The Recall Desk

Recalls

All US Product Recalls

Searchable archive of product recalls from the FDA, USDA FSIS, CPSC, NHTSA, and EPA — translated into plain English, severity-scored from objective criteria, and always linked back to the original government notice.

How to read this listing

Every recall on this list has been scored 1–5 against an objective rubric from the agency’s own classification, presence of reported injuries or deaths, distribution scope, and hazard type. The colored severity badge on each card is the score — 5 (Critical) means deaths or serious injuries are reported, 4 (Severe) means significant injury or hospitalization risk, and 3 (High) means a known hazard with no confirmed harm yet. Filter to 3+ (High and above) if you want to skip routine Class-III notices.

The agency chips above control which streams are visible. FDA spans food, drugs, and medical devices. USDA FSIS is meat, poultry, and processed egg products. CPSC is household consumer products. NHTSA is vehicles, tires, and child seats. Nationwide on a card means the recall affects every US state — most do — while a state list means distribution was limited.

Click any recall to see the full summary, affected lot codes and model numbers, the agency’s own classification, and a direct link back to the original government notice for verification.

Filters

Agency

Category

Minimum severity

Clear filters

3476–3500 of 39948

  • ModerateFDA (Devices)·Z-2245-2025·2025-08-13

    daVinci urology kits contain an inadvertently resterilized visualization system

    DAVINCI UROLOGY medical convenience kits are being recalled because the Clearify Visualization System inside was inadvertently resterilized, causing discoloration. 166 units are affected.

    Product
    DAVINCI UROLOGY. Medical convenience kit.
    Category
    Medical Device
    Distribution
    7 states
  • ModerateFDA (Devices)·Z-2270-2025·2025-08-13

    DxI 600 immunoassay analyzer motor current resets below the required setting

    A DxI 600 Access Immunoassay Analyzer is being recalled because software overwrites the factory PnP X-motor current setting, causing increased motion errors that could delay patient results.

    Product
    DxI 600 Access Immunoassay Analyzer W/Dual Gantry, Part Number A71461
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • ModerateFDA (Devices)·Z-2243-2025·2025-08-13

    daVinci urology and prostate packs contain a resterilized visualization system

    DA VINCI PACK URO/PROSTATE medical convenience kits are being recalled because the Clearify Visualization System inside was inadvertently resterilized, causing discoloration. 21 units affected.

    Product
    DA VINCI PACK URO/PROSTATE. Medical convenience kit.
    Category
    Medical Device
    Distribution
    7 states
  • ModerateFDA (Devices)·Z-2241-2025·2025-08-13

    Cardiac robot packs contain an inadvertently resterilized visualization system

    Cardiac Robot Pack medical convenience kits are being recalled because the Clearify Visualization System inside was inadvertently resterilized, causing discoloration. 20 units are affected.

    Product
    CARDIAC ROBOT PACK. Medical convenience kit.
    Category
    Medical Device
    Distribution
    7 states
  • ModerateFDA (Devices)·Z-2257-2025·2025-08-13

    LAVH packs contain an inadvertently resterilized visualization system

    LAVH Pack medical convenience kits are being recalled because the Clearify Visualization System inside was inadvertently resterilized, causing discoloration. 292 units are affected.

    Product
    LAVH PACK. Medical convenience kit.
    Category
    Medical Device
    Distribution
    7 states
  • ModerateFDA (Devices)·Z-2267-2025·2025-08-13

    DxI 800 immunoassay analyzer motor current resets below the required setting

    A DxI 800 Access Immunoassay Analyzer is being recalled because software overwrites the factory PnP X-motor current setting, causing increased motion errors that could delay patient results.

    Product
    DxI 800 Access Immunoassay Analyzer W/Dual Gantry, Part Number A71457
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • ModerateFDA (Devices)·Z-2263-2025·2025-08-13

    Robotic uro/gyn packs contain an inadvertently resterilized visualization system

    Robotic Uro/Gyn Pack medical convenience kits are being recalled because the Clearify Visualization System inside was inadvertently resterilized, causing discoloration. 372 units are affected.

    Product
    ROBOTIC URO/GYN PACK. Medical convenience kit.
    Category
    Medical Device
    Distribution
    7 states
  • SevereFDA (Devices)·Z-2164-2025·2025-08-06

    Automated Impella Controllers may not detect a connected Impella pump

    The Automated Impella Controller, used with left heart support blood pumps, is being recalled because of a potential issue with the controller not detecting an Impella pump when it is connected. 11,031 units affected.

    Product
    Automated Impella Controller (AIC), used in left heart support blood pump, including the following: 1. Impella Controller, Packaged, AU; Product Code: 0042-0000-AU. 2. Impella Controller, Packaged, CA; Product Code: 0042-0000-CA. 3. Impella Controller, Packaged, EU; Pr
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • SevereFDA (Devices)·Z-2159-2025·2025-08-06

    SafeStar 55 Plus breathing filters show a slow rising CO2 curve

    SafeStar 55 Plus bidirectional breathing system filters are being recalled following reports of a slow increase of the CO2 curve when the gas sample is taken from the filter's port. 9,885,139 units affected.

    Product
    SafeStar 55 Plus Filter. Bidirectionally breathing system filter.
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • SevereFDA (Devices)·Z-2160-2025·2025-08-06

    HME TwinStar HEPA Plus filters show a slow rising CO2 curve

    HME TwinStar HEPA Plus bidirectional breathing system filters are being recalled following reports of a slow increase of the CO2 curve when the gas sample is taken from the filter's port. 1,019,782 units affected.

    Product
    HME TwinStar HEPA Plus Filter. Bidirectionally breathing system filter.
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • SevereFDA (Devices)·Z-2158-2025·2025-08-06

    SafeStar 90 Plus breathing filters show a slow rising CO2 curve

    SafeStar 90 Plus bidirectional breathing system filters are being recalled following reports of a slow increase of the CO2 curve when the gas sample is taken from the filter's gas sampling port. 669,838 units affected.

    Product
    SafeStar 90 Plus Filter. Bidirectionally breathing system filter.
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • SevereFDA (Devices)·Z-2161-2025·2025-08-06

    SafeStar 60A Plus breathing filters show a slow rising CO2 curve

    SafeStar 60A Plus bidirectional breathing system filters are being recalled following reports of a slow increase of the CO2 curve when the gas sample is taken from the filter's port. 374,700 units affected.

    Product
    SafeStar 60A Plus Filter. Bidirectionally breathing system filter.
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • SevereFDA (Devices)·Z-2167-2025·2025-08-06

    Hamilton coaxial breathing circuits may have a crack in blue tubing

    Coaxial breathing circuit sets for HAMILTON-C1, T1 and MR1 ventilators are being recalled because the blue tubing that delivers fresh gas may be cracked, which can cause rebreathing of exhaled gases. 2,560 units affected.

    Product
    Breathing circuit set, coaxial, 1.80 m, with flow sensor and expiratory valve set, REF: 260128, used with HAMILTON-C1, HAMILTON-T1 and HAMILTON-MR1 ventilators.
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-2212-2025·2025-08-06

    Philips MRI Machine Intera 3.0T Quasar Dual Recalled for Fire Risk

    Philips is recalling its Intera 3.0T Quasar Dual MRI system due to a potential fire hazard from gradient coil failures. The recall affects one unit distributed worldwide.

    Product
    Intera 3.0T Quasar Dual Model Number (REF): 781150;
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-2200-2025·2025-08-06

    Philips Achieva 3.0T gradient coils may act as a heat source

    Achieva 3.0T MR systems are being recalled because component failures in the gradient coil may act as a heat source with the potential to produce smoke or fire. 460 units are affected.

    Product
    Achieva 3.0T; Model Numbers (REF): (1) 781177, (2) 781277, (3) 781278, (4) 781344, (5) 781345;
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-2189-2025·2025-08-06

    Philips Allura R8.2.x systems may temporarily lose x-ray imaging functionality

    Philips Allura R8.2.x systems are being recalled because of the potential for temporary loss of imaging (X-ray) functionality due to a software issue. 4,508 units are affected.

    Product
    Philips Allura R8.2.x Systems. Software Verion: R8.2.x. 1. Allura Xper FD10, Model Number: 722026. 2. Allura Xper FD10/10, Model Number: 722027. 3. Allura Xper FD20, Model Number: 722028. 4. Allura Xper FD20/10, Model Number: 722029. 5. Allura Xper FD10 OR Table,
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-2204-2025·2025-08-06

    Philips Ingenia 1.5T gradient coils may act as a heat source

    Ingenia 1.5T MR systems are being recalled because component failures in the gradient coil may act as a heat source with the potential to produce smoke or fire. 2,279 units are affected.

    Product
    Ingenia 1.5T Model Numbers (REF): (1) 781315, (2) 718641, (3) 781396, (4) 782101, (5) 782115, (6) 782140;
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-2235-2025·2025-08-06

    LVP dual Y-site administration sets have potential for external cassette leaks

    LVP Primary Administration Sets, Dual-Inlet, Low-Sorbing, Needle-Free Port, Dual Y-Site, are being recalled because of the potential for external cassette leaks. 348,150 units are affected.

    Product
    LVP Primary Administration Set, Dual-Inlet, Low-Sorbing, Needle-Free Port, Dual Y-Site (Qty 25). Product Code: SET-0032-25.
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-2177-2025·2025-08-06

    HarmonyAIR dual flat panel monitor yokes may detach from spring arm

    Dual Flat Panel Monitor accessories for HarmonyAIR A-Series Surgical Lighting Systems are being recalled because the yoke assembly may detach from the spring arm. 167 units are affected.

    Product
    Dual Flat Panel Monitor (DFPM) accessories for HarmonyAIR A-Series Surgical Lighting Systems
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-2178-2025·2025-08-06

    HarmonyAIR E-Series monitor yokes may detach from the spring arm

    Dual Flat Panel Monitor accessories for HarmonyAIR E-Series Surgical Lighting Systems are being recalled because the yoke assembly may detach from the spring arm. 59 units are affected.

    Product
    Dual Flat Panel Monitor (DFPM) accessories for HarmonyAIR E-Series Surgical Lighting Systems
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-2213-2025·2025-08-06

    Philips MR 5300 gradient coils may act as a heat source

    MR 5300 systems are being recalled because component failures in the gradient coil may act as a heat source with the potential to produce smoke or fire. 551 units are affected.

    Product
    MR 5300 Model Numbers (REF): (1) 782110, (2) 782135 (China ONLY), (3) 782152;
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-2205-2025·2025-08-06

    Philips Ingenia 1.5T S gradient coils may act as heat source

    Ingenia 1.5T S MR systems are being recalled because component failures in the gradient coil may act as a heat source with the potential to produce smoke or fire. 136 units are affected.

    Product
    Ingenia 1.5T S Model Number (REF): 781347;
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-2218-2025·2025-08-06

    Philips MR 7700 upgrade gradient coils may act as a heat source

    Upgrade to MR 7700 systems are being recalled because component failures in the gradient coil may act as a heat source with the potential to produce smoke or fire. 26 units are affected.

    Product
    Upgrade to MR 7700 Model Number (REF): 782130;
    Category
    Medical Device
    Distribution
    Distributed nationwide