Varian Eclipse Treatment Planning System Recalled for Dose Calculation Error
Varian is recalling Eclipse Treatment Planning Systems because a clock synchronization issue can cause incorrect dose calculations, potentially leading to under- or over-dosing of patients.
What this means for you
Illness, injury, or structural failure has been reported. Stop using the product immediately and contact the manufacturer for a refund or repair.
Our severity reasoning: The recall involves a medical device defect that can lead to significant patient harm (under- or over-dosing) during cancer treatment, fitting the criteria for a Severe recall due to the potential for serious injury.
Plain-English summary
Varian Medical Systems, Inc. is recalling the Eclipse Treatment Planning System (Model H48, versions 8.9 and 10.0) due to a software anomaly. The issue occurs when the clock on the Eclipse Client Workstation is not synchronized with the Database/System Server. In these instances, an outdated assigned Hounsfield Unit (HU) value may be used for dose calculation.
The use of these incorrect values can lead to under-dosing or over-dosing of patients during radiotherapy treatments. The recall affects 6,087 units distributed worldwide, including nationwide in the United States and internationally.
Healthcare facilities using the affected systems should contact Varian Medical Systems for instructions on how to resolve the synchronization issue and ensure accurate dose calculations.
The recalled product
- Product
- Varian brand Eclipse Treatment Planning System, Model Number: H48; Eclipse 8.9 and 10.0 Reference/FSCA Identifier: CP-08922; Product is manufactured and distributed by Varian Medical Systems Inc., Palo Alto, CA The Eclipse Treatment Planning System (Eclipse TPS) is used to p
- Manufacturer
- Varian Medical Systems, Inc. Oncology Systems
- Affected units
- 6,087
Is your product affected?
Check your packaging against the codes below. If any of them match, the product is part of this recall.
Lot numbers (20)
- CP-08922: LIST OF SERIAL NUMBERS: H480001
- H480130
- H480301
- H480446
- H480623
- H480795
- H480976
- H481088
- H481222
- H481382
- H481544
- H481777
- H481931
- H480002
- H480131
- H480302
- H480448
- H480624
- H480798
- H480981
Distribution
Distributed nationwide across the United States.
Related recalls
Same manufacturer · Varian Medical Systems, Inc. Oncology Systems
See all →- HighVarian High Energy Clinacs Radiation Treatment System Recall for Beam Symmetry
FDA (Devices) · 2013-05-22
- ModerateVarian RPM Respiratory Gating System Recall for Waveform Anomaly
FDA (Devices) · 2013-05-01
- HighVarian Eclipse Treatment Planning System Recalled for Dose Calculation Error
FDA (Devices) · 2013-02-27
- HighVarian ARIA Radiation Oncology System Recall for DICOM Export Anomaly
FDA (Devices) · 2012-12-26
- HighVarian ARIA Radiation Oncology Software Recall Due to Treatment Plan Error
FDA (Devices) · 2012-11-07
Same hazard · incorrect dose calculation
See all →- HighRayStation Radiation Therapy Software Recalled for Incorrect Dose Calculation
FDA (Devices) · 2023-07-05
- HighRayStation software recall due to incorrect total dose display
FDA (Devices) · 2021-05-26
- HighRayStation Treatment Planning Software Recalled for Dose Calculation Errors
FDA (Devices) · 2020-02-05
- HighElekta Monaco Radiation Treatment Planning System Software Recall
FDA (Devices) · 2019-12-18
- HighElekta Monaco RTP System Recalled for Potential Dose Calculation Errors
FDA (Devices) · 2019-10-02