Sysmex XN Series Hematology Analyzers Recalled for Fire Risk
Sysmex America Inc is recalling XN-10 and XN-11 hematology analyzers due to a risk of fire caused by misaligned internal protective sheets.
The agency’s classification
FDA (Devices): Class II
Class II is a situation in which use of or exposure to a violative product may cause temporary or medically reversible adverse health consequences, or where the probability of serious adverse health consequences is remote.
This is the agency’s own classification, taken from its notice. The severity score below is our assessment, not the agency’s. How recall classes work
What this means for you
Our assessment, not the agency’s: severity 4/5 (Severe).
The hazard poses a serious risk of illness, injury, or structural failure. Stop using the product immediately and contact the manufacturer for a refund or repair.
Our severity reasoning: The source text explicitly reports that the analyzers caught fire, which constitutes a significant property-damage report and a risk of serious injury, meeting the criteria for a Severe (4) rating.
Plain-English summary
Sysmex America Inc is recalling 3,682 units of the Automated Hematology Analyzer XN Series, specifically models XN-10 and XN-11. The recall is due to customer complaints that the analyzers caught fire. The fire risk is caused by the misalignment of an internal protective sheet that is designed to shield electronic components from salty reagents.
When the protective sheet is misaligned, salt deposits can build up on unprotected electronic components. These deposits may combust due to the heat generated by the electronics when the instrument cycles. A completed corrective and preventive action (CAPA) identified the root cause as a supplier providing a defective sensor ready valve and the misalignment of the electronic protective shield during manufacturing.
The affected units were distributed nationwide in the United States and in Canada. Users of these devices should contact Sysmex America Inc for instructions on how to address the defect and prevent potential fire hazards.
The recalled product
- Product
- Sysmex, Automated Hematology Analyzer XN Series, Sysmex Corporation, 1-5-1 Wakinohama-Kaigandori, Chuo-ku, Kobe 651-0073 Japan
- Manufacturer
- Sysmex America Inc
- Hazard
- Affected units
- 3,682
Is your product affected?
Check your packaging against the codes below. If any of them match, the product is part of this recall.
Lot numbers (20)
- **Model: XN-10
- Product Number: AP795756
- Serial Numbers: 11001 29656
- 29668-29786
- 29788-29798
- 29800-29801
- 29803
- 29804
- 29807
- 29812
- 29816-29867 **Model: XN-11
- Product Number: CS297931
- Serial Numbers: A1001-A1124
- A1127
- A1130
- A1135-A1140
- A1142-A1144
- A1146-A1149
- A1151
- A1153-A1159
Distribution
Distributed nationwide across the United States.
Related recalls
Same manufacturer · Sysmex America Inc
See all →- ModerateLaboratory tube sorters recalled over a lift that may drop unexpectedly
FDA (Devices) · 2026-01-28
- HighSysmex PS-10 Sample Preparation System Recalled for False Test Results
FDA (Devices) · 2025-02-12
- HighSysmex PS-10 Sample Preparation System Recalled for Insufficient Antibody
FDA (Devices) · 2020-07-29
- HighSysmex CF-70 Blood Smear Transporter Recalled for Software Mismatch Hazard
FDA (Devices) · 2020-04-22
- SevereSysmex CV-11 Sample Unit Recalled for Combustion Hazard
FDA (Devices) · 2020-04-22
Same hazard · fire
See all →- SevereVQQQ Gas Patio Heaters Recalled for Fire and Burn Hazards
CPSC · 2026-10-22
- HighForest River Recalls 2023-2026 Pause Travel Trailers for Wiring Defect
NHTSA · 2026-10-20
- SevereForest River Recalls 2026-2027 Cedar Creek Trailers for Fire Risk
NHTSA · 2026-10-16
- SevereAlliance RV Recalls Valor and Delta Trailers Due to Fire Risk
NHTSA · 2026-10-16
- SevereSpartan RV Chassis K2 Recalled for Fuel Hose Separation Risk
NHTSA · 2026-10-15
Similar recalls
- HighSysmex XN-Series Hematology Analyzers Recalled for Potential Elevated Values
FDA (Devices) · 2016-06-22
- ModerateSysmex pocH-100i Automated Hematology Analyzer Recalled for Data Omission
FDA (Devices) · 2013-11-06
- HighSysmex PS-10 Sample Preparation System Recalled for Insufficient Antibody
FDA (Devices) · 2020-07-29
- ModerateSysmex XE-2100 Hematology System Recalled for Reticulocyte Count Bias
FDA (Devices) · 2012-11-14
- ModerateSysmex XE-2100C Hematology System Recalled for Reticulocyte Count Bias
FDA (Devices) · 2012-11-14
- ModerateSysmex Fluorocell PLT Reagent Recalled for Inaccurate Platelet Counts
FDA (Devices) · 2018-09-05