The Recall Desk

Recalls

All US Product Recalls

Searchable archive of product recalls from the FDA, USDA FSIS, CPSC, NHTSA, and EPA — translated into plain English, severity-scored from objective criteria, and always linked back to the original government notice.

How to read this listing

Every recall on this list has been scored 1–5 against an objective rubric from the agency’s own classification, presence of reported injuries or deaths, distribution scope, and hazard type. The colored severity badge on each card is the score — 5 (Critical) means the agency’s own highest classification or reported deaths or serious injuries, 4 (Severe) means significant injury or hospitalization risk, and 3 (High) means a known hazard with no confirmed harm yet. Filter to 3+ (High and above) if you want to skip routine Class-III notices.

The agency chips above control which streams are visible. FDA spans food, drugs, and medical devices. USDA FSIS is meat, poultry, and processed egg products. CPSC is household consumer products. NHTSA is vehicles, tires, and child seats. Nationwide on a card means the recall affects every US state — most do — while a state list means distribution was limited.

Click any recall to see the full summary, affected lot codes and model numbers, the agency’s own classification, and a direct link back to the original government notice for verification.

Filters

Agency

Category

Minimum severity

Clear filters

17476–17500 of 40248

  • HighFDA (Devices)·Z-1984-2020·2020-05-20

    Dover Silicone Foley Catheter Recalled for Temperature Sensor Inaccuracy

    FDA recalls 468 Dover Foley catheters with temperature sensors that failed accuracy tests. Incorrect readings could lead to misdiagnosis.

    Product
    Dover 100% Silicone Foley Catheter with Temperature Sensor 14FR 5CC2W, Product Code: 50514T - Product Usage: for Drainage of the Urinary bladder and simultaneous monitoring of temperature. Prescription use only.
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-1987-2020·2020-05-20

    Dover Foley Catheter Recalled for Temperature Sensor Calibration Failures

    Covidien is recalling Dover Foley catheters with temperature sensors that failed to meet calibration tolerances, posing a risk of incorrect diagnosis.

    Product
    Dover 100% Silicone Silver-Hydrogel Coated Foley Catheter with Temperature Sensor 18FR 5CCCC2W, Product Code: 50518ICT - Product Usage: for Drainage of the Urinary bladder and simultaneous monitoring of temperature. Prescription use only.
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-1990-2020·2020-05-20

    Dover Silicone Foley Catheter Recalled for Temperature Sensor Inaccuracy

    FDA is recalling Dover silicone Foley catheters because temperature sensors failed to meet calibration standards, risking incorrect patient diagnoses.

    Product
    Dover 400mL, 100% Silicone Premium Urine Meter Foley Tray, 16 Fr/Ch (5.3 mm), 5 mL Catheter Pre-connected to 2000 mL Drainage Bag with Luer-Lock Sampling, Temperature Sensor and Adhesive Catheter Securement, Sterile, Product Code: P4P16TSD - Product Usage: for Drainage of the…
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-1954-2020·2020-05-20

    Cardiac Assist Recalls TandemLung Kits Due to Assembly Error

    Cardiac Assist, Inc. is recalling two TandemLung Kit units because an assembly error may prevent the pump from priming.

    Product
    TandemHeart pump is assembled into kits: TandemLung Kit - DL31, Product: 5730-3118 - Product Usage: provides temporary extracorporeal circulatory system support with the unique ability to locally deliver an anticoagulant through the infusion system.
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-1952-2020·2020-05-20

    Cardiac Assist Recalls TandemHeart ProtekDuo Kits Due to Assembly Error

    Cardiac Assist, Inc. is recalling 46 TandemHeart ProtekDuo Kits due to an assembly error that may cause failure to prime.

    Product
    TandemHeart pump is assembled into kits: ProtekDuo Kit - DL3, Product: 5720-3118 - Product Usage: provides temporary extracorporeal circulatory system support with the unique ability to locally deliver an anticoagulant through the infusion system.
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-1979-2020·2020-05-20

    Access Scientific Recalls Saline Flush Syringes Due to Packaging Holes

    Access Scientific LLC is recalling 570 units of BD PosiFlush SF Saline Flush Syringes used in POWERWAND XL catheters because holes in the packaging compromised sterility.

    Product
    BD PosiFlush SF Saline Flush Syringe 10 mL used in POWERWAND XL 4Fr 10cm Extended Dwell Catheter. REF/UDI / 94108/00859821006773 - Product Usage: is used to gain access to the vascular system to sample blood and administer fluids intravenously. It may be used for power injection…
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-1983-2020·2020-05-20

    Access Scientific Recalls Saline Flush Syringes Due to Packaging Holes

    Access Scientific LLC is recalling 110 units of BD PosiFlush SF Saline Flush Syringes because holes in the packaging compromised sterility.

    Product
    BD PosiFlush SF Saline Flush Syringe 10 mL used in POWERWAND XL 5Fr 8cm Max Barrier Kit. REF/UDI / 94116/10859821006237 - Product Usage: is used to gain access to the vascular system to sample blood and administer fluids intravenously. It may be used for power injection of…
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-1951-2020·2020-05-20

    Cardiac Assist Recalls 38 TandemHeart ProtekDuo Kits Due to Assembly Error

    Cardiac Assist, Inc. is recalling 38 TandemHeart ProtekDuo Kits because an assembly error may prevent the pump from priming.

    Product
    TandemHeart pump is assembled into kits: ProtekDuo Kit - DL29, Product: 5720-2916 - Product Usage: provides temporary extracorporeal circulatory system support with the unique ability to locally deliver an anticoagulant through the infusion system.
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-1967-2020·2020-05-20

    BD Alaris PCEA Administration Set Recalled for Potential Leaking

    BD is recalling 110,350 Alaris PCEA Administration Sets because they may leak at the luer connection, potentially causing inadequate pain control.

    Product
    BD Alaris PCEA Administration Set, Microbore Tubing with Yellow Identification Stripe ASV, Catalog Number 30893-07 - Product Usage: Microbore Tubing with Yellow Identification Stripe Anti-siphon Valve Use Aseptic Technique Set cannot be used for gravity infusion. 1. Connect…
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-1956-2020·2020-05-20

    Merit Medical Recalls Custom Sheath Introducer Kits Due to Labeling Error

    Merit Medical Systems is recalling 15 Custom Sheath Introducer Kits because they were incorrectly labeled and contained the wrong sheath assembly.

    Product
    Custom Sheath Introducer Kit REF K21-00031 - Product Usage: intended use is to provide access and facilitate the percutaneous introduction of various devices into vein.
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-2007-2020·2020-05-20

    Edwards Lifesciences Recalls Sterile Oxymetry Oligon Catheter Kits

    Edwards Lifesciences is recalling 597 sterile catheter kits because their packaging was compromised, potentially affecting sterility.

    Product
    Sterile Oxymetry Oligon Catheter Kit. Pressure injectable. REF/UDI:XA3816SAFKT/ (01)00690103197181
    Category
    Medical Device
    Distribution
    15 states
  • HighFDA (Devices)·Z-1963-2020·2020-05-20

    DeRoyal Sterile Custom Kits Recalled Due to Packaging Holes

    DeRoyal Industries is recalling sterile custom kits packaged with BD PosiFlush syringes because holes in the packaging may compromise sterility.

    Product
    DeRoyal Sterile Custom Kits packaged as a Piggy Back with the BD PosiFlush SF Saline Flush Syringe, labeled as follows: a. H* Vein Pack Pgybk, 89-6209.03 b. H* Vein Pack, 89-6209.04 c. PICC Line Tray Pgybk, 89-6150.12 d. Umbilical Arterial Cath, 89-9370.01 e.…
    Category
    Medical Device
    Distribution
    8 states
  • ModerateFDA (Devices)·Z-1998-2020·2020-05-20

    ICU Medical Recalls Central Venous Catheters Due to Guidewire Issue

    ICU Medical is recalling 10 Central Venous Catheters because the included guidewire cannot pass through the needles.

    Product
    Central Venous Catheters are indicated for the assessment of hemodynamic status through right atrial, right ventricular, and pulmonary artery and/or wedge pressure monitoring for patients. Model # 41401-25.
    Category
    Medical Device
    Distribution
    19 states
  • ModerateFDA (Devices)·Z-1964-2020·2020-05-20

    Bard PowerPort ClearVUE Slim Implantable Port Recalled for Incorrect Catheter Lock Size

    Bard Peripheral Vascular Inc is recalling 200 units of PowerPort ClearVUE Slim Implantable Ports because the catheter locks may be 8Fr instead of the labeled 6Fr.

    Product
    BARD Access Systems PowerPort ClearVUE Slim Implantable Port With Smooth Septum and Attachable 6 F Polyurethane Open-Ended Single-Lumen Venous Catheter, Product #5676300 - Product Usage: All materials are biocompatible, can be used with virtually all injectable solutions…
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • ModerateFDA (Devices)·Z-1999-2020·2020-05-20

    ICU Medical Recalls Central Venous Catheters Due to Guidewire Issue

    ICU Medical is recalling 60 units of Model 41406-23 Central Venous Catheters because the included guidewires cannot pass through the needles.

    Product
    Central Venous Catheters are indicated for the assessment of hemodynamic status through right atrial, right ventricular, and pulmonary artery and/or wedge pressure monitoring for patients. Model # 41406-23
    Category
    Medical Device
    Distribution
    19 states
  • ModerateFDA (Devices)·Z-1928-2020·2020-05-20

    Edwards HemoSphere Oximetry Cable Recalled for Inconsistent Connector Crimping

    Edwards Lifesciences is recalling 35 HemoSphere oximetry cables due to intermittent communication caused by inconsistent crimping of connectors.

    Product
    Oxymetry Cable, Rxonly. REF/UDI: HEMOXSC100/(01)00690103196986 - Product Usage: The HemoSphere oximetry cable is a multi-use device that connects with a compatible Edwards monitor on one end and any approved Edwards oximetry catheter on the other end to continuously measure…
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • ModerateFDA (Devices)·Z-1942-2020·2020-05-20

    Karl Storz Endoscopy Recalls Insufflation Tubing Due to Ethylene Oxide Residues

    Karl Storz Endoscopy is recalling sterile insufflation tubing because it may contain higher ethylene oxide residues than the acceptable level.

    Product
    Insufflation tubing to be used with varies insufflator units. REF 031200-10; REF 031322-10; REF 031532-10; REF 031622-10
    Category
    Medical Device
    Distribution
    27 states
  • ModerateFDA (Devices)·Z-1970-2020·2020-05-20

    C.R. Bard Recalls SureStep Foley Tray Systems Due to Residue

    C.R. Bard is recalling 8,392 SureStep Foley Tray Systems because a white paper-type residue may be attached to the catheter.

    Product
    SureStep Foley Tray System - Product Usage: Intended for use in the drainage and/or collection and/or measurement of urine. Generally, drainage is accomplished by inserting the catheter through the urethra and into the bladder. However, drainage is sometimes accomplished by…
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • ModerateFDA (Devices)·Z-2002-2020·2020-05-20

    ICU Medical Recalls Central Venous Catheters Due to Guidewire Issue

    ICU Medical is recalling 15 units of Central Venous Catheters because the included guidewire cannot pass through the needles.

    Product
    Central Venous Catheters are indicated for the assessment of hemodynamic status through right atrial, right ventricular, and pulmonary artery and/or wedge pressure monitoring for patients. Model # 50456-01
    Category
    Medical Device
    Distribution
    19 states
  • ModerateFDA (Devices)·Z-2000-2020·2020-05-20

    ICU Medical Recalls Central Venous Catheters Due to Guidewire Defect

    ICU Medical is recalling 10 Central Venous Catheters because the included guidewire cannot pass through the needles. The defect prevents proper use of the device.

    Product
    Central Venous Catheters are indicated for the assessment of hemodynamic status through right atrial, right ventricular, and pulmonary artery and/or wedge pressure monitoring for patients. Model # 41411-25
    Category
    Medical Device
    Distribution
    19 states
  • ModerateFDA (Devices)·Z-2001-2020·2020-05-20

    ICU Medical Recalls Central Venous Catheters Due to Guidewire Issue

    ICU Medical is recalling 235 Central Venous Catheters because the included guidewire cannot pass through the needles.

    Product
    Central Venous Catheters are indicated for the assessment of hemodynamic status through right atrial, right ventricular, and pulmonary artery and/or wedge pressure monitoring for patients. Model # 41414-23
    Category
    Medical Device
    Distribution
    19 states
  • ModerateFDA (Devices)·Z-1971-2020·2020-05-20

    BD Biosciences Recalls CD45 Reagent Due to False Positive Reactions

    Becton, Dickinson and Company is recalling 1,473 vials of Analyte specific reagent (CD45) because of false positive reactions. The product is used for in vitro diagnostic testing.

    Product
    Analyte specific reagent. REF/UDI: 641408/641408-4012-1-1 (US); 641417/641417-1157-1-01 (OUS); 641399/641399-4012-1-01 (US Research Use Only) - Product Usage: CD45 is intended for in vitro diagnostic use in the identification of cells expressing the CD45 antigen, using a BD FAC…
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • ModerateFDA (Devices)·Z-1962-2020·2020-05-20

    Gore DrySeal Flex Introducer Sheath Recalled for Incorrect Size Labeling

    W. L. Gore & Associates is recalling 141 Gore DrySeal Flex Introducer Sheaths because 16Fr devices were mislabeled as 14Fr.

    Product
    Gore DrySeal Flex Introducer Sheath, Catalog #DSF1433, 33cm 14FR, Sterile. The responsible firm on the label is W. L. Gore & Associates, Inc., Flagstaff, AZ.
    Category
    Medical Device
    Distribution
    0 states
  • ModerateFDA (Devices)·Z-1997-2020·2020-05-20

    ICU Medical Recalls Central Venous Catheters Due to Guidewire Defect

    ICU Medical is recalling 260 units of Central Venous Catheters because the included guidewire cannot pass through the needles.

    Product
    Central Venous Catheters are indicated for the assessment of hemodynamic status through right atrial, right ventricular, and pulmonary artery and/or wedge pressure monitoring for patients. Model # 41401-23.
    Category
    Medical Device
    Distribution
    19 states
  • ModerateFDA (Devices)·Z-1996-2020·2020-05-20

    ICU Medical Recalls Central Venous Catheters Due to Guidewire Issue

    ICU Medical is recalling 20 Central Venous Catheters because the included guidewire cannot pass through the needles. The product was distributed to US consignees and Canada.

    Product
    Central Venous Catheters are indicated for the assessment of hemodynamic status through right atrial, right ventricular, and pulmonary artery and/or wedge pressure monitoring for patients. LOT Number 4330832. Product not distributed in US, UDI: 1 0840619 04197 4
    Category
    Medical Device
    Distribution
    19 states