The Recall Desk

Recalls

All US Product Recalls

Searchable archive of product recalls from the FDA, USDA FSIS, CPSC, NHTSA, and EPA — translated into plain English, severity-scored from objective criteria, and always linked back to the original government notice.

How to read this listing

Every recall on this list has been scored 1–5 against an objective rubric from the agency’s own classification, presence of reported injuries or deaths, distribution scope, and hazard type. The colored severity badge on each card is the score — 5 (Critical) means the agency’s own highest classification or reported deaths or serious injuries, 4 (Severe) means significant injury or hospitalization risk, and 3 (High) means a known hazard with no confirmed harm yet. Filter to 3+ (High and above) if you want to skip routine Class-III notices.

The agency chips above control which streams are visible. FDA spans food, drugs, and medical devices. USDA FSIS is meat, poultry, and processed egg products. CPSC is household consumer products. NHTSA is vehicles, tires, and child seats. Nationwide on a card means the recall affects every US state — most do — while a state list means distribution was limited.

Click any recall to see the full summary, affected lot codes and model numbers, the agency’s own classification, and a direct link back to the original government notice for verification.

Filters

Agency

Category

Minimum severity

Clear filters

21376–21400 of 40175

  • SevereFDA (Devices)·Z-0767-2019·2019-02-06

    West Pharma Vial2BAG DC 13mm Recalled for Adverse Events

    West Pharma Services IL, Ltd. is recalling Vial2BAG DC 13mm devices due to reported adverse events and potential inadequate drug admixing.

    Product
    Vial2BAG DC 13mm, REF 6070111, Single use only, Rx Only
    Category
    Medical Device
    Distribution
    1 state
  • SevereFDA (Devices)·Z-0762-2019·2019-02-06

    Edwards Lifesciences Recalls Swan Ganz Catheters Due to Lumen Assembly Error

    Edwards Lifesciences is recalling 277 Swan Ganz Thermodilution Catheters because incorrect lumen assembly causes reversed pressure readings.

    Product
    Edwards Lifesciences Swan Ganz Thermodilution Catheter, REF131F 7J Product Usage: The Swan-Ganz Thermodilution Catheter provides a diagnostic tool for physicians to rapidly determine hemodynamic pressures and cardiac output when used with a compatible cardiac output…
    Category
    Medical Device
    Distribution
    0 states
  • SevereFDA (Devices)·Z-0763-2019·2019-02-06

    Edwards Lifesciences Recalls Swan Ganz Catheters Due to Lumen Reversal

    Edwards Lifesciences is recalling 875 units of Swan Ganz Thermodilution Catheters because incorrect lumen assembly causes reversed pressure readings.

    Product
    Edwards Lifesciences Swan Ganz Thermodilution Catheter, REF 131F7P (5 pack of 131F7) Product Usage: The Swan-Ganz Thermodilution Catheter provides a diagnostic tool for physicians to rapidly determine hemodynamic pressures and cardiac output when used with a compatible cardiac…
    Category
    Medical Device
    Distribution
    0 states
  • HighFDA (Devices)·Z-0784-2019·2019-02-06

    Roche Anti-HAV IgM Test System Recalled for Erroneous Results

    Roche Diagnostics is recalling Anti-HAV IgM test systems due to potential erroneous result messages on the cobas e 602 module.

    Product
    Anti-HAV IgM test System, Material No. 11820591160 Product Usage: The Roche Elecsys Anti-HAV IgM immunoassay is used for the in vitro qualitative detection of IgM antibodies to hepatitis A virus (anti-HAV IgM) in human serum and plasma (potassium EDTA, lithium or sodium…
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-0785-2019·2019-02-06

    Medtronic Recalls Reservoir Y Adapters Due to Tearing and Cracking

    Medtronic is recalling 1,100 Reservoir Y Adapters because the Y-connectors may tear or crack, potentially causing blood loss and circuit breaches during autotransfusion.

    Product
    RESERVOIR "Y" ADAPTER, (a) BT920 (b) BT926 Suction and Anticoagulation assemblies are indicated for use with an autotransfusion system between the suction tip and reservoir.
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-0790-2019·2019-02-06

    XTant Medical Recalls Calix Lumbar Spinal Implants for Sterilization Concerns

    XTant Medical is recalling 632 Calix Lumbar Spinal Implant System units due to a change in packaging and sterilization method to address potential risks of compromised sterilization.

    Product
    Calix Lumbar Spinal Implant System: 22mm x 10mm x 7mm PLIF ASSY PC, REF X034-0172PC 22mm x 10mm x 8mm PLIF ASSY PC, REF X034-0173PC 26mm x 10mm x 6mm PLIF ASSY PC, REF X034-0240PC 26mm x 10mm x 8mm PLIF ASSY PC, REF X034-0241PC 26mm x 10mm x 10mm PLIF ASSY PC, REF…
    Category
    Medical Device
    Distribution
    12 states
  • HighFDA (Devices)·Z-0779-2019·2019-02-06

    Haemonetics Recalls Cell Saver Elite Processing Kits Due to Leakage Risk

    Haemonetics is recalling 4,576 Cell Saver Elite Processing Kits because the 70ml centrifuge bowl may leak during cell salvage processing.

    Product
    Cell Saver Elite Processing Kit, 70 ml, Product ID: CSE-P-70 Indications for Use: The Haemonetics Cell Saver 5/5+ System and its related accessory components are intended to be used for the recovery and processing of blood shed during or subsequent to an operation or as a…
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-0788-2019·2019-02-06

    GE Healthcare Recalls LOGIQ P6 Ultrasound Scanners Due to Probe Heating

    GE Healthcare is recalling 1,897 LOGIQ P6 ultrasound scanners (Models BT07 and BT09) because the probe-head surface can reach increased temperatures during specific scans.

    Product
    GE HEALTHCARE LOGIQ P6 Premium TruScan(TM) Imagining Technology: (a) Model BT09 (b) Model BT07 Product Usage: The device is intended for use by a qualified physician for ultrasound evaluation of Fetal; Abdominal; Pediatric; Small Organ (breast, testes, thyroid); Neonatal…
    Category
    Medical Device
    Distribution
    0 states
  • HighFDA (Devices)·Z-0786-2019·2019-02-06

    Medtronic Recalls Suction and Anticoagulation Assemblies Due to Connector Defects

    Medtronic is recalling 7,587 Suction and Anticoagulation Assemblies because Y-connectors may tear or crack, potentially causing leaks and blood loss.

    Product
    SUCTION / ANTICOAGULATION ASSEMBLY (a) BTC93 (b) BTC98 Suction and Anticoagulation assemblies are indicated for use with an autotransfusion system between the suction tip and reservoir.
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-0780-2019·2019-02-06

    Siemens Recalls Atellica IM 1300 Analyzer Software Due to Multiple Issues

    Siemens Healthcare Diagnostics is recalling 217 Atellica IM 1300 Analyzers worldwide due to multiple issues identified in software version 1.14.2 and lower.

    Product
    Atellica IM 1300 Analyzer. In vitro diagnostic testing of clinical specimens. Siemens Material Number (SMN): 1066001 - Product Usage: The Atellica Solution is a multi-component system for in vitro diagnostic testing of clinical specimens. The system is intended for the…
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-0782-2019·2019-02-06

    Siemens Atellica Analyzer Software Recall for Multiple System Issues

    Siemens Healthcare Diagnostics is recalling Atellica analyzer software versions 1.14.2 and lower due to multiple identified issues affecting in vitro diagnostic testing.

    Product
    Atellica CH 930 Analyzer. In vitro diagnostic testing of clinical specimens Siemens Material Number (SMN): 1 067000 - Product Usage: The Atellica Solution is a multi-component system for in vitro diagnostic testing of clinical specimens. The system is intended for the…
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-0791-2019·2019-02-06

    3M Surgical Clipper Recalled for Overheating Risk During Charging

    3M is recalling 19,860 Surgical Clippers due to reports of overheating when charging from modified sine wave power inverters in mobile EMS settings.

    Product
    3M(TM) Surgical Clipper by Remington(R) REF 9602 The 3M Surgical Clipper Classic, 9602, is intended to remove body hair and head hair from patients in preparation for any medical procedure requiring hair removal.
    Category
    Medical Device
    Distribution
    0 states
  • HighFDA (Devices)·Z-0778-2019·2019-02-06

    Haemonetics Recalls Cell Saver 5/5+ Bowl Sets Due to Leakage

    Haemonetics is recalling 11,208 units of Cell Saver 5/5+ Bowl Sets because the 70ml centrifuge bowls may leak during cell salvage processing.

    Product
    Cell Saver 5/5+ Bowl Set, 70ml, Product ID: 0291E-00, 0291A-00 Indications for Use: The Haemonetics Cell Saver 5/5+ System and its related accessory components are intended to be used for the recovery and processing of blood shed during or subsequent to an operation or as a…
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-0781-2019·2019-02-06

    Siemens Atellica Analyzer Software Recall for Multiple System Issues

    Siemens Healthcare Diagnostics is recalling Atellica analyzer software versions 1.14.2 and lower due to multiple identified issues affecting in vitro diagnostic testing systems.

    Product
    Atellica IM 1600 Analyzer: In vitro diagnostic testing of clinical specimens - Product Usage: The Atellica Solution is a multi-component system for in vitro diagnostic testing of clinical specimens. The system is intended for the qualitative and quantitative analysis of various…
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • ModerateFDA (Devices)·Z-0774-2019·2019-02-06

    Wilson-Cook Medical Recalls Mislabelled Captura Serrated Biopsy Forceps

    Wilson-Cook Medical is recalling 8 units of Captura Serrated Biopsy Forceps in France, Ireland, and Japan due to mislabeling.

    Product
    COOK MEDICAL Captura SERRATED BIOPSY FORCEPS WITH SPIKE,2.4 mm, 230 cm, with Spike, DBF-2.4SN-230SP-20-S
    Category
    Medical Device
    Distribution
    0 states
  • ModerateFDA (Devices)·Z-0783-2019·2019-02-06

    Siemens Recalls Atellica Sample Handler Prime Due to Software Issues

    Siemens Healthcare Diagnostics is recalling 410 Atellica Sample Handler Prime units because software version 1.14.2 and lower contain multiple issues.

    Product
    Atellica Sample Handler Prime. In vitro diagnostic testing of clinical specimens - Siemens Material Number (SMN): 11069001 - Product Usage: The Atellica Solution is a multi-component system for in vitro diagnostic testing of clinical specimens. The system is intended for the…
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • ModerateFDA (Devices)·Z-0789-2019·2019-02-06

    Abbott Recalls Alinity i Processing Modules Due to Gear Pump Assembly Issue

    Abbott is recalling Alinity i Processing Modules because a gear pump assembly issue causes foaming and dried residue in the wash buffer reservoir.

    Product
    Alinity i Processing Module, (a) Part Number 03R65-01 (b) Part Number A-30108552-01 (c) Part Number A-30109694-01 Product Usage: The Alinity i processing module is a fully automated immunoassay analyzer allowing random and continuous access, as well as priority and…
    Category
    Medical Device
    Distribution
    0 states
  • ModerateFDA (Devices)·Z-0775-2019·2019-02-06

    Wilson-Cook Medical Recalls Mislabeled Captura Serrated Forceps

    Wilson-Cook Medical is recalling 6 units of Captura Serrated Forceps due to mislabeling that swapped spike and no-spike models.

    Product
    COOK MEDICAL Captura SERRATED LARGE FORCEP-NO SPIKE, 2.4 mm, 230 cm, DBF-2.4SL- 230-20-S
    Category
    Medical Device
    Distribution
    0 states
  • LowFDA (Devices)·Z-0777-2019·2019-02-06

    Riverpoint Medical Recalls 18G Brachy Grids for Incorrect CE Labeling

    Riverpoint Medical is recalling 443 units of 18G Brachy Grids because they are labeled with a CE mark despite not being approved in the European Union.

    Product
    RP RIVERPOINT MEDICAL 18G Brachy Grid B&K(TM) style, REF RPG-18 Product Usage: The Riverpoint Brachytherapy Grid is used during Brachytherapy procedures to assist with the precise insertion of the needle or other probes. The Riverpoint Brachytherapy Grid is placed against…
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • CriticalFDA (Devices)·Z-0737-2019·2019-01-30

    Draeger Medical Recalls VentStar Basic 250 Breathing Circuits Due to Assembly Error

    Draeger Medical is recalling 170 units of VentStar Basic 250 breathing circuits because they were incorrectly pre-assembled, creating a risk of incorrect connection or short circuiting.

    Product
    VentStar Basic 250, Catalog Number MP00350 Breathing Circuits/Anesthesia Sets without a Water trap: Breathing circuit for conduction of respiratory gases between anesthesia machine or ventilator and patient. Intended for single use only. Breathing Circuits/Anesthesia Sets…
    Category
    Medical Device
    Distribution
    12 states
  • SevereFDA (Devices)·Z-0736-2019·2019-01-30

    Draeger Medical Recalls VentStar Basic 180 Breathing Circuits

    Draeger Medical is recalling 325 units of VentStar Basic 180 breathing circuits due to incorrect pre-assembly that may cause incorrect connections.

    Product
    VentStar Basic 180, Catalog Number MP00349 Breathing Circuits/Anesthesia Sets without a Water trap: Breathing circuit for conduction of respiratory gases between anesthesia machine or ventilator and patient. Intended for single use only. Breathing Circuits/Anesthesia Sets…
    Category
    Medical Device
    Distribution
    12 states
  • SevereFDA (Devices)·Z-0738-2019·2019-01-30

    Draeger Medical Recalls VentStar Basic Breathing Circuits Due to Assembly Error

    Draeger Medical is recalling 525 units of VentStar Basic (P) 180 breathing circuits because they were incorrectly pre-assembled, posing a risk of incorrect connection or short circuiting.

    Product
    VentStar Basic (P) 180, Catalog Number MP00351 Breathing Circuits/Anesthesia Sets without a Water trap: Breathing circuit for conduction of respiratory gases between anesthesia machine or ventilator and patient. Intended for single use only. Breathing Circuits/Anesthesia…
    Category
    Medical Device
    Distribution
    12 states
  • SevereFDA (Devices)·Z-0740-2019·2019-01-30

    Draeger Medical Recalls VentStar Watertrap Breathing Circuits Due to Assembly Error

    Draeger Medical is recalling 25 units of VentStar Watertrap (P) 180 breathing circuits because they were incorrectly pre-assembled, posing a risk of incorrect connection.

    Product
    VentStar Watertrap (P) 180, Catalog Number MP00361 Breathing Circuits/Anesthesia Sets without a Water trap: Breathing circuit for conduction of respiratory gases between anesthesia machine or ventilator and patient. Intended for single use only. Breathing Circuits/Anesthesia…
    Category
    Medical Device
    Distribution
    12 states
  • SevereFDA (Devices)·Z-0753-2019·2019-01-30

    Teleflex Percuvance Percutaneous Shaft Recall for Potential Tip Breakage

    Teleflex Medical is recalling Percuvance Percutaneous Shafts because the distal tip may deform or break off during use, potentially compromising the tool's security.

    Product
    Percuvance(R) Percutaneous Shaft 29cm, PCVSH3 The Percutaneous Surgical System with 5mm attachments is indicated for the means to penetrate Intended Use: soft tissue to access certain areas of the abdomen. The system is used to grasp, manipulate, cut, cauterize and deliver…
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • SevereFDA (Devices)·Z-0754-2019·2019-01-30

    Teleflex Percuvance Percutaneous Shaft Recall Due to Tip Breakage Risk

    Teleflex Medical is recalling 1,962 units of Percuvance Percutaneous Shafts because the distal tip may deform or break off during use, potentially requiring surgical intervention.

    Product
    Percuvance(R) Percutaneous Shaft 36cm, PCVSHL3 The Percutaneous Surgical System with 5mm attachments is indicated for the means to penetrate Intended Use: soft tissue to access certain areas of the abdomen. The system is used to grasp, manipulate, cut, cauterize and deliver…
    Category
    Medical Device
    Distribution
    Distributed nationwide