Medline pelviscopy packs contain recalled Mastisol liquid adhesive
The Medline SMSC Pelviscopy Pack-LF is being recalled because it contains a recalled lot of Ferndale Laboratories Mastisol Liquid Adhesive whose butyrate tubing may crack.
The agency’s classification
FDA (Devices): Class II
Class II is a situation in which use of or exposure to a violative product may cause temporary or medically reversible adverse health consequences, or where the probability of serious adverse health consequences is remote.
This is the agency’s own classification, taken from its notice. The severity score below is our assessment, not the agency’s. How recall classes work
What this means for you
Our assessment, not the agency’s: severity 2/5 (Moderate).
Minor labeling or packaging issue, or a voluntary precautionary recall. Risk of harm is low but real — follow the manufacturer's instructions.
Our severity reasoning: FDA Class II recall where the described consequence is that a component becomes unusable rather than any patient hazard, which fits the Moderate band.
Plain-English summary
The Medline SMSC Pelviscopy Pack-LF, SKU DYNJ0161619M, UDI-DI 10198459028274, lot 25LMI820, is being recalled. The action covers 49,654 kits in total across the connected Medline Mastisol recalls.
The stated reason is that the kits contain a recalled lot of Ferndale Laboratories Mastisol Liquid Adhesive. According to the notice, the butyrate tubing may crack during actuation and render the product unusable for application. The FDA classified the action as Class II. The source does not describe any patient hazard beyond the product being unusable.
Distribution was nationwide in the United States, with distribution outside the United States pending.
Facilities should identify affected stock using the kit number, UDI-DI and lot number above and follow the instructions issued by Medline.
The recalled product
- Product
- Medline convenience kit: SMSC-PELVISCOPY PACK-LF DYNJ0161619M
- Manufacturer
- Medline Industries, LP
- Affected units
- 49,654
Is your product affected?
Check your packaging against the codes below. If any of them match, the product is part of this recall.
Lot numbers (1)
- UDI-DI 10198459028274 Lot 25LMI820
Distribution
Distributed nationwide across the United States.
Part of a larger recall action
This product is one of 34 recalled by Medline Industries, LP under a single FDA (Devices) recall action. The agency files each product separately, so the size of the action is not visible from this page alone.
See all 34 products in this recall →Related recalls
Same manufacturer · Medline Industries, LP
See all →- SevereMedline Bariatric Rollator Recalled Due to Front Wheel Detachment Risk
FDA (Devices) · 2026-10-14
- SevereMedline Bariatric Rollator Recalled Due to Front Wheel Assembly Disconnect Risk
FDA (Devices) · 2026-10-14
- SevereMedline Homecare Beds Recalled for Fire Hazard in Hand Control Pendant
FDA (Devices) · 2026-10-07
- SevereMedline Homecare Bed Hand Pendants Recalled for Fire Hazard
FDA (Devices) · 2026-10-07
- HighMedline Instrument Channel Brush Kits Recalled for Sharp Edges
FDA (Devices) · 2026-09-30
Same hazard · component defect
See all →- ModerateMedline EBSI dressing change kits contain recalled Mastisol liquid adhesive
FDA (Devices) · 2026-06-24
- ModerateMedline femoral block trays contain recalled Mastisol liquid adhesive
FDA (Devices) · 2026-06-24
- ModerateMedline breast hernia port kits contain recalled Mastisol liquid adhesive
FDA (Devices) · 2026-06-24
- ModerateCenturion dialysis and dressing change kits contain recalled Mastisol adhesive
FDA (Devices) · 2026-06-24
- ModerateMedline EBSI kits contain recalled Mastisol liquid adhesive
FDA (Devices) · 2026-06-24
Similar recalls
- ModerateLaparoscopy procedure packs recalled over cracking MASTISOL adhesive tubing inside
FDA (Devices) · 2026-02-04
- ModerateGynecology and prostate packs recalled over cracking adhesive tubing
FDA (Devices) · 2026-02-04
- ModerateMaxillofacial and plastics packs recalled over cracking MASTISOL adhesive tubing
FDA (Devices) · 2026-02-04
- ModerateCircumcision packs recalled over MASTISOL adhesive tubing that can crack
FDA (Devices) · 2026-02-04
- ModerateBreast biopsy and dermatology packs recalled over cracking adhesive tubing
FDA (Devices) · 2026-02-04
- ModerateC-section packs recalled over MASTISOL adhesive tubing that can crack
FDA (Devices) · 2026-02-04