Roche Recalls COBAS AmpliPrep Sample Tube Input Barcode Clips
Roche Molecular Systems is recalling certain lots of COBAS AmpliPrep barcode clips because they can cause sample tubes to stick, leading to aborted instrument runs.
What this means for you
Minor labeling or packaging issue, or a voluntary precautionary recall. Risk of harm is low but real — follow the manufacturer's instructions.
Our severity reasoning: The recall is classified as FDA Class II. The hazard involves an operational failure (aborted runs) rather than a direct safety risk to patients or erroneous results, fitting the criteria for a moderate severity issue.
Plain-English summary
Roche Molecular Systems, Inc. is recalling 2,132 boxes of COBAS AmpliPrep Sample Tube Input Barcode Clips, catalog number 03137040001. The affected lots are 11049503, 11053503, 11054503, 11525503, and 11526503. These products were distributed nationwide in the United States.
The recall was initiated because the barcode clips can cause sample input tubes to stick. When this occurs, the COBAS AmpliPrep instrument lifts both the sample input tube and the barcode clip from the sample rack, resulting in an aborted run. The source text notes that no erroneous results would be produced if the issue occurs.
Users of the COBAS AmpliPrep Instrument should check their inventory for the specific catalog and lot numbers listed above. If the affected products are found, they should stop using them and contact Roche Molecular Systems for instructions on return or replacement.
The recalled product
- Product
- COBAS AmpliPrep Sample Tube Input Barcode Clips, catalog number 03137040001. Input S-tube barcode clips hold sample input tubes in the COBAS AmpliPrep Instrument (CAP) sample racks and have a barcode to identify the sample.
- Manufacturer
- Roche Molecular Systems, Inc.
- Hazard
- Affected units
- 2,132
Is your product affected?
Check your packaging against the codes below. If any of them match, the product is part of this recall.
Lot numbers (4)
- Lot Numbers: 11049503
- 11053503
- 11054503
- 11525503 and 11526503.
Distribution
Distributed nationwide across the United States.
Related recalls
Same manufacturer · Roche Molecular Systems, Inc.
See all →- HighRoche cobas Influenza Test Recall Due to False Negative Results
FDA (Devices) · 2023-02-15
- HighRoche Influenza A/B and RSV Test Kit Recalled for False Negative Results
FDA (Devices) · 2023-02-15
- HighRoche cobas Flu/COVID assay recalled for false negative influenza results
FDA (Devices) · 2023-02-15
- HighMedical Device Recall: Roche cobas 5800 False Results Due to Baseline Anomalies
FDA (Devices) · 2022-10-26
- ModerateCobas Liat respiratory virus analyzer units recalled for false Influenza B results
FDA (Devices) · 2022-04-06
Same hazard · operational failure
See all →- HighBaxter TruSystem 7500 Hybrid medical device software defect in emergency mode
FDA (Devices) · 2025-04-02
- HighBeckman Coulter DxI 9000 immunoassay analyzer recalled for software-related motion errors
FDA (Devices) · 2025-02-12
- HighMEERA CL operating table stops responding during surgery when IR-Hand Control errors occur
FDA (Devices) · 2024-11-20
- HighMobile Operating Table Charger Defect Poses Risk of Operational Failure
FDA (Devices) · 2024-07-17
- HighALPHAMAXX Operating Table Defective Charger Recalled for Operational Failure Risk
FDA (Devices) · 2024-07-17