The Recall Desk

Recalls

All US Product Recalls

Searchable archive of product recalls from the FDA, USDA FSIS, CPSC, NHTSA, and EPA — translated into plain English, severity-scored from objective criteria, and always linked back to the original government notice.

How to read this listing

Every recall on this list has been scored 1–5 against an objective rubric from the agency’s own classification, presence of reported injuries or deaths, distribution scope, and hazard type. The colored severity badge on each card is the score — 5 (Critical) means deaths or serious injuries are reported, 4 (Severe) means significant injury or hospitalization risk, and 3 (High) means a known hazard with no confirmed harm yet. Filter to 3+ (High and above) if you want to skip routine Class-III notices.

The agency chips above control which streams are visible. FDA spans food, drugs, and medical devices. USDA FSIS is meat, poultry, and processed egg products. CPSC is household consumer products. NHTSA is vehicles, tires, and child seats. Nationwide on a card means the recall affects every US state — most do — while a state list means distribution was limited.

Click any recall to see the full summary, affected lot codes and model numbers, the agency’s own classification, and a direct link back to the original government notice for verification.

Filters

Agency

Category

Minimum severity

Clear filters

4851–4875 of 27364

  • HighFDA (Devices)·Z-2452-2024·2024-08-07

    Boston Scientific Capio SLIM Device May Fail During Surgical Procedures

    Some lots of the Capio SLIM suture capturing device may fail to consistently catch sutures during surgery, potentially requiring extended procedures or additional medical intervention.

    Product
    Capio SLIM Suture Capturing Device, 1-Pack, UPN: M0068318250. Capio SLIM Suture Capturing Devices are used by physicians to facilitate the consistent placement of sutures during open surgical procedures.
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-2458-2024·2024-08-07

    BD Insyte Autoguard BC IV Catheter Recalled for Tubing Defect and Leakage

    Becton Dickinson has recalled certain BD Insyte Autoguard BC IV catheters due to potential holes in the catheter tubing that could cause leakage during insertion. The recall involves 26,400 units distributed in the United States.

    Product
    BD Insyte Autoguard BC, Shielded IV Catheter with Blood Control Technology, REF: 382533, 20 GA x 1.00 in (1.1x25 mm) 63 mL/min, Rx Only. BD Insyte Autoguard BC Shielded IV Catheter is intended to be inserted into a patient's peripheral vascular system for short term use to sample
    Category
    Medical Device
    Distribution
    1 state
  • HighFDA (Devices)·Z-2456-2024·2024-08-07

    Blood Culture Identification Panel May Produce False Positive Results

    The BIOFIRE BCID2 Panel may produce false positive Serratia marcescens results when used with certain culture media bottles, potentially leading to inappropriate changes in patient therapy.

    Product
    BIOFIRE Blood Culture Identification 2 (BCID2) Panel, REF: RFIT-ASY-0147, for FILMARRAY Systems, when used in conjunction with Culture Media Bottles: BACT/ALERT FA PLUS, REF: 410851, and BACT/ALERT PF PLUS, REF: 410853
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-2451-2024·2024-08-07

    Capio SLIM Suture Capturing Device may fail to catch sutures consistently

    Boston Scientific is recalling some lots of Capio SLIM Suture Capturing Devices because they may not consistently catch sutures during surgery, potentially requiring prolonged procedures or additional medical intervention.

    Product
    Capio SLIM Suture Capturing Device, Box of 5, Outer Package UPN: M0068318261, Inner Package UPN: M0068318260. Capio SLIM Suture Capturing Devices are used by physicians to facilitate the consistent placement of sutures during open surgical procedures.
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-2416-2024·2024-08-07

    SlideView VS200 slide viewer laser shutter malfunction recall

    The SlideView VS200 slide viewer has a broken laser fuse caused by a short circuit from a poorly positioned cable. This prevents the shutter from operating properly, leaving it stuck in an open or closed position.

    Product
    SlideView VS200 slide viewer,
    Category
    Medical Device
    Distribution
    0 states
  • SevereFDA (Devices)·Z-2409-2024·2024-07-31

    NIM Vital Nerve Monitoring Systems Recalled for False Negative Risk

    Medtronic Xomed is recalling NIM Vital Nerve Monitoring systems because they may fail to detect nerve damage or dysfunction. Approximately 7,804 units were distributed worldwide.

    Product
    NIM Vital Nerve Monitoring systems (all serial numbers manufactured) with NIM Vital System versions: v1.0.0, v1.1.1, v1.3.2, v1.4.3. NIM Vital Nerve Integrity Monitor with below product descriptions. 1. CONSOLE NIM4CM01 NIM 4.0, Model Number: NIM4CM01. 2. CONSOLE NIM4CM
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-2414-2024·2024-07-31

    Dxl 9000 Immunoassay Analyzer Recalled for Corrosion of USB Cable Contacts

    Beckman Coulter is recalling 143 Dxl 9000 Immunoassay Analyzers due to nonconforming USB cables with brass instead of gold contacts. Brass may corrode, potentially disabling internal communication and delaying patient test results.

    Product
    Dxl 9000 Access Immunoassay Analyzer, Catalog #C11137.
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-2438-2024·2024-07-31

    Sensis Vibe diagnostic systems software crash during event documentation

    Siemens Sensis Vibe diagnostic systems may crash when using documentation features during once-per-study event reporting. Three systems are affected; no injuries reported.

    Product
    Sensis Vibe systems with software version VD15B in combination with Sensis High End Server -Intended to be used as a diagnostic and administrative tool supporting hemodynamic catheterizations and/or electrophysiology studies, for cardiac as well as interventional Radiology as wel
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-2436-2024·2024-07-31

    Perfusor Space infusion pump syringe incompatibility may cause dosing error

    B Braun is recalling Perfusor Space infusion pump kits because Cardinal Health Monoject syringes have manufacturing changes that make them incompatible, potentially causing overdose, underdose, or delayed therapy.

    Product
    Brand Name: Perfusor Space Product Name: PERFUSOR SPACE ACM-KIT Model/Catalog Number: 8713040U Product Description: Perfusor Space, Infusion Pump Kit, 1 each Component: N/A
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-2426-2024·2024-07-31

    ABL800 FLEX Blood Gas Analyzers May Report Inaccurate pH Results

    Radiometer Medical ABL800 FLEX blood gas analyzers (models 393-800, 393-801) may report inaccurate pH results due to bacterial growth in calibration solution bottles. Approximately 27,127 units distributed worldwide are affected.

    Product
    ABL800 FLEX Model Numbers 393-800 and 393-801 The ABL800 FLEX analyzers are intended for: " In Vitro testing of samples of whole blood for the parameters pH, pO2, pCO2, cK+, cNa+, cCa2+, cCl , cGlu, cLac, cCrea, ctBil, and co-oximetry parameters (ctHb, sO2, and the hemoglob
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-2437-2024·2024-07-31

    Perfusor Space Syringe Pump Recalled for Monoject Syringe Incompatibility Risk

    B Braun's Perfusor Space infusion pump may be incompatible with Cardinal Health Monoject syringes affected by manufacturing changes, creating risks of overdose, underdose, and delayed therapy.

    Product
    Brand Name: Perfusor Space Product Name: Perfusor Space Syringe Pump, Non-Wireless Model/Catalog Number: 8713032U Product Description: Perfusor Space Syringe Pump, Non-Wireless, Infusion Pump, 1 each Component: N/A
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-2424-2024·2024-07-31

    BD Nano 2nd Gen Pen Needles recalled for post-expiration distribution

    Embecta Corp. is recalling BD Nano 2nd Gen Pen Needles that were distributed at or after their February 29, 2024 expiration date. Using expired needles cannot guarantee sterility and integrity.

    Product
    BD Nano 2nd Gen Pen Needles 32G X 4mm (5-count sample contained in Patient Counseling Sample Toolbox -Catalog 6484755/Lot Number: 3750144)-Sterile, single use devices designed for the parenteral administration of a drug from a cartridge contained in a drug pen injector Model/C
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-2430-2024·2024-07-31

    Abbott Infinity 5 Pulse Generator May Fail Prematurely

    Abbott is recalling 5,900 Infinity 5 implantable pulse generators because the replacement indicator and end-of-service date may be shorter than labeled, potentially causing premature device failure. Patients should contact their doctor about device status.

    Product
    Abbott Infinity 5, Implantable Pulse Generator, REF: 6661, SterileEO, Rx Only
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-2428-2024·2024-07-31

    Beckman Coulter Clinical Reagent Recalled Due to Incorrect Lipemia Check Settings

    Beckman Coulter has recalled 13,565 units of IGM clinical reagent due to incorrect lipemia check settings that may produce false low results or report high results as normal. Affected units were distributed worldwide.

    Product
    Beckman Coulter, REF: OSR61173, IGM 4x14 mL R1, 4x11 mL R2
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-2432-2024·2024-07-31

    Abbott Infinity 7 Implantable Pulse Generators Recalled for Potentially Shortened Service Life

    Abbott is recalling 1198 Infinity 7 Implantable Pulse Generators due to potentially incorrect service life indicators and end-of-service dates in product labeling, which could result in loss of therapy.

    Product
    Abbott Infinity 7, Implantable Pulse Generator, REF: 6663, SterileEO, Rx Only
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-2434-2024·2024-07-31

    Perfusor Space Syringe Pump Recalled Due to Monoject Syringe Incompatibility

    B Braun Medical's Perfusor Space Syringe Pump may not properly recognize Cardinal Health Monoject syringes due to manufacturing changes, risking medication overdose, underdose, or therapy delays.

    Product
    Brand Name: Perfusor Space Product Name: Perfusor Space Syringe Pump, Wireless Model/Catalog Number: 8713031U Software Version: 8713031U Product Description: Perfusor Space Syringe Pump, Infusion Pump, 1 each Component: N/A
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-2441-2024·2024-07-31

    Sensis Vibe medical diagnostic system software crashes during documentation events

    A software issue in certain Siemens Sensis Vibe diagnostic systems may cause the application to crash when documentation features are used during clinical reporting events.

    Product
    Sensis Vibe systems with software version VD15B-Intended to be used as a diagnostic and administrative tool supporting hemodynamic catheterizations and/or electrophysiology studies, for cardiac as well as interventional Radiology as well as surgical studies. Equipped by modules,
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-2417-2024·2024-07-31

    Call station devices may fail to transmit emergency alarms to staff

    Securitas Healthcare's Arial Call Station (models CSK200 and CSK200MR) may fail to transmit alarms and experience premature battery depletion. The FDA recall affects 1,268 units distributed in the U.S. and Canada.

    Product
    Arial Call Station, Model Numbers CSK200 and CSK200MR, is a wireless, battery-operated call device installed inside resident living areas to allow immediate staff response.
    Category
    Medical Device
    Distribution
    9 states
  • HighFDA (Devices)·Z-2429-2024·2024-07-31

    Abbott Infinity 5 Pulse Generator May Have Inaccurate Battery-Life Labels

    Abbott Infinity 5 implantable pulse generators have inaccurate end-of-service dates in labeling, which may result in loss of therapy. Approximately 5,900 units are affected and distributed worldwide.

    Product
    Abbott Infinity 5, Implantable Pulse Generator, REF: 6660, SterileEO, Rx Only
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-2435-2024·2024-07-31

    Perfusor Space syringe pump recalled for Cardinal Health syringe incompatibility

    B Braun is recalling Perfusor Space syringe pumps because manufacturing changes to Cardinal Health Monoject syringes may cause incompatibility, resulting in dosing errors or delayed treatment.

    Product
    Brand Name: Perfusor Space Product Name: Perfusor PCA Syringe Pump Model/Catalog Number: 8713080U Product Description: Perfusor¿ Space Syringe Pump, Infusion Pump, 1 each Component: N/A
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-2431-2024·2024-07-31

    Abbott Infinity 7 Pulse Generator May Have Shorter-Than-Labeled Service Date

    The Abbott Infinity 7 implantable pulse generator's replacement indicator and end-of-service date may be shorter than the product labeling states, potentially leading to loss of therapy for patients.

    Product
    Abbott Infinity 7, Implantable Pulse Generator, REF: 6662, SterileEO, Rx Only
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-2427-2024·2024-07-31

    Medtronic A610 Clinician Programmer pocket adaptors limited to head-only MRI

    Medtronic recalled A610 Clinician Programmer software and pocket adaptors (Models 64001, 64002) due to MRI safety restrictions. Patients with these implants are limited to head-only MRI procedures.

    Product
    A610 Clinician Programmer (CP) Software Application (app) with the following versions: Version 2.0.x (2.0.4584, 2.0.4594, 2.0.4605, 2.0.4630, and 2.0.4648), Version 3.0.x (3.0.1057, 3.0.1062, 3.0.1081, and 3.0.1098), Version 4.0.x (4.0.1052). Includes the below product descripti
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-2440-2024·2024-07-31

    Sensis Vibe Cardiac Imaging System Software May Crash During Documentation

    Siemens' Sensis Vibe cardiac imaging system software may crash when documentation functionality is used during diagnostic event recording. Two affected systems have been recalled nationwide.

    Product
    Sensis Vibe systems with software version VD15B in combination with AXIOM Sensis Post Processing Workstation-Intended to be used as a diagnostic and administrative tool supporting hemodynamic catheterizations and/or electrophysiology studies, for cardiac as well as intervention
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-2445-2024·2024-07-31

    LINK SymphoKnee Patella Sizing Template recalled due to incorrect markings

    LINK BIO CORP is recalling the LINK SymphoKnee Patella Sizing Template (Item Number 881-509/00, Lot C225066) because markings on two arms of the instrument were mixed up during manufacturing.

    Product
    LINK SymphoKnee Patella Sizing Template, Item Number 881-509/00
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-2418-2024·2024-07-31

    Wireless Nurse Call Station Recalled for Potential Alarm Communication Failure

    Arial Call Station wireless nurse call systems (models CSK200-1069 and CSK200-1069MR) are recalled due to risk of alarm communication failure and premature battery depletion. 77 units affected across multiple U.S. states.

    Product
    Arial Call Station, Model Numbers CSK200-1069 and CSK200-1069MR, is a wireless nurse call station.
    Category
    Medical Device
    Distribution
    13 states