The Recall Desk

Recalls

All US Product Recalls

Searchable archive of product recalls from the FDA, USDA FSIS, CPSC, NHTSA, and EPA — translated into plain English, severity-scored from objective criteria, and always linked back to the original government notice.

How to read this listing

Every recall on this list has been scored 1–5 against an objective rubric from the agency’s own classification, presence of reported injuries or deaths, distribution scope, and hazard type. The colored severity badge on each card is the score — 5 (Critical) means the agency’s own highest classification or reported deaths or serious injuries, 4 (Severe) means significant injury or hospitalization risk, and 3 (High) means a known hazard with no confirmed harm yet. Filter to 3+ (High and above) if you want to skip routine Class-III notices.

The agency chips above control which streams are visible. FDA spans food, drugs, and medical devices. USDA FSIS is meat, poultry, and processed egg products. CPSC is household consumer products. NHTSA is vehicles, tires, and child seats. Nationwide on a card means the recall affects every US state — most do — while a state list means distribution was limited.

Click any recall to see the full summary, affected lot codes and model numbers, the agency’s own classification, and a direct link back to the original government notice for verification.

Filters

Agency

Category

Minimum severity

Clear filters

22376–22400 of 40175

  • HighFDA (Devices)·Z-3206-2018·2018-09-26

    Coviden Temperature Sensing Foley Catheters Recalled for Inaccurate Readings

    Coviden is recalling temperature sensing Foley catheters because they may display lower body temperature readings than the patient's actual temperature.

    Product
    400mL, Premium, Urine Meter, 16 FR, Silicone Temperature Sensing Foley Traywith Cath-Secure Plus, Sterile, Item Code P4P16TSD
    Category
    Medical Device
    Distribution
    24 states
  • HighFDA (Devices)·Z-2783-2018·2018-09-26

    Degania Silicone Recalls Temperature Sensing Foley Catheters Due to Defective Sensors

    Degania Silicone is recalling 2,461 temperature sensing Foley catheters because defective sensors may display lower body temperatures than the patient's actual temperature.

    Product
    Temperature sensing catheter 18FR, Catalogue Number 102203101863DO Product Usage: This is sterile, single use, two-way silicone Foley Catheter with a thermistor embedded in the third lumen. It is intended for use in the drainage/ collection of urine from the urinary bladder…
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-2774-2018·2018-09-26

    Degania Silicone Foley Catheters Recalled for Defective Temperature Sensors

    Degania Silicone is recalling 500 units of temperature-sensing Foley catheters because defective sensors may display inaccurate, lower body temperature readings.

    Product
    Temperature Sensing 100% Silicone Foley Catheter, sterile, Catalogue Number 102201101880MD Product Usage: This is sterile, single use, two-way silicone Foley Catheter with a thermistor embedded in the third lumen. It is intended for use in the drainage/ collection of urine…
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-2772-2018·2018-09-26

    Degania Silicone Recalls Temperature Sensing Foley Catheters Due to Defective Sensors

    Degania Silicone, Ltd. is recalling 1,000 temperature sensing Foley catheters because defective sensors may display lower body temperatures than actual.

    Product
    Temperature Sensing 100% Silicone Foley Catheter, sterile, Catalogue Number 102201100880MD Product Usage: This is sterile, single use, two-way silicone Foley Catheter with a thermistor embedded in the third lumen. It is intended for use in the drainage/ collection of urine…
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-3210-2018·2018-09-26

    Medline Recalls Endoscopic Cleaning Brushes Due to Detachment Risk

    Medline Industries is recalling single-use endoscopic cleaning brushes because they may detach during cleaning, potentially leaving parts inside the scope.

    Product
    Endoscopic Cleaning Brush, single use, sold in kits and labeled as the following: a. Item Number DYK1002DCB, BRUSH, CLEANING, DOUBLE ENDED, W/ HANDLE - 300 EA/CS, 50EA/BX, 6BX/CS; b. Item Number DYK1002DBDE, BRUSH, CLEANING, DBL, DIF ENDS - 500 EA/CS, 50 EA/BX, 10…
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-2782-2018·2018-09-26

    Degania Silicone Recalls Temperature Sensing Foley Catheters Due to Defective Sensors

    Degania Silicone is recalling 34,627 temperature sensing Foley catheters because defective sensors may display lower body temperatures than the patient's actual temperature.

    Product
    Temperature sensing catheter 16FR, Catalogue Number 102203101663DO Product Usage: This is sterile, single use, two-way silicone Foley Catheter with a thermistor embedded in the third lumen. It is intended for use in the drainage/ collection of urine from the urinary bladder…
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-2771-2018·2018-09-26

    Degania Foley Temperature Sensor Catheters Recalled for Defective Sensors

    Degania Silicone is recalling Foley Temperature Sensor Catheters because defective sensors may display lower body temperatures than the patient's actual temperature.

    Product
    Foley Temperature Sensor Catheter 14FR, Catalogue Number 102201101480JP Product Usage: This is sterile, single use, two-way silicone Foley Catheter with a thermistor embedded in the third lumen. It is intended for use in the drainage/ collection of urine from the urinary…
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-2768-2018·2018-09-26

    Degania Foley Temperature Sensor Catheters Recalled for Defective Sensors

    Degania Silicone is recalling 250 Foley Temperature Sensor Catheters because defective sensors may display lower body temperatures than the patient's actual temperature.

    Product
    Foley Temperature Sensor Catheter 8FR, Catalogue Number 102201100880JP Product Usage: This is sterile, single use, two-way silicone Foley Catheter with a thermistor embedded in the third lumen. It is intended for use in the drainage/ collection of urine from the urinary…
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-3211-2018·2018-09-26

    Accuray Recalls CyberKnife Calibration Film Due to Dimensional Inaccuracy

    Accuray is recalling 259 units of Ashland Box Film Ball Cube II calibration film because manufacturing defects may cause up to 0.5 mm of positional inaccuracy in CyberKnife systems.

    Product
    Ashland(TM) Box Film Ball Cube II EBT3, P/N: 027863-002 rev E Product is used to calibrate CyberKnife systems
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-2762-2018·2018-09-26

    Degania Silicone Recalls Temperature Sensor Foley Catheters Due to Defective Sensors

    Degania Silicone is recalling 1,025 Temperature Sensor Catheters because defective sensors may display lower body temperatures than actual patient readings.

    Product
    Temperature Sensor Catheter 18FR, Catalogue Number 102201101863BI Product Usage: This is sterile, single use, two-way silicone Foley Catheter with a thermistor embedded in the third lumen. It is intended for use in the drainage/ collection of urine from the urinary bladder…
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-3192-2018·2018-09-26

    Invivo Xper Flex Cardio Physiomonitoring System Recalled for Firmware Error

    Invivo Corporation is recalling 38 Xper Flex Cardio Physiomonitoring Systems distributed in the Netherlands due to incorrect firmware that may cause inaccurate ECG and heart rate displays.

    Product
    Xper Flex Cardio Physiomonitoring System Model # 453564669081 Software Version 1.6.0.0388
    Category
    Medical Device
    Distribution
    0 states
  • HighFDA (Devices)·Z-2781-2018·2018-09-26

    Degania Silicone Recalls Temperature Sensing Foley Catheters Due to Defective Sensors

    Degania Silicone, Ltd. is recalling 4,372 temperature sensing Foley catheters because defective sensors may display lower body temperatures than the patient's actual temperature.

    Product
    Temperature sensing catheter 14FR, Catalogue Number 102203101463DO Product Usage: This is sterile, single use, two-way silicone Foley Catheter with a thermistor embedded in the third lumen. It is intended for use in the drainage/ collection of urine from the urinary bladder…
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-3186-2018·2018-09-26

    Polymedco PATHFAST NTproBNP Cartridges Recalled for Measurement Bias

    Polymedco is recalling 272 PATHFAST NTproBNP cartridges due to a manufacturing defect that may cause measurement values to be 12% to 20% higher than actual.

    Product
    PATHFAST NTproBNP Catalog Number: PF1061-KUS PATHFAST¿ NTproBNP is a product for in-vitro diagnostic use with the in vitro diagnostic system PATHFAST¿ for the quantitative measurement of N-terminal-pro B-type Natriuretic Peptide (NT-proBNP) in heparinized or EDTA whole…
    Category
    Medical Device
    Distribution
    8 states
  • HighFDA (Devices)·Z-3209-2018·2018-09-26

    Exactech Recalls Optetrak Logic CC Stem Extension Due to Thread Defects

    Exactech is recalling one unit of the Optetrak Logic CC Stem Extension because it may have non-conforming internal threads.

    Product
    Optetrak Logic CC Stem Extension, 22mm x 160mm The OPTETRAK Logic CC Total Knee System is indicated for use in skeletally mature individuals undergoing primary surgery for total knee replacement due to osteoarthritis, osteonecrosis, rheumatoid arthritis and/or post-traumatic…
    Category
    Medical Device
    Distribution
    1 state
  • HighFDA (Devices)·Z-2991-2018·2018-09-26

    Mobius AIRO Mobile CT System Recalled for Software Fault

    Mobius Imaging is recalling 147 AIRO Mobile CT Systems because a software fault prevents the tube current modulation feature from working.

    Product
    AIRO Mobile CT System Model # MobiCT-32
    Category
    Medical Device
    Distribution
    0 states
  • HighFDA (Devices)·Z-2755-2018·2018-09-26

    Degania Silicone Recalls Temperature Sensor Foley Catheters Due to Defective Sensors

    Degania Silicone is recalling 11,488 temperature sensor Foley catheters because defective sensors may display lower body temperatures than actual.

    Product
    Temperature sensor Foley catheter 8FR, Catalogue Number 102205100880AB Product Usage: This is sterile, single use, two-way silicone Foley Catheter with a thermistor embedded in the third lumen. It is intended for use in the drainage/ collection of urine from the urinary…
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-2776-2018·2018-09-26

    Degania Silicone Recalls Temperature Sensing Foley Catheters Due to Defective Sensors

    Degania Silicone, Ltd. is recalling 100 units of temperature sensing Foley catheters because defective sensors may display lower body temperatures than actual.

    Product
    Temperature Sensing 100% Silicone Foley Catheter, sterile, Catalogue Number 102201101080MD Product Usage: This is sterile, single use, two-way silicone Foley Catheter with a thermistor embedded in the third lumen. It is intended for use in the drainage/ collection of urine…
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-3191-2018·2018-09-26

    Swissray Medical Recalls ddR Formula B X-ray Systems for Bucky Movement Risk

    Swissray Medical is recalling 27 ddR Formula B X-ray systems because the bucky component may move unexpectedly, posing a risk of injury to patients.

    Product
    ddR Formula B X-ray System, Model ddR Formula B X-ray system used for imaging
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-2780-2018·2018-09-26

    Degania Silicone Recalls Thermistor Foley Catheters Due to Defective Sensors

    Degania Silicone is recalling Thermistor Foley catheters because defective sensors may display lower body temperatures than the patient's actual temperature.

    Product
    Thermistor Foley catheter, nonsterile, Catalogue Number 102201101463MD Product Usage: This is sterile, single use, two-way silicone Foley Catheter with a thermistor embedded in the third lumen. It is intended for use in the drainage/ collection of urine from the urinary…
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-2775-2018·2018-09-26

    Degania Silicone Recalls Temperature Sensing Foley Catheters Due to Defective Sensors

    Degania Silicone is recalling 500 sterile Foley catheters with defective temperature sensors that may display inaccurate readings.

    Product
    Temperature Sensing 100% Silicone Foley Catheter, sterile, Catalogue Number 102201101680MD Product Usage: This is sterile, single use, two-way silicone Foley Catheter with a thermistor embedded in the third lumen. It is intended for use in the drainage/ collection of urine…
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-2784-2018·2018-09-26

    Degania Silicone Recalls Temperature Sensing Foley Catheters Due to Defective Sensors

    Degania Silicone, Ltd. is recalling 1,000 temperature sensing Foley catheters because defective sensors may display lower body temperatures than actual.

    Product
    Temperature sensing catheter 14FR, Catalogue Number 102201101463NU Product Usage: This is sterile, single use, two-way silicone Foley Catheter with a thermistor embedded in the third lumen. It is intended for use in the drainage/ collection of urine from the urinary bladder…
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-2773-2018·2018-09-26

    Degania Silicone Recalls Temperature Sensing Foley Catheters Due to Defective Sensors

    Degania Silicone is recalling 500 temperature-sensing Foley catheters because defective sensors may display lower body temperatures than actual.

    Product
    Temperature Sensing 100% Silicone Foley Catheter, sterile, Catalogue Number 102201101480MD Product Usage: This is sterile, single use, two-way silicone Foley Catheter with a thermistor embedded in the third lumen. It is intended for use in the drainage/ collection of urine…
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-2757-2018·2018-09-26

    Degania Silicone Recalls Temperature Sensor Catheters Due to Defective Sensors

    Degania Silicone, Ltd. is recalling 1,588 Temperature Sensor Catheters because defective sensors may display lower body temperatures than the patient's actual temperature.

    Product
    Temperature Sensor Catheter 8FR, Catalogue Number 102201100863BI Product Usage: This is sterile, single use, two-way silicone Foley Catheter with a thermistor embedded in the third lumen. It is intended for use in the drainage/ collection of urine from the urinary bladder…
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-2777-2018·2018-09-26

    Degania Silicone Recalls Thermistor Foley Catheters Due to Defective Sensors

    Degania Silicone is recalling 2,784 Thermistor Foley catheters because defective sensors may display lower body temperatures than the patient's actual temperature.

    Product
    Thermistor Foley catheter, nonsterile, Catalogue Number 102201101863MD Product Usage: This is sterile, single use, two-way silicone Foley Catheter with a thermistor embedded in the third lumen. It is intended for use in the drainage/ collection of urine from the urinary…
    Category
    Medical Device
    Distribution
    Distributed nationwide