Siemens Dimension Reagent Management System Power Cord Recall
Siemens Healthcare Diagnostics is recalling 2,554 Dimension Reagent Management System power cords due to a fire hazard caused by partial cord insertion.
The agency’s classification
FDA (Devices): Class II
Class II is a situation in which use of or exposure to a violative product may cause temporary or medically reversible adverse health consequences, or where the probability of serious adverse health consequences is remote.
This is the agency’s own classification, taken from its notice. The severity score below is our assessment, not the agency’s. How recall classes work
What this means for you
Our assessment, not the agency’s: severity 4/5 (Severe).
The hazard poses a serious risk of illness, injury, or structural failure. Stop using the product immediately and contact the manufacturer for a refund or repair.
Our severity reasoning: The recall involves a potential fire hazard, which constitutes a significant injury/property-damage risk under the rubric for a score of 4.
Plain-English summary
Siemens Healthcare Diagnostics, Inc. is recalling 2,554 units of the Dimension Reagent Management System (RMS) power cord. The affected units have serial numbers ranging from 99060001 to 99062554. These power cords are components for the Siemens Healthcare Diagnostics Dimension clinical chemistry analyzers, which are fully automated, microprocessor-controlled systems used to measure analytes in human body fluids.
The recall was initiated because partial insertion of the power cord into the RMS power input module can cause the plug to overheat, potentially creating a fire hazard. The source text notes that RMS units without a power cord retention clip may be affected.
The affected products were distributed across the United States, including Washington, DC, Puerto Rico, the Virgin and Pacific Islands, and all 50 states. Consumers and healthcare facilities should check their equipment for the affected serial numbers and contact Siemens Healthcare Diagnostics for instructions on how to resolve the issue.
The recalled product
- Product
- Dimension(R) Reagent Management System (RMS) - Power Cord. The Siemens Healthcare Diagnostics Dimension(R) clinical chemistry analyzers are floor model, fully automated, microprocessor-controlled, integrated instrument systems that use prepackaged Flex(R) reagent test…
- Manufacturer
- Siemens Healthcare Diagnostics, Inc.
- Hazard
- Affected units
- 2,554
Is your product affected?
Check your packaging against the codes below. If any of them match, the product is part of this recall.
Lot numbers (1)
- Serial numbers 99060001 - 99062554
Distribution
Related recalls
Same manufacturer · Siemens Healthcare Diagnostics, Inc.
See all →- HighSiemens Atellica CH C-Reactive Protein Reagent Recalled for Falsely Elevated Results
FDA (Devices) · 2026-10-07
- SevereSiemens Recalls Emit 2000 Quinidine Assay Due to Precision Issues
FDA (Devices) · 2026-08-26
- SevereSiemens Quinidine Assay Recalled for Precision Errors
FDA (Devices) · 2026-08-26
- HighSiemens Atellica CH Urine Albumin In Vitro Diagnostic Recall
FDA (Devices) · 2026-08-12
- LowSyva Emit Disopyramide Enzyme Immunoassay Calibrator Recalled for Sodium Azide Concentration
FDA (Devices) · 2026-08-12
Same hazard · fire
See all →- SevereVQQQ Gas Patio Heaters Recalled for Fire and Burn Hazards
CPSC · 2026-10-22
- HighAutocar Recalls Xpeditor Trucks for Fuel Hose Separation Risk
NHTSA · 2026-10-21
- SevereAudi Recalls 2024-2025 Q4 E-Tron Models Due to Battery Fire Risk
NHTSA · 2026-10-21
- HighForest River Recalls 2023-2026 Pause Travel Trailers for Wiring Defect
NHTSA · 2026-10-20
- SevereForest River Recalls 2026-2027 Cedar Creek Trailers for Fire Risk
NHTSA · 2026-10-16
Similar recalls
- HighSiemens Reagent Management System Recalled for Electrical Safety Hazard
FDA (Devices) · 2015-12-30
- HighSiemens Reagent Management System Recalled for Electrical Hazard
FDA (Devices) · 2015-12-30
- ModerateSiemens Dimension Chemistry Systems Recalled for Non-Standard Power Plugs
FDA (Devices) · 2015-05-20
- HighSiemens Dimension Chemistry Wash Recalled for Result Shifts
FDA (Devices) · 2014-01-01
- LowSiemens Dimension Integrated Chemistry Systems Recalled for Photometer Misassembly
FDA (Devices) · 2014-08-06
- ModerateSiemens Dimension Ammonia Reagent Cartridges Recalled for Calibration Issues
FDA (Devices) · 2017-02-15