Searchable archive of product recalls from the FDA, USDA FSIS, CPSC, NHTSA, and EPA — translated into plain English, severity-scored from objective criteria, and always linked back to the original government notice.

How to read this listing

Every recall on this list has been scored 1–5 against an objective rubric from the agency’s own classification, presence of reported injuries or deaths, distribution scope, and hazard type. The colored severity badge on each card is the score — 5 (Critical) means the agency’s own highest classification or reported deaths or serious injuries, 4 (Severe) means significant injury or hospitalization risk, and 3 (High) means a known hazard with no confirmed harm yet. Filter to 3+ (High and above) if you want to skip routine Class-III notices.

The agency chips above control which streams are visible. FDA spans food, drugs, and medical devices. USDA FSIS is meat, poultry, and processed egg products. CPSC is household consumer products. NHTSA is vehicles, tires, and child seats. Nationwide on a card means the recall affects every US state — most do — while a state list means distribution was limited.

Click any recall to see the full summary, affected lot codes and model numbers, the agency’s own classification, and a direct link back to the original government notice for verification.

Filters

Agency

Category

Minimum severity

Clear filters

20451–20475 of 27647

  • HighFDA (Devices)·Z-0679-2017·2016-12-07

    Teleflex Recalls Arrow OnControl Aspiration System Trays Due to Sterility Concerns

    Teleflex Medical is recalling specific lots of Arrow OnControl Ported Aspiration System Trays because the outer sterile package may have an incomplete seal.

    Product
    ARROW OnControl Ported Aspiration System Tray, Sterile, Rx only, The Arrow OnControl Bone Marrow Aspiration System is intended for bone marrow aspiration of the iliac crest of adult and pediatric patients. The Arrow OnControl Bone Marrow Biopsy System is intended for bone marrow…
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • SevereFDA (Devices)·Z-0653-2017·2016-11-30

    AtriCure Recalls COBRA Fusion Ablation Systems Due to Component Separation Risk

    AtriCure is recalling COBRA Fusion Ablation Systems because a part may separate and remain in the body, potentially causing permanent impairment or damage.

    Product
    COBRA Fusion 50 Ablation System, COBRA Fusion 150 Ablation System, and COBRA Fusion Magnetic Retriever System. Model Numbers: 700-001, 700-001S, 700-001MI, 700-002, and 700-003. Intended to ablate cardiac tissue during cardiac surgery using radiofrequency (RF) energy when…
    Category
    Medical Device
    Distribution
    37 states
  • HighFDA (Devices)·Z-0635-2017·2016-11-30

    Merit Medical Procedure Kits Recalled for Compromised Sterile Barrier

    Merit Medical Systems is recalling 53 Procedure Kits due to a compromised sterile barrier. The kits were distributed nationwide to specific states and Japan.

    Product
    Procedure Kit. Catalog Numbers K12-04830B and K12-YP030A
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-0655-2017·2016-11-30

    Hologic Recalls SURESOUND22 Uterine Sounds Due to Sterile Barrier Breach

    Hologic is recalling SURESOUND22 uterine sound multipacks because a breach in the sterile barrier compromises product sterility.

    Product
    SURESOUND22US Multipack (US Distribution) containing 2 packs of six devices. SURESOUND22 Multipack (OUS Distribution) containing 2 packs of six devices. For uterine sound
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-0632-2017·2016-11-30

    Merit Medical Recalls Manifold Kits Due to Compromised Sterile Barrier

    Merit Medical Systems is recalling 40 Manifold Kits (Catalog No. K09-11867AP) because the sterile barrier is compromised, posing a risk of infection.

    Product
    Manifold Kit. Catalog Number K09-11867AP
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-0634-2017·2016-11-30

    Merit Medical Recalls Pressure Monitoring Kits Due to Sterile Barrier Compromise

    Merit Medical Systems is voluntarily recalling 130 MeriTrans Pressure Monitoring Kits because a compromised sterile barrier may affect product sterility.

    Product
    Pressure Monitoring Kit (MeriTrans"). Catalog Number K11-00232G
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-0650-2017·2016-11-30

    DePuy ACTIS Flex Reamers Recalled Due to Breakage Risk

    DePuy Orthopaedics is voluntarily recalling ACTIS Flex Reamers because the instruments may break and leave pieces in patients.

    Product
    ACTIS Flex Reamers SZ 0/1; 2/3; 4/5; 6/7; 8/9; 10/11; 12, Product is sold non-sterile, these instruments are packed non-sterile, within a rigid tube with foam protectors, with label and IFU
    Category
    Medical Device
    Distribution
    11 states
  • HighFDA (Devices)·Z-0622-2017·2016-11-30

    Toshiba Aquilion Prime CT System Recalled for Cooling Oil Leak Risk

    Toshiba American Medical Systems is recalling Aquilion Prime CT systems because cooling oil in the x-ray tube heat exchanger may leak during operation.

    Product
    Aquilion Prime TSX-303A Aquilion PRIME is a multislice helical CT system with an 40-row detector. High-speed rotation allows rapid data acquisition and shortens scan times, while the fast reconstruction unit further improves throughput, reducing the time required for diagnosis.
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-0375-2017·2016-11-30

    Philips CT Scanner Recall Due to Tabletop Movement Failure

    Philips is recalling 1,363 CT scanners because the patient support tabletop may fail to move horizontally during scans due to a mechanical linkage failure.

    Product
    Philips Healthcare System X-ray Tomography Computed :
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-0633-2017·2016-11-30

    Merit Medical Procedure Kits Recalled for Compromised Sterile Barrier

    Merit Medical Systems is voluntarily recalling specific Procedure Kits due to a compromised sterile barrier, which may affect the sterility of the contents.

    Product
    Waste Management Kit. Catalog Number K10-04381AP
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-0652-2017·2016-11-30

    Siemens Dimension Vista Chemistry 1 Calibrator Recall for Low Calcium Recovery

    Siemens is recalling Dimension Vista Chemistry 1 Calibrators (lots 5GM081 and 5GM082) because they may cause discrepant low calcium recovery in quality control and patient samples.

    Product
    Dimension Vista Chemistry 1 Calibrator The Chemistry 1 Calibrator (KC110B) is an in vitro diagnostic product for the calibration of Calcium (CA), Cholesterol (CHOL), Creatinine (CREA/CRE2), Glucose (GLU), Lactic Acid (LA), Magnesium (MG), Thyroxine (T4), Thyroxine Uptake (TU)…
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-0651-2017·2016-11-30

    Bausch + Lomb Recalls PeroxiClear Peroxide Solution for Excess Residuals

    Bausch + Lomb is recalling PeroxiClear Peroxide Solution because the product was found to contain excess residual levels of hydrogen peroxide.

    Product
    PeroxiClear Peroxide Solution, (3oz. and 12 oz., 12 oz. Twin Pack, 2x12 oz. Twin Pack for Target stores, 12 oz. for Walmart Vision Center and 3 oz. Starter Kit US), Manufactured by Bausch + Lomb, Rochester, NY 14609. PeroxiClear 3% Hydrogen Peroxide Cleaning & Disinfecting…
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-0645-2017·2016-11-30

    DiaSorin Recalls LIAISON Estradiol II Gen Assay Due to Cross-Reactivity

    DiaSorin is recalling the LIAISON Estradiol II Gen assay because it may produce inaccurate results for patients treated with fulvestrant.

    Product
    LIAISON Estradiol II Gen, 310680, 310680-CN The DiaSorin LIAISON Estradiol II Gen assay is a chemiluminescent immunoassay (CLIA) intended for the quantitative determination of estradiol in human serum. Assay results should be used in conjunction with other clinical and…
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-0376-2017·2016-11-30

    Philips CT Scanner Recall Due to Potential Tabletop Movement Failure

    Philips is recalling 1,363 CT scanners because the patient support tabletop may fail to move horizontally during scans due to a mechanical linkage failure.

    Product
    Philips Healthcare System X-ray Tomography Computed :
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-0623-2017·2016-11-30

    Greatbatch Stiffer Coaxial Micro-Introducer Recalled for Detaching Luer

    Greatbatch Medical is recalling Stiffer Coaxial Micro-Introducers because a rotating luer detached from the dilator hub during use.

    Product
    Greatbatch Stiffer Coaxial Micro-Introducer. Model #/ Name: 10904-001 Stiffer Coaxial Micro-Introducer 4FR; 10904-002 Stiffer Coaxial Micro-Introducer 5FR. The radiopaque Stiffer Coaxial Micro-Introducer (Figure 1) consists of a dilator and sheath, available in 4FR and 5FR.…
    Category
    Medical Device
    Distribution
    2 states
  • HighFDA (Devices)·Z-0614-2017·2016-11-30

    Merge Healthcare Recalls RadSuite Software Due to Image Display Malfunction

    Merge Healthcare is recalling RadSuite software versions 8.30.7.8 through 8.31 because images may not appear readable, potentially delaying care or causing patient re-exposure to radiation.

    Product
    Merge RadSuite,f/k/a: Emageon Advanced Visualization , RadSuite versions 8.30.7.8, 8.30.7.9, 8.30.8, and 8.31. Advanced Visualization (Image Viewing) includes: Full featured 2D imaging, 3D surface and volume rendering, Real-time Multi-Planar Reformatting (MPR), Real-time…
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • SevereFDA (Devices)·Z-0487-2017·2016-11-23

    ICU Medical Recalls SafeSet Reservoir Kits Due to Potential Leak

    ICU Medical is recalling 205 SafeSet Reservoir and blood sampling kits containing Transpac IV transducers due to a potential for leak.

    Product
    Cath Lab Kit For Elkhart Hosp., Item No. 46078-17 The Transpac Disposable Straight Pressure Transducer (DSPT) is an extravascular pressure transducer that interfaces between a catheter and pressure monitor by converting changes in pressure into electrical currents that can be…
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-0508-2017·2016-11-23

    ICU Medical Recalls Pediatric CVL Monitoring Kits Due to Leak Risk

    ICU Medical is recalling 40 Pediatric CVL Monitoring Kits containing Transpac IV transducers due to a potential for leak.

    Product
    PEDIATRIC CVL MONITORING KIT W/03ML FLUSH DEVICE FOR YALE NEW HAVEN HOSP., Item No. 46094-06 The Transpac Disposable Straight Pressure Transducer (DSPT) is an extravascular pressure transducer that interfaces between a catheter and pressure monitor by converting changes in…
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-0530-2017·2016-11-23

    ICU Medical Recalls SafeSet Reservoir and Blood Sampling Kits

    ICU Medical is recalling specific SafeSet Reservoir and blood sampling kits containing Transpac IV transducers due to a potential for leak.

    Product
    CATH LAB KIT w/3 PORT "OFF" MANIFOLD, WASTE BAG, GUIDE WIRE, SPLASH STOP, NEEDLE & TP-4, Item No. 46098The Transpac Disposable Straight Pressure Transducer (DSPT) is an extravascular pressure transducer that interfaces between a catheter and pressure monitor by converting…
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-0602-2017·2016-11-23

    Siemens Artis zee/zeego Angiographic X-ray Systems Recalled for Footswitch Defect

    Siemens is recalling Artis zee/zeego angiographic x-ray systems because the semi-automatic focus switch may fail when the footswitch is pressed multiple times.

    Product
    Artis zee/zeego, Angiographic x-ray system Product Usage: The Artis systems are a family of dedicated angiography systems developed for single and biplane diagnostic imaging and interventional procedures including, but not limited to, pediatric and obese patients. Procedures…
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-0489-2017·2016-11-23

    ICU Medical Recalls Transpac IV Transducer Kits Due to Leak Risk

    ICU Medical is recalling 160 units of Transpac IV transducer kits due to a potential leak. The affected products were distributed nationwide.

    Product
    TRIPLE KIT W/SAFESET" AND 03ML FLUSH DEVICE FOR O.H.S.U., Item No. 46080-38 The Transpac Disposable Straight Pressure Transducer (DSPT) is an extravascular pressure transducer that interfaces between a catheter and pressure monitor by converting changes in pressure into…
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-0421-2017·2016-11-23

    ICU Medical Recalls Transpac IV Monitoring Kits Due to Leak Risk

    ICU Medical is recalling Transpac IV bifurcated monitoring kits because of a potential for leak in the SafeSet Reservoir and blood sampling components.

    Product
    TRANSPAC¿ IV BIFURCATED MONITORING KIT, 72" 2 DISPOSABLE TRANSDUCERS, with 2 STOPCOCKS, 2 3 ml SQUEEZE FLUSH, MACRODRIP (POLE MOUNT), Item No. 42500-19 The Transpac Disposable Straight Pressure Transducer (DSPT) is an extravascular pressure transducer that interfaces between a…
    Category
    Medical Device
    Distribution
    Distributed nationwide