The Recall Desk

Recalls

All US Product Recalls

Searchable archive of product recalls from the FDA, USDA FSIS, CPSC, NHTSA, and EPA — translated into plain English, severity-scored from objective criteria, and always linked back to the original government notice.

How to read this listing

Every recall on this list has been scored 1–5 against an objective rubric from the agency’s own classification, presence of reported injuries or deaths, distribution scope, and hazard type. The colored severity badge on each card is the score — 5 (Critical) means the agency’s own highest classification or reported deaths or serious injuries, 4 (Severe) means significant injury or hospitalization risk, and 3 (High) means a known hazard with no confirmed harm yet. Filter to 3+ (High and above) if you want to skip routine Class-III notices.

The agency chips above control which streams are visible. FDA spans food, drugs, and medical devices. USDA FSIS is meat, poultry, and processed egg products. CPSC is household consumer products. NHTSA is vehicles, tires, and child seats. Nationwide on a card means the recall affects every US state — most do — while a state list means distribution was limited.

Click any recall to see the full summary, affected lot codes and model numbers, the agency’s own classification, and a direct link back to the original government notice for verification.

Filters

Agency

Category

Minimum severity

Clear filters

1251–1275 of 27681

  • HighFDA (Devices)·Z-1689-2026·2026-04-08

    Jiffy composite polisher cups may crumble and break apart easily

    Ultradent Products is recalling one lot of Jiffy Original Composite Polisher Cups because the cups may crumble and break apart easily, which could delay patient treatment.

    Product
    Jiffy Original Composite Polisher Cups (Coarse) 12pk, REF 7011.
    Category
    Medical Device
    Distribution
    31 states
  • HighFDA (Devices)·Z-1751-2026·2026-04-08

    MiniMed 530G insulin pump delivery varies with pump height changes

    Medtronic is recalling MiniMed 530G insulin pumps because raising or lowering the pump relative to the infusion site can cause unintended over- or under-delivery of insulin.

    Product
    MiniMed 530G Insulin Pump (MMT-551, MMT-751)
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-1741-2026·2026-04-08

    MiniMed 780G insulin pump delivery varies with pump height changes

    Medtronic is recalling MiniMed 780G insulin pumps because raising or lowering the pump relative to the infusion site can cause unintended over- or under-delivery of insulin.

    Product
    MiniMed 780G Insulin Pump (MMT-1884, MMT-1885, MMT-1886, MMT-1894, MMT-1895, MMT-1896)
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-1748-2026·2026-04-08

    Paradigm insulin pump delivery varies with pump height changes

    Medtronic is recalling Paradigm insulin pumps models MMT-712 and MMT-715 because raising or lowering the pump relative to the infusion site can cause unintended over- or under-delivery of insulin.

    Product
    Paradigm Insulin Pump (MMT-712, MMT-715)
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-1711-2026·2026-04-08

    Medline anterior hip pack contains recalled Olympus biopsy valves

    Medline is recalling the Anterior Hip Pack because it contains recalled Olympus biopsy valves, where rubber fragments may detach from the valve slit during use.

    Product
    MEDLINE ANTERIOR HIP PACK DYNJ64672B
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-1664-2026·2026-04-08

    Philips Allura Xper FD20 Biplane X-ray may not start reliably

    Philips is recalling the Allura Xper FD20 Biplane because X-ray imaging may not be initiated, or may occur intermittently, when using the wired foot switch.

    Product
    ALLURA Xper FD20 Biplane; System Code: (1) 722008, (2)722013;
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-1764-2026·2026-04-08

    Burlington Medical blockers may lose radiation attenuation over time

    Burlington Medical is recalling blockers because of the potential for attenuation degradation over time, decreasing the lifespan of the product.

    Product
    Burlington Medical, Blockers.
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-1731-2026·2026-04-08

    Civco eTRAX 12G starter kit may misreport needle tip position

    Civco is recalling the eTRAX Needle System Starter Kit 12G for Aurora trackers because an inspection and programming error may cause the needle tip position to be incorrectly identified.

    Product
    eTRAX Needle System Starter Kit 12G (for Aurora Trackers), Part Number 667-149
    Category
    Medical Device
    Distribution
    3 states
  • HighFDA (Devices)·Z-1707-2026·2026-04-08

    Medline thoracic robot kits contain recalled Olympus biopsy valves

    Medline is recalling Thoracic Robots kits, model DYNJ908777B, because they contain recalled Olympus biopsy valves where rubber fragments may detach from the valve slit.

    Product
    THORACIC ROBOTS, DYNJ908777B
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-1673-2026·2026-04-08

    Philips Azurion 3 M12 X-ray may not start from foot switch

    Philips is recalling the Azurion 3 M12 because X-ray imaging may not be initiated, or may occur intermittently, when using the wired foot switch.

    Product
    Azurion 3 M12; System Code: (1)722063, (2)722221, (3)722229;
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-1733-2026·2026-04-08

    Civco eTRAX needle starter kit may misreport needle tip position

    Civco is recalling the eTRAX Needle System Starter Kit 16G for Aurora trackers because an inspection and programming error may cause the needle tip position to be incorrectly identified.

    Product
    eTRAX Needle System Starter Kit 16G(for Aurora Trackers), Part Number 667-151
    Category
    Medical Device
    Distribution
    3 states
  • HighFDA (Devices)·Z-1676-2026·2026-04-08

    Philips Azurion 7 B20 X-ray may not start from foot switch

    Philips is recalling the Azurion 7 B20 because X-ray imaging may not be initiated, or may occur intermittently, when using the wired foot switch.

    Product
    Azurion 7 B20; System Code: (1)722068, (2)722226, (3)722236;
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • SevereFDA (Devices)·Z-1614-2026·2026-04-01

    Trilogy Evo ventilator recalled as Class I over tidal volume

    Philips Respironics is recalling the Trilogy Evo ventilator because using non-pneumatic nebulizers may cause a discrepancy between the set tidal volume and the volume the patient receives.

    Product
    Philips Respironics Trilogy Evo, Software Version 1.05.15.00. Continuous home-use ventilator device.
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • SevereFDA (Devices)·Z-1616-2026·2026-04-01

    Trilogy EV300 ventilator recalled as Class I over tidal volume

    Philips Respironics is recalling the Trilogy EV300 ventilator because using non-pneumatic nebulizers may cause a discrepancy between the set tidal volume and the volume the patient receives.

    Product
    Philips Respironics Trilogy EV300, Software Version 1.05.15.00. Continuous home-use ventilator device.
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • SevereFDA (Devices)·Z-1578-2026·2026-04-01

    DuraMax hemodialysis catheter sheath introducer may not split as intended

    Merit Medical is recalling DuraMax Chronic Hemodialysis Catheters because a design defect in the 16F dual-valved splittable sheath introducer may prevent it splitting as intended.

    Product
    DuraMax Chronic Hemodialysis Catheter, REF: H787103028015/A, H787103028025/A, H787103028031/A, H787103028035/A, H787103028041/A, H787103028045/A, H787103028051/A, H787103028055/A, H787103028061/A, H787103028075/A, H787103028085/A, H787103028095/A, H787103028165/A…
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • SevereFDA (Devices)·Z-1577-2026·2026-04-01

    Merit 16F splittable sheath introducer recalled as Class I bulk product

    Merit Medical is recalling the 16F Dual Valved Splittable Sheath Introducer in bulk non-sterile form because a design defect may prevent it splitting as intended.

    Product
    16F Dual Valved Splittable Sheath Introducer (bulk, non-sterile), REF: FCL-174-00/B
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • SevereFDA (Devices)·Z-1575-2026·2026-04-01

    CentrosFLO hemodialysis catheters recalled as Class I over sheath introducer defect

    Merit Medical is recalling CentrosFLO Hemodialysis Catheters because a design defect in the 16F dual-valved splittable sheath introducer may prevent it splitting as intended.

    Product
    CentrosFLO Hemodialysis Catheters, REF: CENFP15K/A, CENFP15K/B, CENFP17K/A, CENFP19K/A, CENFP23K/A, CENFP27K/A, CENFP31K/A, CENFT15K, CENFT15K/D, CENFT17K, CENFT17K/A, CENFT17K/D, CENFT19K, CENFT19K/A, CENFT19K/B, CENFT19K/D, CENFT23K, CENFT23K/A, CENFT23K/B, CENFT23K/D…
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • SevereFDA (Devices)·Z-1617-2026·2026-04-01

    Trilogy Evo Universal ventilator recalled as Class I over tidal volume

    Philips Respironics is recalling the Trilogy Evo Universal ventilator because using non-pneumatic nebulizers may cause a discrepancy between the set tidal volume and the volume the patient receives.

    Product
    Philips Respironics Trilogy Evo Universal, Software Version 1.05.15.00. Continuous home-use ventilator device.
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • SevereFDA (Devices)·Z-1576-2026·2026-04-01

    ProGuide chronic dialysis catheters recalled as Class I over sheath introducer defect

    Merit Medical is recalling ProGuide Chronic Dialysis Catheters because a design defect in the 16F dual-valved splittable sheath introducer may prevent it splitting as intended.

    Product
    ProGuide Chronic Dialysis Catheters, REF: DC01455550/C, DC21452419/C, DC21452419-NE5/C, DC21452823-NE5/C, DC21454035/C, DC21455550/C
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • SevereFDA (Devices)·Z-1567-2026·2026-04-01

    Erbe flexible cryoprobe 20402-410 recalled as Class I over rupture

    Erbe USA is recalling the Flexible Cryoprobe REF 20402-410 because the probes may rupture or burst during activation. The FDA has classified this recall as Class I.

    Product
    Flexible Cryoprobe (OD 1.7mm, L1.15mm) REF: 20402-410. For surgical use
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • SevereFDA (Devices)·Z-1566-2026·2026-04-01

    Erbe flexible cryoprobe 20402-401 recalled as Class I over rupture

    Erbe USA is recalling the Flexible Cryoprobe REF 20402-401 because the probes may rupture or burst during activation. The FDA has classified this recall as Class I.

    Product
    Flexible Cryoprobe (OD 1.1mm, L1.15mm) w/ oversheath (OD2.6mm, L817mm) REF: 20402-401 STERILE EO. For surgical use
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • SevereFDA (Devices)·Z-1615-2026·2026-04-01

    Trilogy Evo O2 ventilator recalled as Class I over tidal volume

    Philips Respironics is recalling the Trilogy Evo O2 ventilator because using non-pneumatic nebulizers may cause a discrepancy between the set tidal volume and the volume the patient receives.

    Product
    Philips Respironics Trilogy Evo O2, Software Version 1.05.15.00. Continuous home-use ventilator device.
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • SevereFDA (Devices)·Z-1579-2026·2026-04-01

    BioFlo DuraMax catheters recalled as Class I over sheath introducer defect

    Merit Medical is recalling BioFlo DuraMax Catheters because a design defect in the 16F dual-valved splittable sheath introducer may prevent it splitting as intended.

    Product
    BioFlo DuraMax Catheter, REF: H965103028011/A, H965103028021/A, H965103028021/EU, H965103028031/A, H965103028031/EU, H965103028041/A, H965103028041/EU, H965103028051/A, H965103028051/EU, H965103028061/A, H965103028061/EU, H965103028070/A, H965103028070/EU, H965103028080/A…
    Category
    Medical Device
    Distribution
    Distributed nationwide