Arrow hemodialysis catheter kits contain drugs from a recalled supplier
Arrow two-lumen hemodialysis kits are being recalled because they contain lidocaine, bupivacaine and 0.9% sodium chloride subject to a supplier recall. 15,425 units affected.
The agency’s classification
FDA (Devices): Class I
Class I is a situation in which there is a reasonable probability that the use of or exposure to a violative product will cause serious adverse health consequences or death.
This is the agency’s own classification, taken from its notice. The severity score below is our assessment, not the agency’s. How recall classes work
What this means for you
Our assessment, not the agency’s: severity 4/5 (Severe).
The hazard poses a serious risk of illness, injury, or structural failure. Stop using the product immediately and contact the manufacturer for a refund or repair.
Our severity reasoning: FDA Class I recall, which the rubric floors at Severe; no death is reported in the source, so it does not escalate to Critical.
Plain-English summary
Arrow International is recalling two-lumen hemodialysis kits, covering REF AK-11802-F, AK-22122, AK-22142-CDC and AK-25142-CDC among the catalog numbers listed in the recall notice. The action covers 15,425 units.
Lidocaine, bupivacaine and 0.9% sodium chloride, the subject of a supplier recall, were included in the kits and sets. According to the manufacturer, the supplier was found to have quality issues and limited reports of drug ineffectiveness, and if sterility of a water-based injectable product is compromised this could lead to serious or life-threatening injuries. The FDA classified the action as Class I, its most serious classification.
Distribution was nationwide in the United States across a large number of states. The kits are supplied to healthcare facilities rather than sold to consumers.
Facilities should identify affected stock using the REF numbers, UDI-DI codes and lot numbers in the recall notice, quarantine it, and follow the instructions issued by Arrow International.
The recalled product
- Product
- Description/REF: HEMODIALYSIS KIT: 2-LUMEN 8 FR X 11 CM/AK-11802-F; HEMODIALYSIS KIT: 2-LUMEN 12 FR X 16 CM/AK-22122; HEMODIALYSIS KIT 2L: 14 FR X 15CM AGB/AK-22142-CDC; HEMODIALYSIS KIT 2L: 14 FR X 20CM AGB/AK-25142-CDC; HEMODIALYSIS KIT: 2L 14 FR X 15 CM/ASK-22142-UMD…
- Manufacturer
- ARROW INTERNATIONAL, LLC
- Affected units
- 15,425
Is your product affected?
Check your packaging against the codes below. If any of them match, the product is part of this recall.
Lot numbers (20)
- REF(Material)/UDI-DI/Lot(Batch): AK-11802-F/10801902193275
- 10801902197129/33F24G0344
- 33F24J0511
- 33F24L0637
- 33F25A0473
- 33F25E0060
- 33F25J0473
- 33F24A0848
- AK-22122/10801902193275/33F24H0232
- 33F24J0523
- AK-22142-CDC/10801902193275/33F24J0527
- 33F24K0129
- 33F24L0649
- 33F25C0115
- 33F25C0959
- 33F25E0940
- 33F25J0481
- 33F25L0428
- 33F24G0364
- AK-25142-CDC/10801902193275
Distribution
Part of a larger recall action
This product is one of 35 recalled by ARROW INTERNATIONAL, LLC under a single FDA (Devices) recall action. The agency files each product separately, so the size of the action is not visible from this page alone.
See all 35 products in this recall →Related recalls
Same manufacturer · ARROW INTERNATIONAL, LLC
See all →- SevereArrow drainage and thoracentesis kits contain recalled supplier drugs
FDA (Devices) · 2026-06-24
- SevereArrow paracentesis kit JHH contains drugs from a recalled supplier
FDA (Devices) · 2026-06-24
- SevereArrow access trays and PSI kits contain recalled supplier drugs
FDA (Devices) · 2026-06-24
- SevereArrow peritoneal lavage catheter kits contain recalled supplier drugs
FDA (Devices) · 2026-06-24
- SevereArrow access trays and StimuCath kits contain recalled supplier drugs
FDA (Devices) · 2026-06-24
Same hazard · contained drug recall
See all →- SevereArrow single lumen CVC kits contain drugs from recalled supplier
FDA (Devices) · 2026-06-24
- SevereArrow paracentesis kits contain drugs from a recalled supplier
FDA (Devices) · 2026-06-24
Similar recalls
- HighArrow International Recalls Two-Lumen Hemodialysis Catheterization Kits
FDA (Devices) · 2017-06-14
- HighArrow Hemodialysis Catheter Kits Recalled for Sterility Concerns
FDA (Devices) · 2018-04-18
- HighArrow International Recalls Multi-Lumen Hemodialysis Catheterization Kits Due to Sterility Concerns
FDA (Devices) · 2018-04-18
- HighArrow International Recalls Hemodialysis Catheter Kits Due to Packaging Seal Issues
FDA (Devices) · 2017-06-21
- ModerateArrow International Recalls Hemodialysis Catheterization Kits with Incorrect Guide
FDA (Devices) · 2018-02-28
- ModerateArrow International Recalls Two-Lumen Central Venous Access Kits
FDA (Devices) · 2018-06-06