The Recall Desk

Recalls

All US Product Recalls

Searchable archive of product recalls from the FDA, USDA FSIS, CPSC, NHTSA, and EPA — translated into plain English, severity-scored from objective criteria, and always linked back to the original government notice.

How to read this listing

Every recall on this list has been scored 1–5 against an objective rubric from the agency’s own classification, presence of reported injuries or deaths, distribution scope, and hazard type. The colored severity badge on each card is the score — 5 (Critical) means deaths or serious injuries are reported, 4 (Severe) means significant injury or hospitalization risk, and 3 (High) means a known hazard with no confirmed harm yet. Filter to 3+ (High and above) if you want to skip routine Class-III notices.

The agency chips above control which streams are visible. FDA spans food, drugs, and medical devices. USDA FSIS is meat, poultry, and processed egg products. CPSC is household consumer products. NHTSA is vehicles, tires, and child seats. Nationwide on a card means the recall affects every US state — most do — while a state list means distribution was limited.

Click any recall to see the full summary, affected lot codes and model numbers, the agency’s own classification, and a direct link back to the original government notice for verification.

Filters

Agency

Category

Minimum severity

Clear filters

3951–3975 of 27346

  • HighFDA (Devices)·Z-0247-2025·2024-11-06

    Surgical patties and strips recalled for elevated endotoxin levels

    Integra LifeSciences is recalling surgical patties and strips (Model 801409) used in surgery due to higher-than-expected endotoxin levels detected in raw materials. No illnesses have been reported.

    Product
    SURG PAT XRAY 3X3 Model/Catalog Number: 801409. The surgical patties and surgical strips are indicated for the use in protection of tissue, including brain and other tissues of the central nervous system, during surgery.
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-0252-2025·2024-11-06

    Surgical Patties and Strips Recalled for Endotoxin Contamination

    Integra LifeSciences is recalling surgical patties and strips due to higher-than-expected endotoxin levels in raw material, which may result in out-of-specification endotoxin in finished products.

    Product
    SURG STRP 1X6-200 Model/Catalog Number: 801453. The surgical patties and surgical strips are indicated for the use in protection of tissue, including brain and other tissues of the central nervous system, during surgery.
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-0251-2025·2024-11-06

    Surgical Patties and Strips Recalled for Elevated Endotoxin Levels

    Integra LifeSciences is recalling surgical patties and strips (Model 801452) due to higher-than-expected endotoxin levels in raw materials that may result in out-of-specification contamination. Approximately 4,818 units are affected.

    Product
    SURG STRP 3/4X6 Model/Catalog Number: 801452. The surgical patties and surgical strips are indicated for the use in protection of tissue, including brain and other tissues of the central nervous system, during surgery.
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-0240-2025·2024-11-06

    Surgical Patties and Strips Recalled for Elevated Endotoxin Contamination

    Integra LifeSciences recalls 33,478 units of surgical patties and strips due to higher-than-expected endotoxin levels in raw materials that may have contaminated finished products.

    Product
    SURG PAT XRAY 3/4X3/4 Model/Catalog Number: 801401. The surgical patties and surgical strips are indicated for the use in protection of tissue, including brain and other tissues of the central nervous system, during surgery.
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-0265-2025·2024-11-06

    Thyroglobulin Antibody Assay Instructions Incorrectly State Biotin Interference Resolution

    Beckman Coulter is recalling over 465,000 thyroglobulin antibody test kits because their instructions incorrectly state that biotin interference was addressed. The false information could lead to false-low test results.

    Product
    Access Thyroglobulin Antibody II, REF A32898, containing Instructions for Use, IFU REF B14994, B63978, and A38168. The Access Thyroglobulin Antibody II (TgAb) assay is used for the quantitative determination of thyroglobulin antibody levels in human serum and plasma using the Ac
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-0246-2025·2024-11-06

    Surgical Patties and Strips Recalled for Elevated Endotoxin Levels

    Codman Surgical Patties and Strips are being recalled due to higher-than-expected endotoxin levels in raw materials. Approximately 93,782 units were distributed worldwide for use in protecting tissue during surgery.

    Product
    SURG PAT XRAY 1X3 Model/Catalog Number: 801408. The surgical patties and surgical strips are indicated for the use in protection of tissue, including brain and other tissues of the central nervous system, during surgery.
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-0239-2025·2024-11-06

    Surgical Patties and Strips Recalled for Out-of-Specification Endotoxin

    Integra LifeSciences is recalling Codman Surgical Patties and Strips (Model 801400) due to higher-than-expected endotoxin levels in raw materials that resulted in out-of-specification finished products.

    Product
    SURG PAT XRAY 1/2X1/2 Model/Catalog Number: 801400. The surgical patties and surgical strips are indicated for the use in protection of tissue, including brain and other tissues of the central nervous system, during surgery.
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-0249-2025·2024-11-06

    Surgical patties and strips recalled for endotoxin contamination

    Integra LifeSciences is recalling surgical patties and strips due to higher-than-expected endotoxin levels in raw materials. The contamination may have resulted in out-of-specification finished products.

    Product
    SURG STRP 1/4X6 Model/Catalog Number: 801450. The surgical patties and surgical strips are indicated for the use in protection of tissue, including brain and other tissues of the central nervous system, during surgery.
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-0250-2025·2024-11-06

    Surgical Patties and Strips Recalled for Elevated Endotoxin Levels

    Integra LifeSciences is recalling surgical patties and strips (Model 801451) worldwide due to higher-than-expected endotoxin levels in raw materials that may result in out-of-specification contamination.

    Product
    SURG STRP 1/2X6 Model/Catalog Number: 801451. The surgical patties and surgical strips are indicated for the use in protection of tissue, including brain and other tissues of the central nervous system, during surgery.
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-0261-2025·2024-11-06

    Ambu VivaSight 2 DLT endotracheal tubes recalled for hyper angulation risk

    Ambu is recalling VivaSight 2 DLT endotracheal tubes due to hyper angulation of the distal end, which increases risk of intubation complications and airway injury.

    Product
    Ambu¿ VivaSight 2 DLT Endobronchial tube, 35 FR (Catalog Number 412351000), 37 Fr (Catalog Number 412371000), 39 Fr (Catalog Number 412391000), 41 Fr (Catalog Number 412411000)
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-0278-2025·2024-11-06

    Knee Prosthesis Implant Recall: Wrong Size Component Shipped

    Smith & Nephew is recalling a knee prosthesis component because packages were found to contain the wrong size implant. The size 3 left component was shipped instead of the size 4 left as labeled.

    Product
    smith&nephew LEIGON OXINIUM CONSTRAINED NONPOROUS FEMORAL COMPONENT, SIZE 4 LEFT, REF 71421164; PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL/POLYMER
    Category
    Medical Device
    Distribution
    1 state
  • HighFDA (Devices)·Z-0233-2025·2024-11-06

    Abbott Alinity m HPV test kit control failures from iron contamination

    Abbott is recalling 13,318 units of its Alinity m HPV AMP Kit due to iron leaching into the lysis solution, causing positive controls to fail and potentially producing false negative test results.

    Product
    Abbott Alinity m HPV AMP Kit, used with the Alinity m System, product codes: a) REF 09N15-095; b) REF 09N15-090; c) REF 09N15-091
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-0253-2025·2024-11-06

    Surgical Patties and Strips Recalled for Higher-Than-Expected Endotoxin Levels

    Integra LifeSciences is recalling surgical patties and strips due to higher-than-expected endotoxin levels in raw material that may result in out-of-specification finished products. The recall affects 2,056 units distributed worldwide.

    Product
    SURG STRP 1-1/2X6 Model/Catalog Number: 801454. The surgical patties and surgical strips are indicated for the use in protection of tissue, including brain and other tissues of the central nervous system, during surgery.
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-0236-2025·2024-11-06

    Codman Surgical Patties Recalled for Endotoxin Contamination

    Integra LifeSciences Corp. is recalling Codman Surgical Patties and Strips due to endotoxin contamination in raw material that may result in out-of-spec finished products. The devices protect tissue during surgery, including the brain.

    Product
    SURG PAT XRAY 1/4X1-1/2 Model/Catalog Number: 801397. The surgical patties and surgical strips are indicated for the use in protection of tissue, including brain and other tissues of the central nervous system, during surgery.
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-0228-2025·2024-11-06

    Hospital Bed Fails to Send Remote Alerts Through Nurse Call System

    Hillrom Centrella Smart+ hospital beds may fail to send remote alerts through the Rauland Responder 5 Nurse Call System, though local bed alerts still function. This could delay nursing staff notification of patient needs.

    Product
    a. Hillrom Centrella Smart+ Medical-Surgical Hospital Bed, Product codes: P7900B000001, P7900B100010, P7900B100024, P7900B100208, P7900B100273, P7900B100304, P7900B000005, P7900B100011, P7900B100025, P7900B100212, P7900B100277, P7900B100307, P7900B000010, P7900B100017, P7900B10
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-0245-2025·2024-11-06

    Codman Surgical Patties and Strips recalled for elevated endotoxin levels

    Integra LifeSciences has recalled Codman Surgical Patties and Strips (Model 801407) due to higher-than-expected endotoxin levels in raw materials, resulting in out-of-specification endotoxin in finished products used in brain surgery.

    Product
    SURG PAT XRAY 1/2X3 Model/Catalog Number: 801407. The surgical patties and surgical strips are indicated for the use in protection of tissue, including brain and other tissues of the central nervous system, during surgery.
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-0262-2025·2024-11-06

    Dural Graft Implants Recalled for Out-of-Specification Endotoxin Levels

    Integra LifeSciences Corp. is recalling 128 units of Duraform Dural Graft Implant (Model 80-1477US, Lot CT007694) due to endotoxin levels exceeding specifications. No adverse events have been reported.

    Product
    Duraform Dural Graft Implant, Size 3x3 , Model Number 80-1477US Lot CT007694
    Category
    Medical Device
    Distribution
    0 states
  • HighFDA (Devices)·Z-0259-2025·2024-11-06

    Pregnancy Test Kits Recalled for False Negative Results

    Osom Ultra hCG pregnancy test kits are recalled due to false negative results. The devices failed to detect hCG in samples with confirmed hCG presence.

    Product
    Osom Ultra hCG Combo Test- IVD immunoassay for the qualitative detection of human chorionic gonadotropin (hCG) in serum or urine for the early confirmation of pregnancy. Model/Catalog Number: 1004
    Category
    Medical Device
    Distribution
    1 state
  • HighFDA (Devices)·Z-0238-2025·2024-11-06

    Surgical patties recalled due to higher-than-expected endotoxin levels

    Integra LifeSciences Corp. is recalling surgical patties and strips due to higher-than-expected endotoxin levels in raw material that may have resulted in out-of-specification finished products used during surgery.

    Product
    SURG PAT XRAY 1/4X1/4 Model/Catalog Number: 801399. The surgical patties and surgical strips are indicated for the use in protection of tissue, including brain and other tissues of the central nervous system, during surgery.
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-0264-2025·2024-11-06

    Olympus Electrosurgical Generator ESG-410 Recalled for Audio Buffer System Error

    Olympus Corporation recalls 279 units of its ESG-410 electrosurgical generator nationwide due to increased complaints of system error E1226 (Could not write to Audio Buffer) occurring during surgical procedures.

    Product
    Olympus Electrosurgical Generator, Model Number ESG-410 (SEP) Version WA91327U
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-0263-2025·2024-11-06

    Westmed Flex Extenders recalled for connector failure risk

    Westmed is recalling Flex Extenders because connectors may fail to hold properly, potentially causing the breathing circuit to leak or disconnect and creating life-threatening conditions. Over 50,000 units are affected across multiple U.S. states.

    Product
    1) Flex Extenders, 5.2" x 15mm OD x 15mm ID REF S6411P 2) Flex Extender, 5.2" x 15mm OD x 22/15mm ID REF 7 5132 The Westmed Flex Extenders are non invasive, single use/disposable, large bore, corrugated plastic tubes that provide a flexible, low resistance lightweight conne
    Category
    Medical Device
    Distribution
    30 states
  • HighFDA (Devices)·Z-0256-2025·2024-11-06

    Surgical Patties and Strips recalled for endotoxin contamination in raw materials

    Integra LifeSciences is recalling Surgical Patties and Strips (Model 801457) due to higher-than-expected endotoxin levels in raw materials. These surgical devices protect tissue during operations, including brain and nervous system procedures.

    Product
    SURG STRP 3-1/2X6 Model/Catalog Number:801457. The surgical patties and surgical strips are indicated for the use in protection of tissue, including brain and other tissues of the central nervous system, during surgery.
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-0282-2025·2024-11-06

    Smith & Nephew Knee Prosthesis Component Recalled for Incorrect Labeling

    Smith & Nephew recalled a knee prosthesis component after a complaint indicated the package contained the wrong size and type of implant instead of what was labeled. This product substitution could result in incorrect implant during surgery.

    Product
    smith&nephew LEGION OXINIUM POSTERIOR STABILIZED NONPOROUS FEMORAL COMPONENT, SIZE 5N, RIGHT, NARROW, REF 71421275; Uncoated knee femur prosthesis, metallic
    Category
    Medical Device
    Distribution
    7 states
  • HighFDA (Devices)·Z-0229-2025·2024-11-06

    Olympus Endoscopic Flushing Pump Recalled Due to Function Loss

    Olympus is recalling the OFP-2 Endoscopic Flushing Pump due to intermittent loss of function caused by internal component connection failure. Approximately 445 units distributed nationwide are affected.

    Product
    Olympus Endoscopic Flushing Pump, Model/Catalog Number: OFP-2 (K10001141)
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-0244-2025·2024-11-06

    Surgical patties and strips recalled for elevated endotoxin levels in raw materials

    Integra LifeSciences has recalled Codman surgical patties and strips due to higher-than-expected endotoxin levels in raw materials, which may result in out-of-specification endotoxin in finished products.

    Product
    SURG PAT XRAY 1/2X2 Model/Catalog Number: 801406. The surgical patties and surgical strips are indicated for the use in protection of tissue, including brain and other tissues of the central nervous system, during surgery.
    Category
    Medical Device
    Distribution
    Distributed nationwide