The Recall Desk
Agency classClass IIOur severity:ModerateFDA (Devices)·Z-2165-2015·Announced 2015-07-29

Fiber Stripper Recalled for Cleaning and Sterilization Validation Issues

American Medical Systems is recalling Fiber Strippers because validation data for cleaning and sterilization methods does not meet FDA guidance for reusable products.

The agency’s classification

FDA (Devices): Class II

Class II is a situation in which use of or exposure to a violative product may cause temporary or medically reversible adverse health consequences, or where the probability of serious adverse health consequences is remote.

This is the agency’s own classification, taken from its notice. The severity score below is our assessment, not the agency’s. How recall classes work

What this means for you

Our assessment, not the agency’s: severity 2/5 (Moderate).

Minor labeling or packaging issue, or a voluntary precautionary recall. Risk of harm is low but real — follow the manufacturer's instructions.

Our severity reasoning: The recall is classified as FDA Class II. The hazard involves a failure to meet validation guidance for cleaning and sterilization, which is a significant quality issue but does not involve reported injuries or deaths, fitting the Moderate severity criteria for non-critical labeling or validation issues.

Plain-English summary

American Medical Systems Innovation Center - Silicon Valley is recalling the Fiber Stripper, 200µm, Autoclavable. This tool is used to strip the protective jacket off the end of an Endostat or SureFlex Fiber. The device does not come into direct contact with the patient.

The recall is being issued because validation data related to cleaning instructions and sterilization methods for the Reusable Fibers and Reusable Stripper and Cleaver Accessories do not meet the new FDA guidance on qualifying reusable products. The affected devices are Model 80-10001-003, manufactured between April 8, 2012, and April 8, 2015.

A total of 16,710 devices are affected. The products were distributed worldwide, including in the United States and numerous other countries. Users should stop using the affected devices and contact the manufacturer for instructions on return or replacement.

The recalled product

Product
Fiber Stripper, 200¿m, Autoclavable; The Fiber Stripper is a tool provided for stripping the protective jacket off the end of an Endostat or SureFlex Fiber. The Fiber Stripper does not come in to direct contact with the patient. It is available in several sizes, corresponding…

Is your product affected?

Check your packaging against the codes below. If any of them match, the product is part of this recall.

Lot numbers (4)

  • Model 80-10001-003
  • All product manufactured from April 8
  • 2012 to April 8
  • 2015

Distribution

Distributed nationwide across the United States.

Part of a larger recall action

This product is one of 32 recalled by American Medical Systems Innovation Center - Silicon Valley under a single FDA (Devices) recall action. The agency files each product separately, so the size of the action is not visible from this page alone.

See all 32 products in this recall →

Same manufacturer · American Medical Systems Innovation Center - Silicon Valley

See all →