The Recall Desk

Recalls

All US Product Recalls

Searchable archive of product recalls from the FDA, USDA FSIS, CPSC, NHTSA, and EPA — translated into plain English, severity-scored from objective criteria, and always linked back to the original government notice.

How to read this listing

Every recall on this list has been scored 1–5 against an objective rubric from the agency’s own classification, presence of reported injuries or deaths, distribution scope, and hazard type. The colored severity badge on each card is the score — 5 (Critical) means deaths or serious injuries are reported, 4 (Severe) means significant injury or hospitalization risk, and 3 (High) means a known hazard with no confirmed harm yet. Filter to 3+ (High and above) if you want to skip routine Class-III notices.

The agency chips above control which streams are visible. FDA spans food, drugs, and medical devices. USDA FSIS is meat, poultry, and processed egg products. CPSC is household consumer products. NHTSA is vehicles, tires, and child seats. Nationwide on a card means the recall affects every US state — most do — while a state list means distribution was limited.

Click any recall to see the full summary, affected lot codes and model numbers, the agency’s own classification, and a direct link back to the original government notice for verification.

Filters

Agency

Category

Minimum severity

Clear filters

6526–6550 of 39844

  • HighFDA (Devices)·Z-2468-2024·2024-08-14

    GE MR450w 1.5T MRI software defect may cause implant overheating

    GE MR450w 1.5T MRI systems may overheat MR-conditional implants when using specific imaging sequences with Low SAR Mode. No injuries reported; 1,607 units affected worldwide.

    Product
    Optima MR450w 1.5T, with affected software versions: DV24.0, DV25.0, DV25.1, DV26.0, DV29.1, MR30.0, MR30.1; Nuclear Magnetic Resonance Imaging System
    Category
    Medical Device
    Distribution
    0 states
  • HighFDA (Devices)·Z-2530-2024·2024-08-14

    Baxter SIGMA Spectrum Infusion Pump recall due to inadequate pre-release testing

    Baxter Healthcare is recalling the SIGMA Spectrum Infusion Pump (Serial 2158913) distributed nationwide in Florida. The recall is due to improperly performed testing procedures prior to release, which may affect device function.

    Product
    Baxter SIGMA Spectrum Infusion Pump, Product Code: 35700BAX2
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-2476-2024·2024-08-14

    GE SIGNA MR360 MRI System Software Defect May Overheat Implants

    GE SIGNA MR360 MRI systems with software versions SV25.5 or SV25.6 may exceed prescribed RF exposure limits in certain scan configurations, potentially overheating MR-conditional implants. The recall affects 81 units worldwide.

    Product
    SIGNA MR360, with affected software versions: SV25.5, SV25.6; Nuclear Magnetic Resonance Imaging System
    Category
    Medical Device
    Distribution
    0 states
  • HighFDA (Devices)·Z-2510-2024·2024-08-14

    Philips IntelliVue Patient Monitors lack critical software monitoring options

    Philips IntelliVue Patient Monitors (models MX400/450/500/550) with software versions L.x or M.x are missing three monitoring capability options (MOS, M06, M20) due to software changes, creating potential for delayed or incorrect treatment. The recall involves 18 units distributed nationwide and internationally.

    Product
    IntelliVue Patient Monitor MX500 with L.x and M.x; K.x provided with L.x entitlements Model Number: 866064. Indicated for use by health care professionals for monitoring and recording of, and to generate alarms for, multiple physiological parameters of adults, pediatrics, and
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-2464-2024·2024-08-14

    GE Discovery MR750 MRI System Software May Cause Implant Overheating

    GE HealthCare is recalling certain MR750 MRI systems with specific software versions. When using particular imaging sequences in Low SAR Mode without the Optimized T2 FLAIR option, the radiofrequency field can exceed safe limits, risking overheating of MR-conditional implants.

    Product
    Discovery MR750 3.0T, with affected software versions: DV24.0, DV25.0, DV25.1, DV26.0 to DV26.4, DV29.1, MR30.0, MR30.1; Nuclear Magnetic Resonance Imaging System
    Category
    Medical Device
    Distribution
    0 states
  • ModerateFDA (Devices)·Z-2504-2024·2024-08-14

    Soltive Premium SuperPulsed Laser System Wireless Footswitch Pairing Difficulties

    The Soltive Premium SuperPulsed Laser System may experience difficulties pairing its wireless footswitch, potentially delaying surgical procedures. The recall affects units distributed worldwide.

    Product
    Soltive Premium SuperPulsed Laser System, Model TFL-PLS
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • ModerateFDA (Devices)·Z-2505-2024·2024-08-14

    Surgical Laser System Recalled Due to Wireless Footswitch Pairing Difficulties

    Olympus Corporation is recalling the Soltive Pro SuperPulsed Laser System due to difficulties pairing its wireless footswitch, which could delay surgical procedures. The recall affects 253 units worldwide.

    Product
    Soltive Pro SuperPulsed Laser System, Model TFL-SLS
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • ModerateFDA (Devices)·Z-2506-2024·2024-08-14

    Wireless Footswitch for Soltive Laser System Recalled for Pairing Difficulties

    Olympus Corporation of the Americas is recalling 1,838 units of the Soltive SuperPulsed Laser System Wireless Footswitch due to pairing difficulties that may delay surgical procedures.

    Product
    Soltive SuperPulsed Laser System Wireless Footswitch, Model TFL-AFSWL
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • LowFDA (Devices)·Z-2535-2024·2024-08-14

    NavaClick Syringe and Needle Model LIN02 Recalled for Incorrect Carton GTIN Label

    Lineage Biomedical is recalling NavaClick Syringe and Needle Model LIN02 due to an incorrect GTIN number on carton packaging. The recall affects 215 cartons distributed in Michigan and South Carolina.

    Product
    NavaClick Syringe and Needle, Model Number LIN02
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-2450-2024·2024-08-07

    Hip Replacement Prosthesis Recalled Due to Increased Postoperative Fracture Risk

    Zimmer hip replacement prostheses are recalled due to increased risk of postoperative femoral fracture. Affected devices are distributed nationwide; patients should consult their surgeon.

    Product
    CPT Hip Joint Metal/Polymer Semi-Constrained Cemented Prosthesis, used in Hip Arthroplasty; Item Numbers 00811400000 00811400010 00811400100 00811400110 00811400200 00811400210 00811400218 00811400230 00811400300 00811400310 00811400318 00811400330 00811400400 008114
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-2452-2024·2024-08-07

    Boston Scientific Capio SLIM Device May Fail During Surgical Procedures

    Some lots of the Capio SLIM suture capturing device may fail to consistently catch sutures during surgery, potentially requiring extended procedures or additional medical intervention.

    Product
    Capio SLIM Suture Capturing Device, 1-Pack, UPN: M0068318250. Capio SLIM Suture Capturing Devices are used by physicians to facilitate the consistent placement of sutures during open surgical procedures.
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-2455-2024·2024-08-07

    Clinical Chemistry Analyzer Software Bugs May Cause Erroneous Patient Results

    Beckman Coulter's DxC 500 AU analyzer has two software bugs that may cause erroneous patient results due to misreported calibrations or expired reagents. No injuries or illnesses have been reported.

    Product
    DxC 500 AU Clinical Chemistry Analyzer, REF C63520 is an automated chemistry analyzer that measures analytes in samples, in combination with appropriate reagents, calibrators, quality control (QC) material and other accessories. This system is for in vitro diagnostic use only.
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-2458-2024·2024-08-07

    BD Insyte Autoguard BC IV Catheter Recalled for Tubing Defect and Leakage

    Becton Dickinson has recalled certain BD Insyte Autoguard BC IV catheters due to potential holes in the catheter tubing that could cause leakage during insertion. The recall involves 26,400 units distributed in the United States.

    Product
    BD Insyte Autoguard BC, Shielded IV Catheter with Blood Control Technology, REF: 382533, 20 GA x 1.00 in (1.1x25 mm) 63 mL/min, Rx Only. BD Insyte Autoguard BC Shielded IV Catheter is intended to be inserted into a patient's peripheral vascular system for short term use to sample
    Category
    Medical Device
    Distribution
    1 state
  • HighFDA (Devices)·Z-2451-2024·2024-08-07

    Capio SLIM Suture Capturing Device may fail to catch sutures consistently

    Boston Scientific is recalling some lots of Capio SLIM Suture Capturing Devices because they may not consistently catch sutures during surgery, potentially requiring prolonged procedures or additional medical intervention.

    Product
    Capio SLIM Suture Capturing Device, Box of 5, Outer Package UPN: M0068318261, Inner Package UPN: M0068318260. Capio SLIM Suture Capturing Devices are used by physicians to facilitate the consistent placement of sutures during open surgical procedures.
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-2456-2024·2024-08-07

    Blood Culture Identification Panel May Produce False Positive Results

    The BIOFIRE BCID2 Panel may produce false positive Serratia marcescens results when used with certain culture media bottles, potentially leading to inappropriate changes in patient therapy.

    Product
    BIOFIRE Blood Culture Identification 2 (BCID2) Panel, REF: RFIT-ASY-0147, for FILMARRAY Systems, when used in conjunction with Culture Media Bottles: BACT/ALERT FA PLUS, REF: 410851, and BACT/ALERT PF PLUS, REF: 410853
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-2416-2024·2024-08-07

    SlideView VS200 slide viewer laser shutter malfunction recall

    The SlideView VS200 slide viewer has a broken laser fuse caused by a short circuit from a poorly positioned cable. This prevents the shutter from operating properly, leaving it stuck in an open or closed position.

    Product
    SlideView VS200 slide viewer,
    Category
    Medical Device
    Distribution
    0 states
  • HighFDA (Devices)·Z-2457-2024·2024-08-07

    Arctic Sun Temperature Management System software failure may disable critical safety alarms

    Arctic Sun Temperature Management Systems may fail to reach target temperature and critically fail to alert clinicians. The software defect affects approximately 10,507 units worldwide and risks hypothermia or hyperthermia.

    Product
    Arctic Sun Temperature Management Systems, Name/REF: Arctic Sun STAT, NA/6000-00-00; Arctic Sun STAT Loaner, NA/6000-00-00L; Arctic Sun Stat EU UK/600001; Arctic Sun Stat Taiwan/600006; Arctic Sun Stat Japan/600020; Arctic Sun 5000, NA/5000-00-00; Arctic Sun 5000, NA w/ T
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • ModerateFDA (Devices)·Z-2454-2024·2024-08-07

    SOFT-VU KUMPE Catheters Recalled for Mislabeled Inner Pouch

    Angiodynamics is recalling SOFT-VU KUMPE 5F X 65CM angiographic catheters (595 boxes nationwide) because inner pouch labels may show information for a different catheter model while shelf-box labels are correct.

    Product
    SOFT-VU KUMPE 5F X 65CM -Used for patients in need of angiographic diagnosis. Catalog Number: 10732701
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • ModerateFDA (Devices)·Z-2453-2024·2024-08-07

    SOFT-VU KUMPE Diagnostic Catheter Recalled for Mislabeled Inner Pouch

    Angiodynamics recalls SOFT-VU KUMPE diagnostic catheters because inner pouch labels may show incorrect catheter length information, while the outer box label is accurate. 519 boxes distributed nationwide.

    Product
    SOFT-VU KUMPE 5F X 40CM -Used for patients in need of angiographic diagnosis. Catalog Number: 10732702
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • SevereFDA (Devices)·Z-2409-2024·2024-07-31

    NIM Vital Nerve Monitoring Systems Recalled for False Negative Risk

    Medtronic Xomed is recalling NIM Vital Nerve Monitoring systems because they may fail to detect nerve damage or dysfunction. Approximately 7,804 units were distributed worldwide.

    Product
    NIM Vital Nerve Monitoring systems (all serial numbers manufactured) with NIM Vital System versions: v1.0.0, v1.1.1, v1.3.2, v1.4.3. NIM Vital Nerve Integrity Monitor with below product descriptions. 1. CONSOLE NIM4CM01 NIM 4.0, Model Number: NIM4CM01. 2. CONSOLE NIM4CM
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-2418-2024·2024-07-31

    Wireless Nurse Call Station Recalled for Potential Alarm Communication Failure

    Arial Call Station wireless nurse call systems (models CSK200-1069 and CSK200-1069MR) are recalled due to risk of alarm communication failure and premature battery depletion. 77 units affected across multiple U.S. states.

    Product
    Arial Call Station, Model Numbers CSK200-1069 and CSK200-1069MR, is a wireless nurse call station.
    Category
    Medical Device
    Distribution
    13 states
  • HighFDA (Devices)·Z-2417-2024·2024-07-31

    Call station devices may fail to transmit emergency alarms to staff

    Securitas Healthcare's Arial Call Station (models CSK200 and CSK200MR) may fail to transmit alarms and experience premature battery depletion. The FDA recall affects 1,268 units distributed in the U.S. and Canada.

    Product
    Arial Call Station, Model Numbers CSK200 and CSK200MR, is a wireless, battery-operated call device installed inside resident living areas to allow immediate staff response.
    Category
    Medical Device
    Distribution
    9 states
  • HighFDA (Devices)·Z-2428-2024·2024-07-31

    Beckman Coulter Clinical Reagent Recalled Due to Incorrect Lipemia Check Settings

    Beckman Coulter has recalled 13,565 units of IGM clinical reagent due to incorrect lipemia check settings that may produce false low results or report high results as normal. Affected units were distributed worldwide.

    Product
    Beckman Coulter, REF: OSR61173, IGM 4x14 mL R1, 4x11 mL R2
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-2433-2024·2024-07-31

    Infusion Pump Syringe Compatibility Issue May Cause Overdose or Underdose

    B Braun is recalling Perfusor Space infusion pumps due to syringe compatibility issues with Cardinal Health Monoject syringes. The incompatibility may cause overdose, underdose, therapy delays, and alarm failures.

    Product
    Brand Name: Perfusor Space Product Name: Perfusor Space Infusion Pump System Model/Catalog Number: 8713030U Product Description: Perfusor Space Infusion Pump System, Infusion Pump, 1 each Component: N/A
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-2437-2024·2024-07-31

    Perfusor Space Syringe Pump Recalled for Monoject Syringe Incompatibility Risk

    B Braun's Perfusor Space infusion pump may be incompatible with Cardinal Health Monoject syringes affected by manufacturing changes, creating risks of overdose, underdose, and delayed therapy.

    Product
    Brand Name: Perfusor Space Product Name: Perfusor Space Syringe Pump, Non-Wireless Model/Catalog Number: 8713032U Product Description: Perfusor Space Syringe Pump, Non-Wireless, Infusion Pump, 1 each Component: N/A
    Category
    Medical Device
    Distribution
    Distributed nationwide