Philips Allura Xper Systems Recalled for Chiller Leak Fire Risk
Philips is recalling specific Allura Xper systems equipped with Laird chillers because a leak in the cooling system could cause liquid to drip onto electrical components, potentially leading to fire.
- Product
- The following systems are only affected if a Laird chiller is installed (11NC: 98960021599): Allura Xper Releases 3 to 8.1: Allura Xper FD10 (C+F). 722003; Allura Xper FD10/10 722005; Allura Xper FD20 722006; Allura Xper FD20/10 and FD20/20 722008; Allura Xper FD10 722010…
- Category
- Medical Device
- Distribution
- Distributed nationwide