bioMerieux VITEK 2 Gram Positive Test Kits Recalled for MRSA Detection Issues
bioMerieux is recalling VITEK 2 Gram Positive Test Kits because customer reports indicated an increased rate of non-detected MRSA.
The agency’s classification
FDA (Devices): Class I
Class I is a situation in which there is a reasonable probability that the use of or exposure to a violative product will cause serious adverse health consequences or death.
This is the agency’s own classification, taken from its notice. The severity score below is our assessment, not the agency’s. How recall classes work
What this means for you
Our assessment, not the agency’s: severity 4/5 (Severe).
The hazard poses a serious risk of illness, injury, or structural failure. Stop using the product immediately and contact the manufacturer for a refund or repair.
Our severity reasoning: The agency classified this recall as Class I — a situation in which there is a reasonable probability that the use of or exposure to a violative product will cause serious adverse health consequences or death. On that classification alone this is scored Severe; a score of Critical is reserved for recalls where a death or serious injury was reported. The agency gives the reason as: "Customer reports indicated an increase in the rate of non-detected MRSA in association with the VITEK 2 AST-P631 product."
Plain-English summary
bioMerieux, Inc. is recalling specific VITEK 2 Gram Positive Test Kits, including the AST-P631, AST-GP67, AST-GP71, AST-GP75, and AST-GP78 models. These products are components for an automated in vitro diagnostic system used for the identification and antimicrobial susceptibility testing of bacteria and yeast.
The recall was initiated after customer reports indicated an increase in the rate of non-detected MRSA (Methicillin-resistant Staphylococcus aureus) in association with the VITEK 2 AST-P631 product. The affected units include multiple Gram Positive Cards with Oxacillin and Cefotoxin across all lots for the listed test kits.
The affected products were distributed worldwide, including in the United States, Puerto Rico, Bermuda, Brazil, Canada, China, and numerous other countries. Healthcare facilities and laboratories that use these diagnostic systems should stop using the affected test kits and contact bioMerieux for further instructions.
The recalled product
- Product
- bioMerieux VITEK 2 Gram Positive Cefoxitin Screen, VITEK 2 Gram Positive AST for Oxacillin, VITEK SYSTEMS General Susceptibility Card Cefotoxin and AST for Oxacillin bioMerieux VITEK 2 AST-P631 bioMerieux VITEK 2 AST-GP67 Test Kit bioMerieux VITEK 2 AST-GP71 Test Kit…
- Manufacturer
- bioMerieux, Inc.
- Affected units
- 13,961,320
Is your product affected?
Check your packaging against the codes below. If any of them match, the product is part of this recall.
Lot numbers (13)
- VITEK 2 AST-P631
- Ref # 414961 UDI 03573026426804 OXSF
- OXSF
- Ref: # 22226 UDI 03573026242060
- OX
- multiple Gram Positive Cards
- all lots of cards with Oxacillin and Cefotoxin. VITEK 2 AST-GP71 Test Kit
- Ref: #410750 UDI 03573026355173
- all lots of cards with Oxacillin and Cefotoxin. VITEK 2 AST-GP75 Test Kit
- Ref: #415670 UDI 03573026437183
- all lots of cards with Oxacillin and Cefotoxin. VITEK 2 AST-GP78 Test Kit
- Ref: #421051 UDI 03573026553982
- all lots of cards with Oxacillin and Cefotoxin.
Distribution
Related recalls
Same manufacturer · bioMerieux, Inc.
See all →- HighBiomerieux VITEK 2 AST-N461 Test Kit Recalled for Card Pouch Damage
FDA (Devices) · 2026-09-30
- HighVITEK 2 AST-N447 Test Kit Potentially Returns False Resistant Results
FDA (Devices) · 2026-07-15
- HighSusceptibility test cards recalled over false resistant antibiotic results
FDA (Devices) · 2026-02-11
- HighAntimicrobial susceptibility test cards recalled over false colistin resistance results
FDA (Devices) · 2025-10-08
- SevereVITEK 2 AST Test Kits: Ceftriaxone Concentration Error Risk
FDA (Devices) · 2024-05-01
Same hazard · mis diagnosis
See all →- SevereVolcano Corp Recalls Philips IntraSight Plus Systems Due to Data Error
FDA (Devices) · 2026-10-07
- HighBioporto ProNephro AKI Reagent Kit Recalled for Non-Validated Rework Process
FDA (Devices) · 2026-08-19
- HighFilmArray NGDS Warrior Panel Diagnostic Test Recalled Due to Strain Mismatch
FDA (Devices) · 2026-08-12
- HighDASH Rapid PCR Instrument Recalled Due to Software Issue Causing False Co-infection Results
FDA (Devices) · 2026-08-12
- HighDiagnostic Genetic Probe Recalled for Unexpected Chromosome Signals
FDA (Devices) · 2021-08-04
Similar recalls
- HighbioMerieux Recalls VITEK 2 Gram Positive Susceptibility Test Cards
FDA (Devices) · 2016-05-11
- SevereBiomerieux Recalls Vitek 2 Gram Positive Susceptibility Test Cards for Labeling Error
FDA (Devices) · 2016-05-18
- SevereBiomerieux Vitek 2 Gram Positive Susceptibility Test Cards Recalled for Labeling Error
FDA (Devices) · 2016-05-18
- ModerateBiomerieux Recalls VITEK 2 Gram-positive Susceptibility Cards Due to False Positives
FDA (Devices) · 2020-04-29
- HighbioMerieux Recalls Vitek 2 Test Kits Due to False ESBL Results
FDA (Devices) · 2019-04-24
- SevereBiomerieux Vitek 2 Gram Positive Test Cards Recalled for Labeling Error
FDA (Devices) · 2016-05-18