Bio-Rad PR4100 Microplate Reader Recalled for Power Supply Connector Risk
Bio-Rad is recalling 153 PR4100 Microplate Readers because the power supply connector can fit incompatible components, risking irreparable damage to the system.
The agency’s classification
FDA (Devices): Class II
Class II is a situation in which use of or exposure to a violative product may cause temporary or medically reversible adverse health consequences, or where the probability of serious adverse health consequences is remote.
This is the agency’s own classification, taken from its notice. The severity score below is our assessment, not the agency’s. How recall classes work
What this means for you
Our assessment, not the agency’s: severity 2/5 (Moderate).
Minor labeling or packaging issue, or a voluntary precautionary recall. Risk of harm is low but real — follow the manufacturer's instructions.
Our severity reasoning: The recall is classified as FDA Class II. The hazard involves equipment damage (irreparable damage to the system) rather than direct physical injury to patients or users, and no injuries or illnesses are reported in the source text.
Plain-English summary
Bio-Rad Laboratories Inc is recalling 153 units of the PR4100 Microplate Reader US, Catalog Number 45591587. The product is a 96-well absorbance reader intended for in vitro diagnostic analysis of samples from the human body.
The recall was initiated because the connector of the power supply was capable of fitting other components provided with the system. The voltage output of the reader power supply is incompatible with these other components. If incorrectly connected, this incompatibility could result in irreparable damage to the system.
The affected units were distributed nationwide in the United States, including in the states of AL, AR, AZ, CA, CO, CT, FL, GA, HI, IL, IN, KS, LA, MA, MD, MI, MN, MO, MS, NC, NJ, NY, OR, PA, SC, TN, TX, VA, VT, WA, WI, and Puerto Rico. Users should check their serial numbers against the list provided in the official recall notice to determine if their unit is affected.
The recalled product
- Product
- PR4100 Microplate Reader US, Catalog Number 45591587 Product Usage: The PR4100 Microplate Reader is a 96 well absorbance reader for the measurement of light absorbance (optical density-OD) of liquid media. The instrument is intended to be used primarily for the in vitro…
- Manufacturer
- Bio-Rad Laboratories Inc
- Affected units
- 153
Is your product affected?
Check your packaging against the codes below. If any of them match, the product is part of this recall.
Lot numbers (20)
- UDI: 00847865014151 Serial Numbers: 140600072
- 160900991
- 161001953
- 1112001938
- 1112001941
- 1112004075
- 1112004079
- 1112004083
- 1112004084
- 1112004086
- 1112004088
- 1112004093
- 1201003822
- 1201003826
- 1201003827
- 1207001007
- 1207001015
- 1207001017
- 1207001018
- 1208001583
Distribution
Related recalls
Same manufacturer · Bio-Rad Laboratories Inc
See all →- HighToxoplasma IgM assay kits recalled over risk of false positive results
FDA (Devices) · 2025-09-10
- HighAPLS IgM Diagnostic Reagent Packs Recalled Due to Incorrect Conjugate
FDA (Devices) · 2023-05-17
- HighBioPlex 2200 Anti-CCP Reagent Kit recalled due to reduced test specificity
FDA (Devices) · 2023-04-19
- HighBioPlex 2200 Anti-CCP Reagent Pack Recalled for Reduced Diagnostic Specificity
FDA (Devices) · 2023-04-19
- HighCMV IgM EIA Diagnostic Test Recalled for Increased False Positives
FDA (Devices) · 2022-12-07
Same hazard · equipment damage
See all →- HighOlympus Cystoscope Outer Sheath Recalled Due to Potential Device Tip Damage
FDA (Devices) · 2026-03-04
- ModerateIncisive CT systems recalled because tube heat exchanger screws may not be tightened
FDA (Devices) · 2025-12-10
- ModerateCT scanners recalled because tube heat exchanger screws may not be tightened
FDA (Devices) · 2025-12-10
- HighPower Express PE Centrifuge adhesive joint failure may cause component detachment
FDA (Devices) · 2024-09-11
- HighBeckman Coulter DxI 9000 Analyzer Recalled for Leaking Wash Buffer Tank
FDA (Devices) · 2024-09-11
Similar recalls
- SevereTecan Infinite M1000 Microplate Reader Recalled for Battery Overheating Risk
FDA (Devices) · 2016-08-31
- HighTecan Infinite M1000 Microplate Reader Recalled for Firmware Bug
FDA (Devices) · 2015-07-08
- ModerateTecan Infinite M1000 PRO Microplate Reader Recalled for Firmware Bug
FDA (Devices) · 2015-07-08
- SeverePhilips PCR Compano Basic Reader Recalled for Power Supply Fire Hazard
FDA (Devices) · 2016-09-07
- SevereTecan Infinite M1000 PRO Microplate Reader Recalled for Battery Overheating Risk
FDA (Devices) · 2016-08-31
- ModerateTecan Infinite 200 PRO Microplate Reader Recalled for Firmware Bug
FDA (Devices) · 2015-07-08