Maquet CS 100i Intra-Aortic Balloon Pump Recalled Due to Device Failure
Maquet Datascope is recalling 12,319 CS 100i Intra-Aortic Balloon Pumps because electrical failures prevented the device from initiating therapy, which has been associated with a patient death.
The agency’s classification
FDA (Devices): Class I
Class I is a situation in which there is a reasonable probability that the use of or exposure to a violative product will cause serious adverse health consequences or death.
This is the agency’s own classification, taken from its notice. The severity score below is our assessment, not the agency’s. How recall classes work
What this means for you
Our assessment, not the agency’s: severity 5/5 (Critical).
Highest-severity recall. The hazard can cause serious injury or death. Stop using the product now and follow the agency's instructions.
Our severity reasoning: The source text explicitly states that the device failure has been associated with a patient death. Per the rubric, if reported deaths exist anywhere in the source text, the score is 5 regardless of other factors.
Plain-English summary
Maquet Datascope Corp - Cardiac Assist Division is recalling 12,319 units of the CS 100i Intra-Aortic Balloon Pump. The recall covers devices distributed worldwide in over 100 countries, including nationwide distribution in the United States.
The recall was initiated because the devices failed to pump due to an electrical test failure code #58 (power up vent test fail), maintenance code #3, and an autofill failure. The source text states that this failure to initiate therapy has been associated with a patient death.
Healthcare providers and patients should stop using the affected devices immediately and contact the manufacturer for further instructions. The specific product codes associated with this recall are 0998-UC-0446HXX and 0998-UC-0479HXX.
The recalled product
- Product
- CS 100i Intra-Aortic Balloon Pump
- Manufacturer
- Maquet Datascope Corp - Cardiac Assist Division
- Hazard
- Affected units
- 12,319
Is your product affected?
Check your packaging against the codes below. If any of them match, the product is part of this recall.
Lot numbers (1)
- 0998-UC-0446HXX 0998-UC-0479HXX
Distribution
Distributed nationwide across the United States.
Part of a larger recall action
This product is one of 3 recalled by Maquet Datascope Corp - Cardiac Assist Division under a single FDA (Devices) recall action. The agency files each product separately, so the size of the action is not visible from this page alone.
See all 3 products in this recall →Related recalls
Same manufacturer · Maquet Datascope Corp - Cardiac Assist Division
See all →- SevereMaquet Cardiosave Rescue IABP Recalled for High Altitude Therapy Interruption
FDA (Devices) · 2018-11-07
- SevereCardiosave Hybrid IABP Recalled for Potential Therapy Interruption at High Altitudes
FDA (Devices) · 2018-11-07
- SevereMaquet Datascope Recalls CARDIOSAVE Hybrid Intra-Aortic Balloon Pumps
FDA (Devices) · 2018-06-06
- SevereMaquet CS 100 Intra-Aortic Balloon Pump Recalled for Electronic Defects
FDA (Devices) · 2017-09-13
- SevereMaquet CS 100i Intra-Aortic Balloon Pump Recalled for Fluid Ingress
FDA (Devices) · 2017-09-13
Same hazard · device failure
See all →- HighStryker Recalls XCEL Bladeless Trocar Due to Cracking Risk
FDA (Devices) · 2026-09-23
- SevereMedline Neonatal Heated Wire Breathing Circuits Recalled for Thermal Damage Risk
FDA (Devices) · 2026-09-02
- HighBoston Scientific Recalls OverStitch Suture Cinch Devices for Deployment Failure
FDA (Devices) · 2026-08-26
- HighCOAPTIUM Connect Nerve Repair Device Relabeled Due to Unsubstantiated Shelf Life
FDA (Devices) · 2026-08-12
- HighCOAPTIUM CONNECT Peripheral Nerve Repair Device Recall for Shelf Life Labeling
FDA (Devices) · 2026-08-12
Similar recalls
- SevereMaquet CS 100i Intra-Aortic Balloon Pump Recalled for Fluid Ingress
FDA (Devices) · 2017-09-13
- SevereMaquet CS 100 Intra-Aortic Balloon Pump Recalled for Electronic Defects
FDA (Devices) · 2017-09-13
- SevereMaquet CS 300 Intra-Aortic Balloon Pump Recalled for Fluid Ingress Risk
FDA (Devices) · 2017-09-13
- SevereDatascope CS100 Intra-Aortic Balloon Pump Recalled for Battery Maintenance Issues
FDA (Devices) · 2019-07-24
- SevereDatascope CS300 Intra-Aortic Balloon Pump Recalled for Battery Maintenance Issues
FDA (Devices) · 2019-07-24
- SevereMaquet Datascope Intra-Aortic Balloon Pump Recall for Fan Assembly Failure
FDA (Devices) · 2014-05-28