Beacon Tip Centimeter Sizing Catheter recalled for cracking marker bands
Cook Incorporated is recalling Beacon Tip Centimeter Sizing Catheters because marker bands may crack or break, which could lead to device fragmentation, prolonged procedures, vessel injury, and in worst-case scenarios, life-threatening harm or death.
The agency’s classification
FDA (Devices): Class I
Class I is a situation in which there is a reasonable probability that the use of or exposure to a violative product will cause serious adverse health consequences or death.
This is the agency’s own classification, taken from its notice. The severity score below is our assessment, not the agency’s. How recall classes work
What this means for you
Our assessment, not the agency’s: severity 4/5 (Severe).
The hazard poses a serious risk of illness, injury, or structural failure. Stop using the product immediately and contact the manufacturer for a refund or repair.
Our severity reasoning: The FDA classified this as Class I, which per the rubric requires a minimum score of 4. The source text explicitly describes worst-case scenarios of life-threatening harm or death, and the hazard involves a medical device used in invasive cardiovascular procedures with potential for vessel injury.
Plain-English summary
Cook Incorporated is recalling Beacon Tip Centimeter Sizing Catheters (model numbers G31213, G31214, G31216, G31220, and G31221) because the marker bands may be at increased risk of cracking or breaking. These are single-use, sterile catheters with gold marker bands used in angiographic procedures, available in 70 cm and 100 cm lengths with various tip configurations.
Potential adverse events from marker band cracking or breakage include increased procedural time, harms associated with device fragmentation or separation, and vessel injury. In a worst-case scenario, device fragmentation and separation may cause life-threatening harm or death.
Approximately 4,245 units were distributed nationwide in the US and internationally to Brunei, Brazil, Canada, China, Colombia, Costa Rica, Hong Kong, Israel, India, Japan, South Korea, Macau, New Zealand, Singapore, El Salvador, Taiwan, and Uruguay. Specific lot numbers for each model are identified in the recall.
Healthcare providers and patients who have received these catheters should consult with Cook Incorporated or their healthcare provider regarding the recall and appropriate next steps.
The recalled product
- Product
- Beacon Tip Centimeter Sizing Catheter RPN/GPN NR5.0-35-70-P-10S-0-CSC-20 G31213 NR5.0-35-100-P-10S-0-CSC-20 G31214 NR5.0-35-100-P-10S-PIG-CSC-20 G31216 NR5.0-35-70-P-10S-VCF-CSC-20 G31220 NR5.0-35-100-P-10S-VCF-CSC-20 G31221 Sizing catheters are single-use, sterile…
- Manufacturer
- Cook Incorporated
- Affected units
- 4,245
Distribution
Distributed nationwide across the United States.
Part of a larger recall action
This product is one of 3 recalled by Cook Incorporated under a single FDA (Devices) recall action. The agency files each product separately, so the size of the action is not visible from this page alone.
See all 3 products in this recall →Related recalls
Same manufacturer · Cook Incorporated
See all →- CriticalAurous Centimeter Sizing Catheter Recalled for Marker Band Cracking Risk
FDA (Devices) · 2026-05-13
- CriticalCook Centimeter Sizing Catheter recall due to marker band cracking
FDA (Devices) · 2026-05-13
- HighCook Medical tracheostomy introducer sets labeled with overstated expiration dates
FDA (Devices) · 2026-04-15
- HighCook Medical CTO-18 wire guides labeled with overstated expiration dates
FDA (Devices) · 2026-04-15
- HighCook Medical pneumothorax tray labeled with an overstated expiration date
FDA (Devices) · 2026-04-15
Same hazard · device fracture
See all →- SevereBoston Scientific Recalls Imager II Angiographic Catheters Due to Tip Fracture Risk
FDA (Devices) · 2026-09-30
- HighHowmedica Recalls Orthopaedic Implants Due to Potential UHMWPE Oxidation
FDA (Devices) · 2026-09-23
- SevereHowmedica Recalls Orthopaedic Implants Due to Potential UHMWPE Oxidation
FDA (Devices) · 2026-09-23
- HighHowmedica Recalls Orthopaedic Implant Components Due to Material Aging
FDA (Devices) · 2026-09-23
- HighHowmedica Recalls Orthopaedic Implants Due to Aged Polyethylene Material
FDA (Devices) · 2026-09-23
Similar recalls
- SevereCook Inc. Recalls Beacon Tip Catheters Due to Tip Fracture Risk
FDA (Devices) · 2016-08-31
- SevereCook Inc. Recalls Slip-Cath Beacon Tip Catheters Due to Tip Fracture
FDA (Devices) · 2016-08-31
- SevereCook Beacon Tip Angiographic Catheters Recalled for Tip Separation Risk
FDA (Devices) · 2025-06-25
- SevereCook Inc. Recalls Beacon Tip Torcon NB Advantage Catheters Due to Tip Fracture
FDA (Devices) · 2016-08-31
- SevereCook Inc. Recalls Selective Salpingography Catheters Due to Tip Fracture Risk
FDA (Devices) · 2016-08-31
- SevereCook Inc. Recalls Slip-Cath Beacon Tip Catheters Due to Tip Separation
FDA (Devices) · 2015-08-12