FDA Recalls Sureflex 550 Lithotripsy Fibers Due to Sterilization Validation Issues
American Medical Systems is recalling Sureflex 550 Lithotripsy Fibers because cleaning and sterilization validation data do not meet FDA guidance for reusable products.
The agency’s classification
FDA (Devices): Class II
Class II is a situation in which use of or exposure to a violative product may cause temporary or medically reversible adverse health consequences, or where the probability of serious adverse health consequences is remote.
This is the agency’s own classification, taken from its notice. The severity score below is our assessment, not the agency’s. How recall classes work
What this means for you
Our assessment, not the agency’s: severity 2/5 (Moderate).
Minor labeling or packaging issue, or a voluntary precautionary recall. Risk of harm is low but real — follow the manufacturer's instructions.
Our severity reasoning: The recall is classified as FDA Class II, but the source text does not report any illnesses, injuries, or hospitalizations. The hazard is related to validation data not meeting guidance, which fits the criteria for a moderate severity score involving low-risk contamination or voluntary precautionary measures without reported harm.
Plain-English summary
The U.S. Food and Drug Administration (FDA) has issued a Class II recall for the Sureflex 550 Lithotripsy Fiber, 5x, Blue (Model R-LLF550). The recall involves 16,710 devices manufactured between April 8, 2012, and April 8, 2015. These laser fibers are intended for use in surgical procedures involving the vaporization, ablation, and fragmentation of calculi and soft tissue.
The recall was initiated because validation data related to cleaning instructions and sterilization methods for the Sureflex Reusable Fibers and associated accessories do not meet FDA guidance on qualifying reusable products. The devices were distributed worldwide, including in the United States and numerous other countries.
Healthcare facilities and distributors should stop using the affected devices and contact American Medical Systems for further instructions on the recall process.
The recalled product
- Product
- Sureflex 550 Lithotripsy Fiber, 5x, Blue; Sure F!ex laser fibers are intended for use in laser-based surgical applications, including, but not limited to endoscopic, laparoscopic and open surgical procedures involving vaporization, ablation, coagulation, hemostasis, excision…
- Affected units
- 16,710
Is your product affected?
Check your packaging against the codes below. If any of them match, the product is part of this recall.
Lot numbers (3)
- Model R-LLF550 All product manufactured from April 8
- 2012 to April 8
- 2015
Distribution
Distributed nationwide across the United States.
Part of a larger recall action
This product is one of 32 recalled by American Medical Systems Innovation Center - Silicon Valley under a single FDA (Devices) recall action. The agency files each product separately, so the size of the action is not visible from this page alone.
See all 32 products in this recall →Related recalls
Same manufacturer · American Medical Systems Innovation Center - Silicon Valley
See all →- ModerateFDA Recalls Disposable Fiber Strippers Due to Sterilization Validation Issues
FDA (Devices) · 2015-07-29
- ModerateFDA Recalls Fiber Cleaver Accessories Due to Sterilization Validation Issues
FDA (Devices) · 2015-07-29
- ModerateFDA Recalls Fiber Cleavers Due to Sterilization Validation Issues
FDA (Devices) · 2015-07-29
- HighFDA Recalls Disposable EndoStat Fiber Strippers Due to Validation Deficiencies
FDA (Devices) · 2015-07-29
- ModerateFDA Recalls Fiber Stripper Tools Due to Sterilization Validation Issues
FDA (Devices) · 2015-07-29
Same hazard · sterilization failure
See all →- SevereSteris Recalls VERIFY Biological Steam Test Packs Due to Spore Population Concerns
FDA (Devices) · 2026-10-14
- HighSteris VERIFY Dual Species Biological Indicator Recall for Spore Population Issues
FDA (Devices) · 2026-10-14
- SevereSteris VERIFY Biological Indicator Recall Due to Spore Population Concerns
FDA (Devices) · 2026-10-14
- SevereSteris VERIFY Biological Steam Test Pack Recalled for Spore Population Defect
FDA (Devices) · 2026-10-14
- SevereSteris VERIFY Biological/Integrator Challenge Pack Recall for Spore Population Defect
FDA (Devices) · 2026-10-14
Similar recalls
- HighPhotoMedex Recalls SLT Fiber Delivery Systems Due to Sterility Concerns
FDA (Devices) · 2013-10-09
- ModerateBoston Scientific Recalls StarMedTec LightTrail Reusable Fibers for Laser Compatibility
FDA (Devices) · 2014-10-15
- HighSOLTIVE SuperPulsed Laser Fibers Recalled for Potential Fiber Breakage During Surgery
FDA (Devices) · 2024-01-24
- HighOlympus Laser Fiber Stripper Recall Due to Sterilization Validation Gap
FDA (Devices) · 2024-01-24
- ModerateFDA Recalls Polysorb Sutures Due to Potential Packaging Integrity Issues
FDA (Devices) · 2020-02-26
- ModerateFDA Recalls Polysorb Sutures Due to Potential Sterility Barrier Issues
FDA (Devices) · 2020-02-26