Stryker Neptune 1 Gold Rover Waste Management System Recalled for Injury Risk
Stryker is recalling Neptune 1 Gold Rover waste management systems due to reports of serious injury and a fatality associated with the Neptune 2 system.
The agency’s classification
FDA (Devices): Class I
Class I is a situation in which there is a reasonable probability that the use of or exposure to a violative product will cause serious adverse health consequences or death.
This is the agency’s own classification, taken from its notice. The severity score below is our assessment, not the agency’s. How recall classes work
What this means for you
Our assessment, not the agency’s: severity 5/5 (Critical).
Highest-severity recall. The hazard can cause serious injury or death. Stop using the product now and follow the agency's instructions.
Our severity reasoning: The source text explicitly states that one event resulted in a fatality, which mandates a severity score of 5 (Critical) regardless of other factors.
Plain-English summary
Stryker Instruments Div. of Stryker Corporation is recalling the Neptune 1 Gold Rover (120V and 230V) Waste Management System. The recall involves 5,925 domestic units and 2 international units distributed nationwide in the USA and in Canada. The devices are intended for use in surgical environments or physician's offices to collect and dispose of surgical fluid waste.
The recall was initiated after Stryker received two reports of serious injury resulting from tissue damage associated with the use of the Neptune 2 system. One of these events resulted in a fatality when a customer connected the Neptune 2 System to a passive chest drainage tube post-operatively. The Instructions for Use (IFU) did not specifically warn against connecting Neptune Waste Management devices, which are high vacuum/high flow devices, to passive drainage tubes.
Customers were instructed on June 12, 2012, to review the revised IFU, distribute it to affected departments, and educate users on this warning. Warning labels are required on all Neptune devices and are pending distribution. Additionally, the FDA advised that the Neptune 1 Silver, Neptune 2 Ultra (120V), and Neptune 2 Ultra (230V) should not be used because they lacked FDA 510(k) clearance. Customers who choose to continue using these specific uncleared devices must complete a Certificate of Medical Necessity and return it to Stryker by October 12, 2012.
The recalled product
- Product
- Neptune 1 Gold Rover (120 Vand 230V ) Waste Management System Part Number 0700-001-000, 0700-002-000(International) Instructions For Use 0700-001-700, 0700-002-707 Intended for use in the surgical environment or physician's office to collect and dispose of surgical fluid…
- Manufacturer
- Stryker Instruments Div. of Stryker Corporation
Is your product affected?
Check your packaging against the codes below. If any of them match, the product is part of this recall.
Lot numbers (1)
- 120V Gold part numbers 0700-001-000 IFU 0700-001-700 230 V Gold 0700-002- 000
Distribution
Distributed nationwide across the United States.
Part of a larger recall action
This product is one of 4 recalled by Stryker Instruments Div. of Stryker Corporation under a single FDA (Devices) recall action. The agency files each product separately, so the size of the action is not visible from this page alone.
See all 4 products in this recall →Related recalls
Same manufacturer · Stryker Instruments Div. of Stryker Corporation
See all →- HighStryker Triton Canister Scanning Labels Recalled Due to Color Variance
FDA (Devices) · 2022-07-27
- ModerateTriton Sponge System Software Instructions Updated for Proper Sponge Handling
FDA (Devices) · 2022-07-27
- HighSurgiCount+ System software error bypasses sponge tracking alerts
FDA (Devices) · 2022-07-27
- SevereStryker Recalls Cranial Perforators Due to Potential Metal Fragment Risk
FDA (Devices) · 2020-11-18
- ModerateStryker Neptune E-SEP Blade Electrode Recalled for Expired Product
FDA (Devices) · 2020-05-06
Same hazard · tissue damage
See all →- HighJuggerLoc/ToggleLoc 2.9mm Drill Guide Warping in Packaging
FDA (Devices) · 2024-11-27
- HighBard Marquee Biopsy Instruments Recalled for Potential Needle Detachment
FDA (Devices) · 2024-11-27
- SevereFDA Updates POLARx Cryoablation Catheter Instructions for Atrio-esophageal Fistula Risk
FDA (Devices) · 2024-11-20
- HighStryker iBur Diamond Round surgical bur recalled for thermal overheating risk
FDA (Devices) · 2024-09-04
- HighStryker iBur Surgical Bur Recalled Due to Excessive Temperature Risk
FDA (Devices) · 2024-09-04
Similar recalls
- ModerateStryker Neptune 2 Waste Management System Recalled for Diverter Valve Failures
FDA (Devices) · 2015-07-29
- ModerateStryker Neptune 2 Rover Recall Due to Castor Binding Defect
FDA (Devices) · 2012-12-12
- HighStryker Neptune 2 Rover Recall Due to Binding Casters
FDA (Devices) · 2012-12-12
- HighStryker Neptune 2 Waste Management System Docking Station Recall
FDA (Devices) · 2015-09-09
- ModerateStryker Neptune 2 Rover Ultra Waste Management System Field Upgrade Recall
FDA (Devices) · 2014-02-12
- HighNeptune Medical Suction Rovers Recalled Due to Loose Bolts
FDA (Devices) · 2025-01-15