The Recall Desk

Recalls

All US Product Recalls

Searchable archive of product recalls from the FDA, USDA FSIS, CPSC, NHTSA, and EPA — translated into plain English, severity-scored from objective criteria, and always linked back to the original government notice.

How to read this listing

Every recall on this list has been scored 1–5 against an objective rubric from the agency’s own classification, presence of reported injuries or deaths, distribution scope, and hazard type. The colored severity badge on each card is the score — 5 (Critical) means the agency’s own highest classification or reported deaths or serious injuries, 4 (Severe) means significant injury or hospitalization risk, and 3 (High) means a known hazard with no confirmed harm yet. Filter to 3+ (High and above) if you want to skip routine Class-III notices.

The agency chips above control which streams are visible. FDA spans food, drugs, and medical devices. USDA FSIS is meat, poultry, and processed egg products. CPSC is household consumer products. NHTSA is vehicles, tires, and child seats. Nationwide on a card means the recall affects every US state — most do — while a state list means distribution was limited.

Click any recall to see the full summary, affected lot codes and model numbers, the agency’s own classification, and a direct link back to the original government notice for verification.

Filters

Agency

Category

Minimum severity

Clear filters

25801–25825 of 27686

  • HighFDA (Devices)·Z-0209-2014·2013-11-20

    Hospira Recalls Plum A+ Infusion Pumps Due to Pressure Sensor Calibration Drift

    Hospira is recalling Plum A+ Infusion Pumps because pressure sensor calibration drift may cause false occlusion alarms or missed occlusions.

    Product
    *** 1) Plum A+ Infusion Pump; List Number: 20678; Labeled in part: PLUM A+3; for use with Hospira MedNet; *** 2) Plum A+ Infusion Pump; List Number: 20679; Labeled in part: PLUM A+; for use with Hospira MedNet; *** 3) Plum A+ Infusion Pump with Hospira MedNet…
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-0231-2014·2013-11-20

    Carestream Vue PACS Software Recall for Image Loading Issues

    Carestream Health is recalling Vue PACS software because specific multi-slice images may fail to load in the viewer application.

    Product
    Carestream Vue PACS; MX-O SA WORKSTATION SW LICENSE The CARESTREAM PACS is an image management system whose intended use is to provide completely scaleable local and wide area PACS solutions for hospital and related institutions/sites, which will archive, distribute, retrieve…
    Category
    Medical Device
    Distribution
    26 states
  • HighFDA (Devices)·Z-0242-2014·2013-11-20

    Carestream Vue PACS Software Recall for Image Loading Issues

    Carestream Health is recalling Vue PACS software because specific multi-slice images may fail to load in the viewer application.

    Product
    Carestream Vue PACS; UPG MX-O SVR 25K EX/YR UNLIMITED The CARESTREAM PACS is an image management system whose intended use is to provide completely scaleable local and wide area PACS solutions for hospital and related institutions/sites, which will archive, distribute, retrieve…
    Category
    Medical Device
    Distribution
    26 states
  • HighFDA (Devices)·Z-0184-2014·2013-11-20

    Fresenius Hemodialysis Blood Lines Recalled for Potential Disconnection Risk

    Fresenius Medical Care is recalling 8,528 cases of 2008K@ Home Blood lines due to a risk of disconnection and blood loss if not tightened correctly.

    Product
    2008K@ Home Blood lines Combiset True Flow Series for Hemodialysis use Catalog Number: 03-2962-3. Intended for use as the extracorporeal blood circuit during Hemodialysis and is indicated for use with conventional and high flux negative pressure hemodialyzer equipment.
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-0201-2014·2013-11-20

    Toshiba Titan 3T MRI System Recalled for SAR Display Error

    Toshiba is recalling 18 Titan 3T MRI systems because they may display lower Specific Absorption Rate (SAR) values than actual when used with specific coils.

    Product
    Titan 3T MRI System, MRT-3010A/5. Product Usage: Usage: MRI system.
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-0281-2014·2013-11-20

    Ray Station Radiation Therapy System Recalled for Dose Calculation Error

    Ray Search Laboratories is recalling Ray Station 3.0, 3.5, and 4.0 systems due to a dose calculation error involving Elekta motorized wedges.

    Product
    Ray Station Radiation Therapy Treatment Planning System, Model Numbers 3.0, 3.5 and 4.0
    Category
    Medical Device
    Distribution
    12 states
  • HighFDA (Devices)·Z-0196-2014·2013-11-20

    Terumo Recalls Sarns Venous Return Cannulae Due to Particulate

    Terumo CVS is recalling Sarns Venous Return Cannulae because in-process inspections found loose fiber particulate exceeding specifications.

    Product
    Sarns" Venous Return Cannulae, 20 Fr with 1/4" flare, 14.5" (37 cm) long. Indicated for venous drainage during cardiopulmonary bypass surgery for dial cannulation of the superior and inferior vena cava.
    Category
    Medical Device
    Distribution
    9 states
  • HighFDA (Devices)·Z-0278-2014·2013-11-20

    Medtronic RestoreUltra and RestoreSensor Neurostimulators Recalled for Loss of Stimulation

    Medtronic is recalling 64,163 RestoreUltra and RestoreSensor neurostimulators worldwide due to a risk of loss of stimulation when switching program groups.

    Product
    Medtronic RestoreUltra, model 37712, and RestoreSensor, model 37714, Multi-program Rechargeable Neurostimulator for Spinal Cord Stimulation.
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-0182-2014·2013-11-20

    Urologix Recalls Prostiva RF Therapy Hand Piece Kits Due to Deployment Force Issue

    Urologix is recalling 142 Prostiva RF Therapy Hand Piece Kits because the hand piece deployment force exceeded maximum specifications.

    Product
    Prostiva RF Therapy Hand Piece Kit, REF 8929, Rx only, sterilized using Ethylene Oxide. The Prostiva RF Model 8929 Hand Piece Kit is the delivery system component of the Prostiva RF Therapy System. Prostiva RF Therapy delivers low-levels of radiofrequency (RF) energy directly…
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-0192-2014·2013-11-20

    DePuy Mitek FMS Outflow Tubing Recalled Due to Valve Malfunction

    DePuy Mitek is recalling FMS Outflow Tubing because the pillow valve may not perform as intended, potentially leading to backflow of irrigation fluid and patient cross-contamination.

    Product
    DePuy Mitek FMS Outflow Tubing with One-Way valve Product Code:284649 Product Usage: The Intermediary Tubing for FMS¿ Fluid Management Systems connects the Inflow Tubing to the arthroscopic sheath during arthroscopic procedures. The Polyvinyl Chloride (PVC) tubing…
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-0284-2014·2013-11-20

    Hospira Recalls Plum A+ Infusion Systems Due to Broken Pressure Sensor Pin

    Hospira is recalling 10,669 Plum A+ Infusion Systems because the distal pressure sensor pin may break. This defect can prevent the device from detecting occlusions in the administration set.

    Product
    The Plum A+ Infusion System is a cassette based multifunction infusion system that allows two lines in and one line out and can be used for standard, piggyback, or concurrent delivery. The Plum A+ Infusion System is designed to deliver parenteral, enteral, or epidural infusions…
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-0275-2014·2013-11-20

    DeRoyal Surgical Eye Spears Recalled for Incorrect X-Ray Detectability Labeling

    DeRoyal Industries is recalling one lot of Surgical Eye Spears because they were labeled as x-ray detectable but contained non-detectable product.

    Product
    DeRoyal (R) Surgical Eye Spears, REF 30-049, Sterile EO, Manufacturer: DeRoyal Industries, Inc. 200 DeBusk Lane, Powell, TN 37849, MADE IN GUATEMALA cleaning of fluids from surgical site in or near the eye
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-0282-2014·2013-11-20

    Draeger Ceiling Supply Units Recalled Due to Broken Drive Screw

    Draeger Medical is recalling 139 Movita 603x/DVE 803x Ceiling Supply Units because a broken drive screw caused the lift arm to fall.

    Product
    Draeger Movita 603x/DVE 803x Ceiling Supply Unit. It is intended for use in operating theaters and intensive care
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-0270-2014·2013-11-20

    GE Healthcare Recalls AW Workstations Due to Image Orientation Error

    GE Healthcare is recalling AW VolumeShare and AW Server software due to a potential safety issue where images may display in the wrong orientation after rotation.

    Product
    AW VolumeShare 4 (ADVANTAGE WORKSTATION 4.3.), AW VolumeShare 5 (AW VOLUMESHARE 5 WITH ANGIOVIZ OPTION), AW Server. K052995: Advantage Workstation 4.3 is a review station, which allows easy selection, review, processing and filming of multi-modality DICOM images from a…
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-0291-2014·2013-11-20

    Beckman Coulter Recalls Access 2 Immunoassay System Due to Fuse Defect

    Beckman Coulter is recalling the Access 2 Immunoassay System because the fuses may have the wrong amperage.

    Product
    Access 2 Immunoassay System, Part Number: 81600N The Access Immunoassay System is an in vitro diagnostic device used for the quantitative, semi-quantitative or qualitative determination of various analyte concentrations found in human body fluids.
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-0203-2014·2013-11-20

    Hospira Recalls Plum A+ Hyperbaric Infusion Pumps Due to Calibration Drift

    Hospira is recalling 179 Plum A+ Hyperbaric Infusion Pumps because distal pressure sensor calibration drift may cause false alarms or occlusions.

    Product
    Plum A+ Hyperbaric Infusion Pump; List Number 11005; Labeled in part: PLUM A+ HYPERBARIC; Product Usage: The Plum A+ Infusion Systems are a multipurpose family of devices that provides precise delivery of multiple therapies across the general spectrum of clinical…
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-0204-2014·2013-11-20

    Infusion pumps are recalled for distal pressure sensor drift

    Hospira recalled 10,669 Plum A+ infusion pumps after reports of distal pressure sensor calibration drift. The issue may cause false occlusion alarms or distal occlusions.

    Product
    Plum A+ Infusion Pump; List Number: 11971; Product Usage: The Plum A+ Infusion Systems are a multipurpose family of devices that provides precise delivery of multiple therapies across the general spectrum of clinical care
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-0207-2014·2013-11-20

    Hospira Recalls Plum A+ Infusion Pumps Due to Pressure Sensor Calibration Drift

    Hospira is recalling 182,970 Plum A+ Infusion Pumps because pressure sensor calibration drift may cause false alarms or missed occlusions.

    Product
    Plum A+ Infusion Pump; List Number: 12391; Product Usage: The Plum A+ Infusion Systems are a multipurpose family of devices that provides precise delivery of multiple therapies across the general spectrum of clinical care
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-0290-2014·2013-11-20

    Hospira Recalls Plum A+ Infusion Systems Due to Broken Sensor Pin Risk

    Hospira Inc. is recalling Plum A+ Infusion Systems because the distal pressure sensor pin may break, potentially preventing the detection of line occlusions.

    Product
    The Plum A+ Infusion System is a cassette based multifunction infusion system that allows two lines in and one line out and can be used for standard, piggyback, or concurrent delivery. The Plum A+ Infusion System is designed to deliver parenteral, enteral, or epidural infusions…
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-0289-2014·2013-11-20

    Hospira Recalls Plum A+3 Infusion Systems Due to Broken Sensor Pin Risk

    Hospira is recalling 21,551 Plum A+3 Infusion Systems because a distal pressure sensor pin may break, potentially preventing the detection of line occlusions.

    Product
    The Plum A+3 Infusion System is a cassette based multifunction infusion system that allows two lines in and one line out and can be used for standard, piggyback, or concurrent delivery. The Plum A+3 Infusion System is designed to deliver parenteral, enteral, or epidural…
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-0206-2014·2013-11-20

    Hospira Recalls Plum A+3 Infusion Pumps Due to Pressure Sensor Drift

    Hospira is recalling 214 Plum A+3 Infusion Pump Systems because distal pressure sensor calibration drift may cause false alarms or occlusions.

    Product
    Plum A+3 Infusion Pump System; List Number: 12348; Product Usage: The Plum A+ Infusion Systems are a multipurpose family of devices that provides precise delivery of multiple therapies across the general spectrum of clinical care
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-0286-2014·2013-11-20

    Hospira Recalls Plum A+3 Infusion Systems Due to Broken Pressure Sensor Pin

    Hospira is recalling 214 Plum A+3 Infusion Systems because a pressure sensor pin may break, potentially preventing the detection of line occlusions.

    Product
    The Plum A+3 Infusion System is a cassette based multifunction infusion system that allows two lines in and one line out and can be used for standard, piggyback, or concurrent delivery. The Plum A+3 Infusion System is designed to deliver parenteral, enteral, or epidural…
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-0197-2014·2013-11-20

    Baxter Recalls SIGMA SPECTRUM Infusion Pumps Due to Alarm Issues

    Baxter Healthcare is recalling SIGMA SPECTRUM Infusion Pumps because they may trigger false upstream occlusion or air-in-line alarms during use.

    Product
    SIGMA SPECTRUM Infusion Pump with Master Drug Library. Intended to be used for the controlled administration of intravenous fluids.
    Category
    Medical Device
    Distribution
    1 state
  • HighFDA (Devices)·Z-0208-2014·2013-11-20

    Hospira Recalls Plum A+ Infusion Pumps Due to Pressure Sensor Drift

    Hospira is recalling 7,691 Plum A+ Infusion Pumps because pressure sensor calibration drift may cause false alarms or undetected occlusions.

    Product
    Plum A+ Infusion Pump; List Number: 12618; Product Usage: The Plum A+ Infusion Systems are a multipurpose family of devices that provides precise delivery of multiple therapies across the general spectrum of clinical care
    Category
    Medical Device
    Distribution
    Distributed nationwide