Beckman Coulter Access Total T3 Immunoassay Recalled for Biotin Interference
Beckman Coulter is recalling Access Total T3 immunoassays because they are susceptible to biotin interference, which may affect test results.
What this means for you
Minor labeling or packaging issue, or a voluntary precautionary recall. Risk of harm is low but real — follow the manufacturer's instructions.
Our severity reasoning: The recall is classified as FDA Class II, but the source text does not report any hospitalizations, injuries, or deaths. The hazard is a susceptibility to interference rather than a direct physical injury or high-risk pathogen, fitting the criteria for a moderate severity voluntary precautionary recall.
Plain-English summary
Beckman Coulter Inc. is recalling the Access Total T3 immunoassay (Catalog No. 33830, A34435H). The recall involves 313,158 units total, with 43,725 units distributed in the United States. The product was distributed to numerous countries worldwide, including the United States, Canada, and various nations in Europe, Asia, Africa, and the Americas.
The recall was initiated because the Access immunoassays are susceptible to biotin interference. This susceptibility may impact the accuracy of the test results. The source text does not specify reported illnesses, injuries, or deaths associated with this recall.
Healthcare providers and laboratories using this product should stop using the affected units and contact Beckman Coulter for further instructions. Consumers should consult their healthcare providers regarding any recent test results that may have been affected by this issue.
The recalled product
- Product
- Access Total T3, A34435H, Catalog No. 33830
- Manufacturer
- Beckman Coulter Inc.
Is your product affected?
Check your packaging against the codes below. If any of them match, the product is part of this recall.
Lot numbers (1)
- all lots
Distribution
Part of a larger recall action
This product is one of 4 recalled by Beckman Coulter Inc. under a single FDA (Devices) recall action. The agency files each product separately, so the size of the action is not visible from this page alone.
See all 4 products in this recall →Related recalls
Same manufacturer · Beckman Coulter Inc.
See all →- HighBeckman Coulter DxC 500 AU Analyzer Software Defect Recall
FDA (Devices) · 2026-08-26
- HighBeckman Coulter DxC 500i Analyzer Modules Recalled for Software Defect
FDA (Devices) · 2026-08-26
- High
- SevereBeckman Coulter MicroScan Panel Update DVD Kits Recalled
FDA (Devices) · 2026-08-12
- HighBeckman Coulter Access Total T4 Calibrator Recalled for Inaccurate Results
FDA (Devices) · 2026-04-29
Same hazard · biotin interference
See all →- HighBeckman Coulter Access Free T4 Thyroxine Assay Recalled for Biotin Interference
FDA (Devices) · 2020-06-10
- HighBeckman Coulter Access Thyroglobulin Antibody II Recalled for Biotin Interference
FDA (Devices) · 2020-06-10
- ModerateBeckman Coulter Access Total T3 Assay Recalled for Biotin Interference
FDA (Devices) · 2020-06-10
- ModerateBeckman Coulter Free T3 Immunoassay Recalled for Biotin Interference
FDA (Devices) · 2020-06-10
- ModerateBeckman Coulter Access GI Monitor Cancer Antigen 19-9 Recalled for Biotin Interference
FDA (Devices) · 2020-06-10