Sysmex XN Series Hematology Analyzers Recalled for Fire Risk
Sysmex America Inc is recalling XN-10 and XN-11 hematology analyzers due to a risk of fire caused by misaligned internal protective sheets.
What this means for you
Illness, injury, or structural failure has been reported. Stop using the product immediately and contact the manufacturer for a refund or repair.
Our severity reasoning: The source text explicitly reports that the analyzers caught fire, which constitutes a significant property-damage report and a risk of serious injury, meeting the criteria for a Severe (4) rating.
Plain-English summary
Sysmex America Inc is recalling 3,682 units of the Automated Hematology Analyzer XN Series, specifically models XN-10 and XN-11. The recall is due to customer complaints that the analyzers caught fire. The fire risk is caused by the misalignment of an internal protective sheet that is designed to shield electronic components from salty reagents.
When the protective sheet is misaligned, salt deposits can build up on unprotected electronic components. These deposits may combust due to the heat generated by the electronics when the instrument cycles. A completed corrective and preventive action (CAPA) identified the root cause as a supplier providing a defective sensor ready valve and the misalignment of the electronic protective shield during manufacturing.
The affected units were distributed nationwide in the United States and in Canada. Users of these devices should contact Sysmex America Inc for instructions on how to address the defect and prevent potential fire hazards.
The recalled product
- Product
- Sysmex, Automated Hematology Analyzer XN Series, Sysmex Corporation, 1-5-1 Wakinohama-Kaigandori, Chuo-ku, Kobe 651-0073 Japan
- Manufacturer
- Sysmex America Inc
- Hazard
- Affected units
- 3,682
Is your product affected?
Check your packaging against the codes below. If any of them match, the product is part of this recall.
Lot numbers (20)
- **Model: XN-10
- Product Number: AP795756
- Serial Numbers: 11001 29656
- 29668-29786
- 29788-29798
- 29800-29801
- 29803
- 29804
- 29807
- 29812
- 29816-29867 **Model: XN-11
- Product Number: CS297931
- Serial Numbers: A1001-A1124
- A1127
- A1130
- A1135-A1140
- A1142-A1144
- A1146-A1149
- A1151
- A1153-A1159
Distribution
Distributed nationwide across the United States.
Related recalls
Same manufacturer · Sysmex America Inc
See all →- ModerateLaboratory tube sorters recalled over a lift that may drop unexpectedly
FDA (Devices) · 2026-01-28
- HighSysmex PS-10 Sample Preparation System Recalled for False Test Results
FDA (Devices) · 2025-02-12
- HighSysmex PS-10 Sample Preparation System Recalled for Insufficient Antibody
FDA (Devices) · 2020-07-29
- SevereSysmex CV-11 Sample Unit Recalled for Combustion Hazard
FDA (Devices) · 2020-04-22
- HighSysmex CF-70 Blood Smear Transporter Recalled for Software Mismatch Hazard
FDA (Devices) · 2020-04-22
Same hazard · fire
See all →- Severe
- Severe
- SeverePrepared Hero Recalls Fire Spray Canisters Due to Rupture Risk
CPSC · 2026-09-10
- SevereSweet Cubby Recalls Children's Loungewear Sets Due to Burn Hazard
CPSC · 2026-09-10
- HighKibbi Recalls 2026-2027 Renegade RVs Due to Fire Risk
NHTSA · 2026-09-05