xper FD20/15 OR Table X-ray may not start

    Philips is recalling the Allura Xper FD20/15 OR Table because X-ray imaging may not be initiated, or may occur intermittently, when using the wired foot switch.

    Product
    Allura Xper FD20/15 OR Table; System Code: 722059;
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-1733-2026·2026-04-08

    Civco eTRAX needle starter kit may misreport needle tip position

    Civco is recalling the eTRAX Needle System Starter Kit 16G for Aurora trackers because an inspection and programming error may cause the needle tip position to be incorrectly identified.

    Product
    eTRAX Needle System Starter Kit 16G(for Aurora Trackers), Part Number 667-151
    Category
    Medical Device
    Distribution
    3 states
  • HighFDA (Devices)·Z-1752-2026·2026-04-08

    Paradigm REAL-Time Veo pump delivery varies with pump height changes

    Medtronic is recalling Paradigm REAL-Time Veo insulin pumps because raising or lowering the pump relative to the infusion site can cause unintended over- or under-delivery of insulin.

    Product
    Paradigm REAL-Time Veo Insulin Pump (MMT-554, MMT-754)
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-1675-2026·2026-04-08

    Philips Azurion 7 B12 X-ray may not start from foot switch

    Philips is recalling the Azurion 7 B12 because X-ray imaging may not be initiated, or may occur intermittently, when using the wired foot switch.

    Product
    Azurion 7 B12; System Code: (1)722067, (2)722225, (3)722235;
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-1661-2026·2026-04-08

    Philips Allura Xper FD10 X-ray may not start from foot switch

    Philips is recalling the Allura Xper FD10 because X-ray imaging may not be initiated, or may occur intermittently, when using the wired foot switch.

    Product
    ALLURA Xper FD10; System Code: (1) 722003, (2)722010, (3)722026;
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-1691-2026·2026-04-08

    Medline ENFit G-tube connector kits may leak from dimensional defect

    Medline is recalling kits containing ENFit G-tube connectors because the connectors were not made to required dimensional specifications and may not form a proper seal.

    Product
    Medline Kits containing ENFIT G-TUBE CONNECTOR 1. SKU DYKM2720 G/J TUBE CARE KIT 2. SKU ENFITDISCKIT ENFIT DISCHARGE KIT
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-1734-2026·2026-04-08

    Civco eTRAX 18G starter kit may misreport needle tip position

    Civco is recalling the eTRAX Needle System Starter Kit 18G for Aurora trackers because an inspection and programming error may cause the needle tip position to be incorrectly identified.

    Product
    eTRAX Needle System Starter Kit 18G (for Aurora Trackers), Part Number 667-152
    Category
    Medical Device
    Distribution
    3 states
  • HighFDA (Devices)·Z-1671-2026·2026-04-08

    Philips Allura Xper FD20/15 X-ray may not start from foot switch

    Philips is recalling the Allura Xper FD20/15 because X-ray imaging may not be initiated, or may occur intermittently, when using the wired foot switch.

    Product
    Allura Xper FD20/15; System Code: 722058;
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-1707-2026·2026-04-08

    Medline thoracic robot kits contain recalled Olympus biopsy valves

    Medline is recalling Thoracic Robots kits, model DYNJ908777B, because they contain recalled Olympus biopsy valves where rubber fragments may detach from the valve slit.

    Product
    THORACIC ROBOTS, DYNJ908777B
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-1708-2026·2026-04-08

    Medline endoscopy kits contain recalled Olympus biopsy valves shedding fragments

    Medline is recalling BAL and ENDO kits because they contain recalled Olympus biopsy valves, where rubber fragments may detach from the valve slit during use.

    Product
    Medline kits containing Olympus biopsy valves: 1. BAL KIT, DYNDA2137B 2. ENDO KIT, DYKE1915
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-1751-2026·2026-04-08

    MiniMed 530G insulin pump delivery varies with pump height changes

    Medtronic is recalling MiniMed 530G insulin pumps because raising or lowering the pump relative to the infusion site can cause unintended over- or under-delivery of insulin.

    Product
    MiniMed 530G Insulin Pump (MMT-551, MMT-751)
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-1704-2026·2026-04-08

    Finesse BTK angioplasty balloon catheter may not meet burst specifications

    The Finesse BTK Multicath injectable angioplasty balloon catheter is being recalled because of the potential for the balloon not to meet burst specifications.

    Product
    Brand Name: Finesse BTK Multicath Product Name: Injectable Angioplasty Balloon Catheter Model/Catalog Number: FS252251502 Product Description: 5 French sheath, 0.014" guidewire compatible over-the-wire injectable angioplasty balloon catheter, balloon diameter 2.5mm, balloon le
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-1669-2026·2026-04-08

    Philips Allura Xper FD20/20 X-ray may not start from foot switch

    Philips is recalling the Allura Xper FD20/20 because X-ray imaging may not be initiated, or may occur intermittently, when using the wired foot switch.

    Product
    Allura Xper FD20/20; System Code: 722038;
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-1657-2026·2026-04-08

    Atellica hemoglobin A1c assay may return depressed patient results

    Siemens is recalling Atellica CH Enzymatic Hemoglobin A1c because results may be depressed when the RCRP reagent is processed on the same analyzer, with bias increasing over time.

    Product
    Atellica CH Enzymatic Hemoglobin A1c (A1c_E); Siemens Material Number (SMN): 11097536;
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-1694-2026·2026-04-08

    Olympus SOLTIVE Premium laser power supply may fail producing smoke

    Olympus is recalling SOLTIVE Premium SuperPulsed Laser Systems because a potential defect in the 24V power supply module may render the system inoperable and produce smoke or a burning smell.

    Product
    Brand Name: Olympus SOLTIVE Premium SuperPulsed Laser System Product Name: SOLTIVE Premium Laser System TFL-PLS Model/Catalog Number: TFL-PLS containing TFL-CPLU Product Description: An electricity powered device assembly which emits a high-power laser beam to vaporize/ablate
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-1684-2026·2026-04-08

    CPX 4 tall height tissue expander infusion needles may be blunt

    Mentor is recalling CPX 4 US Breast Tissue Expanders with Suture Tabs in the smooth tall height range because the infusion sets may contain a dull or blunt needle tip.

    Product
    CPX 4 US Breast Tissue Expanders with Suture Tabs Reference Numbers: SCPX-113TH, SCPX-123TH, SCPX-140TH, SCPX-146TH Smooth Tall Height Tissue Expander, 350cc, 450cc, 650cc, 750cc
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-1695-2026·2026-04-08

    Olympus SOLTIVE Pro laser power supply may fail producing smoke

    Olympus is recalling SOLTIVE Pro SuperPulsed Laser Systems because a potential defect in the 24V power supply module may render the system inoperable and produce smoke or a burning smell.

    Product
    Brand Name: Olympus SOLTIVE Pro SuperPulsed Laser System Product Name: SOLTIVE Pro Laser System TFL-SLS Model/Catalog Number: TFL-SLS containing TFL-CSLU Product Description: An electricity powered device assembly which emits a high-power laser beam to vaporize/ablate soft tis
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-1655-2026·2026-04-08

    Paradise renal denervation catheter distributed despite being marked for scrapping

    Recor Medical is recalling one lot of the Paradise Ultrasound Renal Denervation System Catheter because a nonconforming product that should have been scrapped was distributed instead.

    Product
    Brand Name: Paradise¿ Ultrasound Renal Denervation System Product Name: Paradise ¿ Catheter Model/Catalog Number: PRDS-068-02 Product Description: Common name: Ablation catheter renal denervation
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-1748-2026·2026-04-08

    Paradigm insulin pump delivery varies with pump height changes

    Medtronic is recalling Paradigm insulin pumps models MMT-712 and MMT-715 because raising or lowering the pump relative to the infusion site can cause unintended over- or under-delivery of insulin.

    Product
    Paradigm Insulin Pump (MMT-712, MMT-715)
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-1747-2026·2026-04-08

    MiniMed 620G insulin pump delivery varies with pump height changes

    Medtronic is recalling MiniMed 620G insulin pumps because raising or lowering the pump relative to the infusion site can cause unintended over- or under-delivery of insulin.

    Product
    MiniMed 620G Insulin Pump (MMT-1750)
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-1676-2026·2026-04-08

    Philips Azurion 7 B20 X-ray may not start from foot switch

    Philips is recalling the Azurion 7 B20 because X-ray imaging may not be initiated, or may occur intermittently, when using the wired foot switch.

    Product
    Azurion 7 B20; System Code: (1)722068, (2)722226, (3)722236;
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-1732-2026·2026-04-08

    Civco eTRAX 14G starter kit may misreport needle tip position

    Civco is recalling the eTRAX Needle System Starter Kit 14G for Aurora trackers because an inspection and programming error may cause the needle tip position to be incorrectly identified.

    Product
    eTRAX Needle System Starter Kit 14G (for Aurora Trackers), Part Number 667-150
    Category
    Medical Device
    Distribution
    3 states
  • HighFDA (Devices)·Z-1756-2026·2026-04-08

    Burlington Medical frontal aprons may lose radiation attenuation over time

    Burlington Medical is recalling frontal aprons made with Xenolite 800 attenuation material because of the potential for attenuation degradation over time.

    Product
    Burlington Medical, Frontal Aprons
    Category
    Medical Device
    Distribution
    Distributed nationwide