Joerns Healthcare Recalls P.R.O. Matt and ProCair Mattresses Due to Connector Defect
Joerns Healthcare is recalling 915 P.R.O. Matt and ProCair Mattresses because incorrect connectors were supplied, lacking internal shutoff valves required to prevent air loss.
The agency’s classification
FDA (Devices): Class II
Class II is a situation in which use of or exposure to a violative product may cause temporary or medically reversible adverse health consequences, or where the probability of serious adverse health consequences is remote.
This is the agency’s own classification, taken from its notice. The severity score below is our assessment, not the agency’s. How recall classes work
What this means for you
Our assessment, not the agency’s: severity 2/5 (Moderate).
Minor labeling or packaging issue, or a voluntary precautionary recall. Risk of harm is low but real — follow the manufacturer's instructions.
Our severity reasoning: The recall is classified as FDA Class II, but the source text does not report any hospitalizations, injuries, or deaths. The hazard is a functional defect (air loss) without reported harm, fitting the criteria for a moderate severity score.
Plain-English summary
Joerns Healthcare has identified a potential issue with 915 units of P.R.O. Matt and ProCair Mattresses. The recall involves specific serial numbers where incorrect connectors (part numbers 206-0401 and 206-0405) were provided by a supplier. These connectors lack the required internal shutoff valves, which are necessary to prevent air from escaping the mattress air cells.
The affected devices are non-powered flotation therapy mattresses intended to treat or prevent decubitus ulcers (bed sores). The distribution pattern is listed as worldwide, including the US, UK, Canada, and Holland. The source text does not report any illnesses, injuries, or deaths associated with this defect.
Consumers and healthcare facilities should check the serial numbers of their P.R.O. Matt and ProCair Mattresses against the list provided in the official recall notice. If a device is affected, users should contact Joerns Healthcare for instructions on how to proceed with the recall.
The recalled product
- Product
- P.R.O. Matt: (a) PROMATT PLUS 36 X 76 MATRESS, NYLON, MATERIAL NUMBER PMP3676NZBA (b) PROMATT PLUS 84" MATRESS, NYLON, MATERIAL NUMBER PMP3684NZBA (c) PROMATT PLUS W/SIDEWALLS, 42X80, MATERIAL NUMBER PMS4280NZBA (d) PROMATT PLUS 36IN MATRESS, MATERIAL NUMBER PMP3680NZBA…
- Manufacturer
- Joerns Healthcare
- Affected units
- 915
Is your product affected?
Check your packaging against the codes below. If any of them match, the product is part of this recall.
Lot numbers (15)
- P.R.O. Matt: (a) PROMATT PLUS 36 X 76 MATRESS
- NYLON
- MATERIAL NUMBER PMP3676NZBA
- Serial Numbers: 1000057154 1000057153 (b) PROMATT PLUS 84" MATRESS
- MATERIAL NUMBER PMP3684NZBA
- 42X80
- MATERIAL NUMBER PMS4280NZBA
- MATERIAL NUMBER PMP3680NZBA
- MATERIAL NUMBER PMS3680NZBA
- STRETCH
- MATERIAL NUMBER PMP3680SZBA
- MATERIAL NUMBER PMP4280NZBA
- DURATUB
- MATERIAL NUMBER PMPE3480SZBADT
- Serial Numbers: PMPE3480SZBADT PMPE3480SZBADT PMPE3480SZBADT
Distribution
Distributed nationwide across the United States.
Part of a larger recall action
This product is one of 2 recalled by Joerns Healthcare under a single FDA (Devices) recall action. The agency files each product separately, so the size of the action is not visible from this page alone.
See all 2 products in this recall →Related recalls
Same manufacturer · Joerns Healthcare
See all →- ModerateJoerns Healthcare Recalls ProCair Mattresses Due to Missing Air Shutoff Valves
FDA (Devices) · 2018-06-06
- HighJoerns EasyCare Bed Recalled Due to Care-Lock Feature Failure Risk
FDA (Devices) · 2014-11-05
Same hazard · defective connector
See all →- HighBio-Detek Recalls Medsolutions4U Multi-Function Electrodes Due to Connector Defect
FDA (Devices) · 2014-09-10
- ModerateIntegra Rickham Reservoirs Recalled for Bent or Broken Connectors
FDA (Devices) · 2014-01-15
Similar recalls
- ModerateBaxter Recalls PRO+ MRS Mattress Due to UDI Labeling Error
FDA (Devices) · 2026-09-30
- ModerateBaxter PRO+ 36-inch Mattress Recalled for UDI Labeling Error
FDA (Devices) · 2026-09-30
- ModerateBaxter PRO+ Mattress Recalled for UDI Labeling Error
FDA (Devices) · 2026-09-30
- HighMedical Device Recall: Hill-Rom PRO+ Mattress Microclimate Feature Failure
FDA (Devices) · 2022-07-27
- SevereStryker Position Pro Mattress Recalled for Power Cord Fire Risk
FDA (Devices) · 2016-03-16
- ModeratePrimus Medical Recalls Transcend and Bariatric Mattresses Due to Delamination
FDA (Devices) · 2014-09-17