Steris Recalls Verify Bowie-Dick Test Cards Due to Performance Issues
Steris Corporation is recalling Verify Bowie-Dick Test Cards because they may produce false fail results, potentially indicating sterilizer issues when performance is actually acceptable.
What this means for you
Minor labeling or packaging issue, or a voluntary precautionary recall. Risk of harm is low but real — follow the manufacturer's instructions.
Our severity reasoning: The recall is classified as FDA Class II. The hazard involves false negative results (false fails) in sterilization testing, which is a significant quality issue but does not involve direct physical injury or death to patients, fitting the criteria for a moderate severity recall involving performance defects.
Plain-English summary
Steris Corporation is recalling 629 Verify Bowie-Dick Test Cards, Model #S3098, due to performance issues. The affected cards are identified by serial numbers H20013, H20014, and H20016. These test cards are used to evaluate the effectiveness of air removal from sterilizer chambers during prevacuum pulse steam sterilizer cycles.
The recall was initiated because the test cards are not performing to product specifications. Specifically, the cards have resulted in false fail results when sterilizer performance is within acceptable ranges. This means the cards may indicate that a sterilization cycle was inadequate even when the sterilizer was functioning correctly.
The affected products were distributed nationwide in the United States and in Canada. Users of these test cards should check their inventory for the specific model and serial numbers listed in this recall. If you have these cards, you should stop using them and contact Steris Corporation for further instructions on how to return or dispose of the product.
The recalled product
- Product
- Verify¿ Bowie-Dick Test Card, Model #S3098, Steris Biological Operations, 9325 Pinecone Drive, Mentor, OH 44060 The Verify Bowie-Dick Test Card consists of a card printed on one side with bars of chemical indicator ink. On the opposite side of the card, space is provided for c
- Manufacturer
- Steris Corporation
- Affected units
- 629
Is your product affected?
Check your packaging against the codes below. If any of them match, the product is part of this recall.
Lot numbers (4)
- Model #S3098
- Serial #"s: H20013
- H20014
- and H20016.
Distribution
Related recalls
Same manufacturer · Steris Corporation
See all →- ModerateHarmonyAIR A-Series surgical light heads may show paint delamination
FDA (Devices) · 2026-06-24
- ModerateHarmonyAIR A-Series surgical light intensity control buttons may detach
FDA (Devices) · 2026-06-24
- ModerateALYON surgical light head intensity control buttons may detach
FDA (Devices) · 2026-06-24
- ModerateDisplay adapters recalled after the wrong configuration file was used
FDA (Devices) · 2026-02-18
- CriticalFrozen whole surti papdi recalled after testing positive for Salmonella
FDA (Devices) · 2025-12-17
Same hazard · false negative
See all →- HighHealthmark ProChek-II Medical Tests Recalled for Expired Ingredient
FDA (Devices) · 2026-09-02
- HighSiemens Immulite 2000 Syphilis Screen Recall for Negative Bias
FDA (Devices) · 2026-08-26
- HighSiemens Immulite 2000 Syphilis Screen Recalled for Negative Bias
FDA (Devices) · 2026-08-26
- HighHologic Aptima HPV assay may produce invalid or false negative results
FDA (Devices) · 2026-03-25
- HighHeparin antibody test kits recalled over a defect producing negative results
FDA (Drugs) · 2026-01-21