The Recall Desk

Recalls

All US Product Recalls

Searchable archive of product recalls from the FDA, USDA FSIS, CPSC, NHTSA, and EPA — translated into plain English, severity-scored from objective criteria, and always linked back to the original government notice.

How to read this listing

Every recall on this list has been scored 1–5 against an objective rubric from the agency’s own classification, presence of reported injuries or deaths, distribution scope, and hazard type. The colored severity badge on each card is the score — 5 (Critical) means the agency’s own highest classification or reported deaths or serious injuries, 4 (Severe) means significant injury or hospitalization risk, and 3 (High) means a known hazard with no confirmed harm yet. Filter to 3+ (High and above) if you want to skip routine Class-III notices.

The agency chips above control which streams are visible. FDA spans food, drugs, and medical devices. USDA FSIS is meat, poultry, and processed egg products. CPSC is household consumer products. NHTSA is vehicles, tires, and child seats. Nationwide on a card means the recall affects every US state — most do — while a state list means distribution was limited.

Click any recall to see the full summary, affected lot codes and model numbers, the agency’s own classification, and a direct link back to the original government notice for verification.

Filters

Agency

Category

Minimum severity

Clear filters

20376–20400 of 40124

  • HighFDA (Devices)·Z-1692-2019·2019-06-12

    Philips IntelliVue MX40 Patient Monitor Recalled for Battery Defect

    Philips is recalling IntelliVue MX40 Patient Monitors because they may experience increased power consumption and lack visual or audible notifications when in a low battery state.

    Product
    IntelliVue MX40 Patient Monitor, Product number 865350, Service # USA: 453564262491 and 453564262511 - Product Usage: Used primarily as a traditional telemetry medical device. It connects to the IntelliVue Information Center by way of a wireless network. When the MX40 is…
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-1739-2019·2019-06-12

    Zimmer Biomet Recalls Oxford Partial Knee Drill Guides Due to Corrosion Risk

    Zimmer Biomet is recalling 3 units of Oxford Partial Knee Phase 3 Femoral Drill Guides because incorrect raw materials may cause corrosion.

    Product
    Biomet Oxford Partial Knee Phase 3 / Domed Lateral Femoral Drill Guide: Item Number: 32-420324Phase 3 Femoral Drill Guide Extra Large
    Category
    Medical Device
    Distribution
    7 states
  • HighFDA (Devices)·Z-1726-2019·2019-06-12

    Stryker High Flow II Tube Set Recalled for Leaking and Detachment

    Stryker is recalling High Flow II Tube Sets because they consistently leak and detach from the cassette. The product is a medical device accessory distributed nationwide and internationally.

    Product
    Stryker High Flow II Tube Set, REF 0620-040-660, an accessory to the PNEUMO SURE High Flow Insufflator, Disposable, Rx, Sterile.
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-1701-2019·2019-06-12

    AngioDynamics Solero Microwave Tissue Ablation System Recall for Delayed Procedures

    AngioDynamics is recalling 178 Solero Microwave Tissue Ablation System generators to update manuals and software, reducing the risk of patient harm from procedure delays.

    Product
    AngioDynamics Solero Microwave Tissue Ablation (MTA) System: (1) Solero Generator Refurbished PG Catalog Number: 12740000D UPN: H78712740000D0 (2) Solero Generator Refurbished US PG Catalog Number: 12740000USD ;UPN: H78712740000USD0
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-1712-2019·2019-06-12

    Flowonix Prometra Clinician Programmer Software Recall for Unexpected Pump Shutdowns

    Flowonix Medical Inc is recalling Prometra Clinician Programmer software due to a defect that can cause intrathecal pumps to shut down unexpectedly.

    Product
    Prometra Clinician Programmer (Touchscreen) Software version 2.00.29, Cat. No. 12828, used with Prometra and Prometra II Programmable Pumps. For intrathecal infusion.
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-1737-2019·2019-06-12

    Zimmer Biomet Recalls Oxford Partial Knee Drill Guides Due to Material Error

    Zimmer Biomet is recalling 10 units of Oxford Partial Knee Phase 3 Femoral Drill Guides because incorrect raw material was used, potentially leading to corrosion.

    Product
    Biomet Oxford Partial Knee Phase 3 / Domed Lateral Femoral Drill Guide: Item Number: 32-420322 Phase 3 Femoral Drill Guide Medium
    Category
    Medical Device
    Distribution
    7 states
  • HighFDA (Devices)·Z-1695-2019·2019-06-12

    Philips IntelliVue MX40 Patient Monitor Recalled for Battery Notification Defects

    Philips is recalling IntelliVue MX40 Patient Monitors because they may fail to provide visual or audible alerts when the battery is low.

    Product
    IntelliVue MX40 Patient Monitor, Product number 857146, Service Numbers All: 453564711531 and 453564711541 - Product Usage: Used primarily as a traditional telemetry medical device. It connects to the IntelliVue Information Center by way of a wireless network. When the MX40 is…
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-1702-2019·2019-06-12

    AngioDynamics Solero Microwave Tissue Ablation System Recall for Delayed Procedure Risk

    AngioDynamics is recalling Solero Microwave Tissue Ablation System applicators to reduce the risk of patient harm from delayed procedures via manual and software updates.

    Product
    AngioDynamics Solero Microwave Tissue Ablation (MTA) System: (1) Solero Applicator 14cm PG Catalog Number: 700106001 UPN:H7877001060010; (2)Solero Applicator 14cm PG US Catalog Number:700106001US UPN: H787700106001US0 Product Usage: The Solero Microwave Tissue…
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-1693-2019·2019-06-12

    Philips IntelliVue MX40 Patient Monitor Recalled for Battery Issues

    Philips is recalling IntelliVue MX40 Patient Monitors because they may consume excess power and fail to provide visual or audible low-battery alerts.

    Product
    IntelliVue MX40 Patient Monitor, Product number 865351, Service Numbers All other countries: 453564451791, 453564451811, 453564467721, 453564467741, 453564467761, 453564467781, 453564467801, 453564467821, 453564467841, 453564467861, 453564262531, and 453564262551 - Product…
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-1696-2019·2019-06-12

    Accuray Recalls CyberKnife M6 Systems Due to Collimator Housing Drop Risk

    Accuray is recalling 68 CyberKnife M6 systems because an interlock interruption during automated housing exchange can cause the secondary collimator housing to drop.

    Product
    CyberKnife M6, Part Numbers: a) 0660000: CyberKnife Treatment Delivery System b) 053101-010: CyberKnife M6 FI Treatment Delivery System c) 053201-010: CyberKnife M6 FM Treatment Delivery System d) 053301-010: CyberKnife M6 FIM Treatment Delivery System e) 053401-010…
    Category
    Medical Device
    Distribution
    0 states
  • HighFDA (Devices)·Z-1714-2019·2019-06-12

    Flowonix Prometra Programmable Pump Recalled for Unexpected Shutdowns

    Flowonix Medical Inc. is recalling 809 Prometra Programmable Pumps because a software defect can cause the pump to shut down unexpectedly.

    Product
    Prometra Programmable Pump, Cat. No. 11827 For intrathecal infusion.
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-1727-2019·2019-06-12

    Stryker High Flow Tube Set Recalled for Leaking and Detachment

    Stryker is recalling High Flow Tube Sets with Real-Time Pressure Sensing because they consistently leak and detach from the cassette.

    Product
    Stryker High Flow Tube Set with Real-Time Pressure Sensing (RTP), REF 0620-040-680, an accessory to the PNEUMO SURE High Flow Insufflator, Disposable, Rx, Sterile.
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-1708-2019·2019-06-12

    Stryker AXS Infinity LS Plus Long Sheath Recalled for Brittle Behavior

    Stryker Neurovascular is recalling AXS Infinity LS Plus Long Sheaths because a design issue can cause the device to become brittle and fracture.

    Product
    Stryker Neurovascular, AXS Infinity LS Plus Long Sheath, Rx Only, Sterile EO, USA Customer Service 855-91 NEURO (916-3876): a) REF: INC-11196-70, 70cm, b) REF: INC-11196-80, 80cm, c) REF: INC-11196-90, 90cm Product Usage: Indications for Use: The AXS Infinity LS Plus…
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-1740-2019·2019-06-12

    Zimmer Biomet Recalls Oxford Partial Knee Drill Guides Due to Corrosion Risk

    Zimmer Biomet is recalling 14 units of Oxford Partial Knee Phase 3 Femoral Drill Guides because incorrect raw materials were used, potentially leading to corrosion.

    Product
    Biomet Oxford Partial Knee Phase 3 / Domed Lateral Femoral Drill Guide: Item Number: 32-421060 Phase 3 Femoral Drill Guide Extra Small
    Category
    Medical Device
    Distribution
    7 states
  • HighFDA (Devices)·Z-1694-2019·2019-06-12

    Philips IntelliVue MX40 Patient Monitor Recalled for Battery Notification Issues

    Philips is recalling IntelliVue MX40 Patient Monitors because they may fail to provide visual or audible low-battery alerts, potentially disrupting patient monitoring.

    Product
    IntelliVue MX40 Patient Monitor, Product number 865352, Service Numbers All: 453564262571, 453564262591, 453564615311, 453564615331, 453564615351, and 453564615371 Product Usage: Used primarily as a traditional telemetry medical device. It connects to the IntelliVue Information…
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-1697-2019·2019-06-12

    Accuray CyberKnife VSI Robotic Radiosurgery Systems Recalled for Collimator Housing Drop

    Accuray is recalling 43 CyberKnife VSI systems because an interlock interruption during automated housing exchange can cause the secondary collimator housing to drop.

    Product
    CyberKnife VSI, Part Numbers: a) 032500-010: CyberKnife VSI 9.6.x Robotic Radiosurgery System for Japan b) 033000-010: CyberKnife VSI Robotic Radiosurgery System
    Category
    Medical Device
    Distribution
    0 states
  • HighFDA (Devices)·Z-1728-2019·2019-06-12

    Stryker Heated Tube Set Recall for Leaking and Detachment Issues

    Stryker is recalling 19,083 boxes of Heated Tube Sets with Real-Time Pressure Sensing because they consistently leak and detach from the cassette.

    Product
    Stryker Heated Tube Set with Real-Time Pressure Sensing (RTP), REF 0620-040-690, an accessory to the PNEUMO SURE High Flow Insufflator, Disposable, Rx, Sterile.
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-1700-2019·2019-06-12

    AngioDynamics Solero Microwave Tissue Ablation System Recall for Delayed Procedures

    AngioDynamics is recalling 314 Solero Microwave Tissue Ablation System generators to update manuals and software, reducing the risk of patient harm from procedure delays.

    Product
    AngioDynamics Solero Microwave Tissue Ablation (MTA) System: Solero Generator PG (1) Catalog Number:12740000 UPN: H78712740000 (2) Solero Generator US PG Catalog Number: 12740000US UPN: H78712740000US0 (3) Solero Generator Service Loaner PG Catalog No: 12740000L…
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-1704-2019·2019-06-12

    AngioDynamics Solero Microwave Tissue Ablation System Recall for Procedure Delay Risk

    AngioDynamics is recalling 1,846 Solero Microwave Tissue Ablation System applicators to reduce the risk of patient harm from procedure delays.

    Product
    AngioDynamics Solero Microwave Tissue Ablation (MTA) System: (1)Solero Applicator 29cm PG Catalog Number:700106003 UPN: H7877001060030 (2)Solero Applicator 29cm PG US Catalog Number:700106003US UPN:H787700106003US0 Product Usage: The Solero Microwave Tissue…
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-1698-2019·2019-06-12

    Microline TLS2 Thermal Ligating Shears Recalled for Actuation Failure

    Microline Surgical is recalling 115 TLS2 Thermal Ligating Shears because they fail to actuate, preventing them from performing their intended surgical function.

    Product
    TLS2 Thermal Ligating Shears, Catalog Number 132-131D
    Category
    Medical Device
    Distribution
    2 states
  • HighFDA (Devices)·Z-1719-2019·2019-06-12

    Centurion Medical Recalls Sterile Nail Clippers Due to Compromised Seals

    Centurion Medical Products Corporation is recalling sterile nail clippers because the sterile barrier pouch seals are compromised, potentially affecting sterility.

    Product
    Sterile Nail Clipper Concave Edge (67710)
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-1736-2019·2019-06-12

    Zimmer Biomet Recalls Oxford Partial Knee Drill Guides Due to Corrosion Risk

    Zimmer Biomet is recalling 41 Oxford Partial Knee Phase 3 Femoral Drill Guides because incorrect raw material was used in the screw component, which could lead to corrosion.

    Product
    Biomet Oxford Partial Knee Phase 3 / Domed Lateral Femoral Drill Guide: Item Number: 32-420321 Phase 3 Femoral Drill Guide Small
    Category
    Medical Device
    Distribution
    7 states
  • HighFDA (Devices)·Z-1724-2019·2019-06-12

    Cook Inc. Recalls Pericardiocentesis Catheter Trays Due to Small Distal Hole

    Cook Inc. is recalling two Pericardiocentesis Catheter Trays because the catheter distal end hole may be too small. The affected units were distributed domestically and internationally.

    Product
    Pericardiocentesis Catheter Tray, RPN C-PCSY-850-LHSC-061096, Order # G10030
    Category
    Medical Device
    Distribution
    0 states
  • HighFDA (Devices)·Z-1713-2019·2019-06-12

    Flowonix Prometra Programmer Software Recall for Unexpected Pump Shutdowns

    Flowonix Medical Inc is recalling 40 units of Prometra Programmer software due to a defect that causes intrathecal pumps to shut down unexpectedly.

    Product
    Prometra Programmer (Grand Prime) Software version 2.00.29, Cat. No. 13828, used with Prometra and Prometra II Programmable Pumps. For intrathecal infusion.
    Category
    Medical Device
    Distribution
    Distributed nationwide