Biomerieux Recalls VITEK 2 Gram Negative Susceptibility Cards Due to Moisture Risk
Biomerieux Inc is recalling VITEK 2 Gram Negative Susceptibility cards because compromised container integrity may expose the product to moisture, potentially causing antibiotic degradation.
The agency’s classification
FDA (Devices): Class II
Class II is a situation in which use of or exposure to a violative product may cause temporary or medically reversible adverse health consequences, or where the probability of serious adverse health consequences is remote.
This is the agency’s own classification, taken from its notice. The severity score below is our assessment, not the agency’s. How recall classes work
What this means for you
Our assessment, not the agency’s: severity 2/5 (Moderate).
Minor labeling or packaging issue, or a voluntary precautionary recall. Risk of harm is low but real — follow the manufacturer's instructions.
Our severity reasoning: The recall is classified as FDA Class II, but the hazard is limited to potential loss of potency due to moisture exposure without reports of serious injury or death, fitting the criteria for a moderate severity precautionary recall.
Plain-English summary
Biomerieux Inc is recalling 18,361 cartons of VITEK 2 Gram Negative Susceptibility cards (AST-GN70), REF 413401. The recall was initiated because the integrity of the product container may be compromised, which could expose the product to moisture. This exposure may lead to antibiotic degradation and a loss of potency.
The affected product was distributed nationwide in the United States, including to military and government sites. It was also distributed to numerous international locations, including Algeria, Angola, Argentina, Australia, Austria, Bahrain, Bangladesh, Belarus, Belgium, Bosnia - Herzegovina, Brazil, Brunei, Bulgaria, Burkina Faso, Cambodia, Cameroun, Canada, Chad, Chile, China, Colombia, Congo, Costa Rica, Croatia, Czech Republic, Denmark, Dominican Republic, Dutch Antilles, Ecuador, Egypt, El Salvador, Estonia, Finland, France, Gabon, Gambia, Georgia, Germany, Greece, Grenada, Guatemala, Guinea, Haiti, Honduras, Hungary, India, Indonesia, Iraq, Israel, Italy, Ivory Coast, Japan, Jordan, Kazakhstan, Kenya, Uganda, Kuwait, Laos, Latvia, Lebanon, Lithuania, Malaysia, Mali, Malta, Mexico, Mongolia, Morocco, Mozambique, Myanmar, Netherlands, Nigeria, Norway, Oman, Pakistan, Palestine, Panama, Paraguay, Peru, Philippines, Poland, Portugal, Romania, Russia, Saudi Arabia, Senegal, Serbia, Singapore, Slovenia, South Africa, South Korea, Spain, Sri Lanka, Sweden, Switzerland, Syria, Thailand, Tunisia, Turkey, Ukraine, United Arab Emirates, United Kingdom, Uruguay, Venezuela, Vietnam, and Zimbabwe.
The recall includes specific lot codes: 5900175203 (20/May/18), 5900243103 (27/Jul/18), 5900243123 (27/Jul/18), 5900272203 (25/Aug/18), 5900286103 (08/Sep/18), and 5900292103 (14/Sep/18). Users should check their inventory for these specific reference numbers and lot codes.
The recalled product
- Product
- VITEK¿ 2 Gram Negative Susceptibility card (AST-GN70), REF 413401, 20 cards per carton.
- Manufacturer
- Biomerieux Inc
- Affected units
- 18,361
Distribution
Part of a larger recall action
This product is one of 187 recalled by Biomerieux Inc under a single FDA (Devices) recall action. The agency files each product separately, so the size of the action is not visible from this page alone.
See all 187 products in this recall →Related recalls
Same manufacturer · Biomerieux Inc
See all →- HighBiomerieux VITEK 2 AST-N461 Test Kit Recalled for Card Pouch Damage
FDA (Devices) · 2026-09-30
- HighVITEK 2 AST-N447 Test Kit Potentially Returns False Resistant Results
FDA (Devices) · 2026-07-15
- HighSusceptibility test cards recalled over false resistant antibiotic results
FDA (Devices) · 2026-02-11
- HighAntimicrobial susceptibility test cards recalled over false colistin resistance results
FDA (Devices) · 2025-10-08
- SevereVITEK 2 AST Test Kits: Ceftriaxone Concentration Error Risk
FDA (Devices) · 2024-05-01
Same hazard · moisture exposure
See all →- HighEchelon-14 Micro Catheter Recall Due to Moisture Exposure During Transport
FDA (Devices) · 2022-08-24
- ModerateTeleflex Recalls RUSCH LASERTUBE Devices Due to Laser Guard Foil Separation
FDA (Devices) · 2020-07-15
- ModeratebioMerieux Recalls VITEK 2 AST-N371 Cards Due to Compromised Seals
FDA (Devices) · 2020-04-01
- ModerateAmerican Health Packaging Recalls Desmopressin Acetate Tablets Due to Missing Desiccant
FDA (Drugs) · 2020-02-26
- ModeratePfizer Recalls Caduet Tablets Due to Defective Bottle Seals
FDA (Drugs) · 2020-02-19
Similar recalls
- ModeratebioMerieux Recalls VITEK 2 AST-N371 Cards Due to Compromised Seals
FDA (Devices) · 2020-04-01
- HighbioMerieux Recalls VITEK 2 Gram Positive Susceptibility Test Cards
FDA (Devices) · 2016-05-11
- SevereBiomerieux Recalls Vitek 2 Gram Positive Susceptibility Test Cards for Labeling Error
FDA (Devices) · 2016-05-18
- HighSusceptibility test cards recalled over false resistant antibiotic results
FDA (Devices) · 2026-02-11
- ModerateBiomerieux Recalls VITEK 2 Gram-positive Susceptibility Cards Due to False Positives
FDA (Devices) · 2020-04-29
- LowBiomerieux Recalls Vitek 2 Gram Positive Susceptibility Cards Due to Labeling Error
FDA (Devices) · 2015-04-15