The Recall Desk
Agency classClass IIOur severity:SevereFDA (Devices)·Z-2350-2018·Announced 2018-07-11

FHC Depth Stop Adapter Recall Due to Lead and Stylet Damage

FHC, Inc. is recalling the Depth Stop Adapter because over-tightening the screw can damage implanted leads and stylets, requiring replacement.

The agency’s classification

FDA (Devices): Class II

Class II is a situation in which use of or exposure to a violative product may cause temporary or medically reversible adverse health consequences, or where the probability of serious adverse health consequences is remote.

This is the agency’s own classification, taken from its notice. The severity score below is our assessment, not the agency’s. How recall classes work

What this means for you

Our assessment, not the agency’s: severity 4/5 (Severe).

The hazard poses a serious risk of illness, injury, or structural failure. Stop using the product immediately and contact the manufacturer for a refund or repair.

Our severity reasoning: The recall is classified as FDA Class II and involves significant injury reports, specifically the need to remove and replace implanted leads due to damage.

Plain-English summary

FHC, Inc. is recalling the microTargeting STar Drive System (Motorized), Product Number ST-DS-ME, specifically the FHC Depth Stop Adapter. The recall involves 2,952 units distributed to 48 US states and 17 foreign countries.

The recall is issued because the Depth Stop Adapter may cause damage to the lead and stylet when the depth stop screw is over-tightened onto the lead. This damage can manifest as low impedances or a short circuit between one or multiple electrode pairs in the lead. The source text states that this damage has resulted in the need to remove and replace the implanted lead intraoperatively or in a follow-up procedure.

Healthcare providers and patients should contact FHC, Inc. for instructions on how to proceed with the recall and replacement of affected devices.

The recalled product

Product
microTargeting" STar" Drive System (Motorized), Product Number ST-DS-ME
Manufacturer
FHC, Inc.
Affected units
2,952

Is your product affected?

Check your packaging against the codes below. If any of them match, the product is part of this recall.

Lot numbers (1)

  • UDI: 00873263004432

Distribution

Part of a larger recall action

This product is one of 15 recalled by FHC, Inc. under a single FDA (Devices) recall action. The agency files each product separately, so the size of the action is not visible from this page alone.

See all 15 products in this recall →

Same manufacturer · FHC, Inc.

See all →