The Recall Desk

Recalls

All US Product Recalls

Searchable archive of product recalls from the FDA, USDA FSIS, CPSC, NHTSA, and EPA — translated into plain English, severity-scored from objective criteria, and always linked back to the original government notice.

How to read this listing

Every recall on this list has been scored 1–5 against an objective rubric from the agency’s own classification, presence of reported injuries or deaths, distribution scope, and hazard type. The colored severity badge on each card is the score — 5 (Critical) means deaths or serious injuries are reported, 4 (Severe) means significant injury or hospitalization risk, and 3 (High) means a known hazard with no confirmed harm yet. Filter to 3+ (High and above) if you want to skip routine Class-III notices.

The agency chips above control which streams are visible. FDA spans food, drugs, and medical devices. USDA FSIS is meat, poultry, and processed egg products. CPSC is household consumer products. NHTSA is vehicles, tires, and child seats. Nationwide on a card means the recall affects every US state — most do — while a state list means distribution was limited.

Click any recall to see the full summary, affected lot codes and model numbers, the agency’s own classification, and a direct link back to the original government notice for verification.

Filters

Agency

Category

Minimum severity

Clear filters

6601–6625 of 13543

  • HighFDA (Devices)·Z-1545-2024·2024-04-24

    CryoValve SG Heart Valve Recalled for Staphylococcus aureus Contamination

    Artivion is recalling one CryoValve SG cryopreserved heart valve contaminated with Staphylococcus aureus, detected during organ donation screening. No illnesses have been reported.

    Product
    CryoValve SG Cryopreserved Pulmonary Human Heart Valve
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-1598-2024·2024-04-24

    Siemens Artis X-ray systems cooling malfunction may halt procedures

    Siemens Artis interventional X-ray systems may fail to cool the X-ray tube adequately, causing system shutdown and potentially requiring cancellation of clinical imaging procedures. The recall affects 583 units distributed nationwide.

    Product
    Siemens Artis Models: Interventional Fluoroscopic X-Ray Systems-Models: Artis zee floor- 10094135 Artis zee ceiling- 10094137 Artis zee multi-purpose- 10094139 Artis zee biplane- 10094141 Artis zeego- 10280959 Artis Q floor- 10848280 Artis Q ceiling- 10848281 Artis Q bipl
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-1592-2024·2024-04-24

    Fetal Membrane Rupture Test Kits Recalled for False Positive Results

    Actim PROM dipstik fetal membrane rupture test kits are recalled because personal lubricants can interfere with the test, producing false positive results.

    Product
    Actim PROM dipstik, REF 30831ETUS; qualitative immunoassay rapid test for rupture of fetal membrane
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-1530-2024·2024-04-24

    Cook Incorporated Dilators Recalled for Low Packaging Seal Strength

    Cook Incorporated is recalling 5,998 dilators worldwide because packaging may not meet seal strength specifications, potentially compromising device sterility.

    Product
    Dilator - Intended to be used for dilating puncture sites or catheter tracts for percutaneous placement of devices for vascular and non-vascular applications such as in the venous, arterial, biliary and renal systems. 1) JCD10.0-38-20 2) JCD20.0-38-20 3) JCD22.0-38-20 4) JCD6
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-1586-2024·2024-04-24

    MRI System Patient Support Table Floor Plate May Be Incorrectly Installed

    Philips Intera CV MRI systems may have an incorrectly installed patient support table floor plate. Eight units have been recalled due to this installation defect.

    Product
    Intera CV, Magnetic Resonance System.
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-1534-2024·2024-04-24

    POLYSIN Surgical Sutures Recalled for Incorrect Needle Curvature Defect

    Surgical Specialties is recalling POLYSIN (Polyglycolic Acid) surgical sutures due to incorrect needle curvature. This defect could affect proper suture placement during surgical procedures.

    Product
    POLYSIN (Polyglycolic Acid) suture, HSM9 5-0 BR PGA Part Number G503N
    Category
    Medical Device
    Distribution
    9 states
  • HighFDA (Devices)·Z-1574-2024·2024-04-24

    MRI patient support table floor plate may be incorrectly installed

    Philips Intera 1.0T MRI systems may have patient support table floor plates that are incorrectly installed. The defect poses a potential safety risk to patients. Facilities should verify installation and contact Philips service personnel for inspection.

    Product
    Intera 1.0T Power/Pulsar, Magnetic Resonance System.
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-1581-2024·2024-04-24

    Philips Intera 1.5T MRI System Patient Support Table Installation Defect

    Philips is recalling certain Intera 1.5T MRI systems because the patient support table floor plate may be incorrectly installed, which could affect patient safety.

    Product
    Intera 1.5T Omni/Stellar, Magnetic Resonance System.
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-1536-2024·2024-04-24

    Masimo Rad-G Pulse Oximeter Recall: Potential Automatic Power Loss

    Masimo Corporation is recalling 443 Rad-G pulse oximeters that may automatically power off and on, interrupting patient monitoring. Affected units are distributed worldwide.

    Product
    Masimo Rad-G, Pulse Oximeter with temperature (W/Sensor), REF:9210, Rx Only,
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-1573-2024·2024-04-24

    MRI System Patient Support Table Floor Plate Installation Defect

    Philips Intera 1.0T MRI systems may have an incorrectly installed patient support table floor plate. This defect could affect patient safety during magnetic resonance imaging procedures.

    Product
    Intera 1.0T Omni/Stellar, Magnetic Resonance System.
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-1552-2024·2024-04-24

    Philips Achieva 1.5T MRI System: Patient Support Table Floor Plate Installation Issue

    Philips is recalling 69 units of the Achieva 1.5T Initial MRI system due to potential issues with the patient support table floor plate installation, which may affect patient safety during imaging.

    Product
    Achieva 1.5T Initial, Magnetic Resonance System.
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-1577-2024·2024-04-24

    Philips Intera 1.5T MRI Patient Support Table Floor Plate Installation Defect

    Philips is recalling 46 units of the Intera 1.5T Achieva Nova MRI system due to improper installation of the patient support table floor plate. The defect may compromise patient safety during use.

    Product
    Intera 1.5T Achieva Nova, Magnetic Resonance System.
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-1565-2024·2024-04-24

    Philips Ingenia MRI System Patient Support Table Floor Plate Installation Defect

    Philips is recalling 135 Ingenia 1.5T S MRI systems worldwide because the patient support table floor plate may be incorrectly installed, creating a potential safety risk to patients and staff.

    Product
    Ingenia 1.5T S, Magnetic Resonance System.
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-1584-2024·2024-04-24

    Philips MRI System Patient Support Table Floor Plate Installation Issue

    Philips is recalling 9 Intera 3.0T Quasar Dual MRI systems (Model 781150) distributed worldwide. The patient support table floor plate may be incorrectly installed on these units.

    Product
    Intera 3.0T Quasar Dual, Magnetic Resonance System.
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-1542-2024·2024-04-24

    Smiths Medical syringe pumps recalled for molding defect in barrel clamp guide

    Smiths Medical is recalling certain Model 3010 syringe pumps due to a molding defect in the barrel clamp guide that may prevent proper syringe recognition. Affected units were manufactured or serviced between July 2016 and April 2021.

    Product
    Smiths Medical Model 3010 Syringe pumps: a. Protege Model 3010, b. Medfusion Model 3010A, c. Protege UK Model 3010E
    Category
    Medical Device
    Distribution
    0 states
  • HighFDA (Devices)·Z-1529-2024·2024-04-24

    Gastrointestinal Suture Anchor Recalled Due to Packaging Seal Defect

    Cook Incorporated is recalling the Entuit Secure Adjustable Gastrointestinal Suture Anchor Set due to packaging that may have insufficient seal strength, potentially compromising device sterility. The recalled lot is distributed worldwide.

    Product
    Entuit Secure Adjustable Gastrointestinal Suture Anchor Set - Intended for anchoring the anterior wall of the stomach to the abdominal wall prior to introduction of interventional catheters. REFERENCE PART NUMBER (RPN): GIAS-SRM-ADJ-2 ORDER NUMBER (GPN): G35562
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-1560-2024·2024-04-24

    Philips GYROSCAN T10-NT MRI Systems Patient Support Table Floor Plate Installation Issue

    Philips North America is recalling GYROSCAN T10-NT MRI systems due to potential incorrect installation of the patient support table floor plate. Affected units should be verified for proper installation.

    Product
    GYROSCAN T10-NT, Magnetic Resonance System.
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-1575-2024·2024-04-24

    Philips MRI Patient Support Table Floor Plate May Be Incorrectly Installed

    Philips is recalling 347 Intera 1.5T MRI systems worldwide because the patient support table floor plate may be incorrectly installed, potentially compromising patient support stability during use.

    Product
    Intera 1.5T, Magnetic Resonance System.
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-1562-2024·2024-04-24

    MRI Patient Support Table Floor Plate May Be Incorrectly Installed

    Philips GYROSCAN T5-NT MRI systems may have an incorrectly installed patient support floor plate. Affected units should be inspected to ensure proper installation.

    Product
    GYROSCAN T5-NT, Magnetic Resonance System.
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-1537-2024·2024-04-24

    Masimo Rad-G Pulse Oximeter May Randomly Power Off, Causing Loss of Monitoring

    Approximately 21,723 Masimo Rad-G pulse oximeters may randomly power off and on, causing temporary loss of patient monitoring. The FDA classified this as a Class II recall affecting devices distributed worldwide.

    Product
    Masimo Rad-G, Pulse Oximeter (W/Sensor), REF:9847, Rx Only,
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-1531-2024·2024-04-24

    Dilators With AQ Hydrophilic Coating recalled due to packaging seal defects

    Cook Incorporated is recalling 1,493 Dilators With AQ Hydrophilic Coating due to packaging that may not maintain proper seal strength, potentially compromising device sterility.

    Product
    Dilator With AQ Hydrophilic Coating - Intended to be used for dilating puncture sites or catheter tracts for percutaneous placement of devices for vascular and non-vascular applications such as in the venous, arterial, biliary and renal systems. REFERENCE PART NUMBER (RPN)/ORD
    Category
    Medical Device
    Distribution
    Distributed nationwide