Omnipod 5 iOS App software design issue causes insulin overdose risk
Insulet Corporation is recalling the Omnipod 5 iOS App due to a software issue where decimal separators are not recognized in manual insulin entry, potentially causing insulin overdose and severe hypoglycemia.
The agency’s classification
FDA (Devices): Class II
Class II is a situation in which use of or exposure to a violative product may cause temporary or medically reversible adverse health consequences, or where the probability of serious adverse health consequences is remote.
This is the agency’s own classification, taken from its notice. The severity score below is our assessment, not the agency’s. How recall classes work
What this means for you
Our assessment, not the agency’s: severity 4/5 (Severe).
The hazard poses a serious risk of illness, injury, or structural failure. Stop using the product immediately and contact the manufacturer for a refund or repair.
Our severity reasoning: The FDA classified this as Class II. The defect directly enables insulin overdose, which could cause severe hypoglycemia—a serious adverse health consequence. Although no illnesses are reported in the source text, the mechanism of harm is confirmed and potentially severe, meeting the Class II + serious health risk criterion for a Severe rating.
Plain-English summary
Insulet Corporation is recalling the Omnipod 5 Automated Insulin Delivery System iOS Application (Model Number PT-000664/M009-S-AP, all software versions 1.X.X) due to a software design defect. The application fails to recognize commas as decimal separators and does not allow users to enter a decimal separator on manual entry screens if they have manually changed their phone settings to a region that uses a comma as the decimal separator instead of a period, or have selected that format as a preference.
When the decimal separator is not recognized, the software may incorrectly increase the insulin dose. For example, a user intent to enter 0.1 unit may be processed as 1 unit. This miscalculation could result in over-delivery of insulin and potentially severe hypoglycemia.
The Omnipod 5 iOS App is available only in the US through the Apple App Store and is used to activate and deactivate insulin delivery Pods, display alerts and alarms, and send insulin delivery commands to the Pod. Users who have changed their phone's regional settings to use a comma as a decimal separator should check their device settings and be aware of this potential risk when using manual insulin entry.
Affected users should contact Insulet Corporation for instructions on how to address this issue or update their app when a corrected version becomes available.
The recalled product
- Product
- Omnipod 5 Automated Insulin Delivery System, iOS Application. Model Number: PT-000664/M009-S-AP. Software Version: all Product Description: The Omnipod 5 iOS App software is available to download through the Apple app store. The Omnipod 5 App is the user interface…
- Manufacturer
- Insulet Corporation
- Affected units
- 122,823
Distribution
Distribution scope not specified by the agency.
Related recalls
Same manufacturer · Insulet Corporation
See all →- CriticalOmnipod DASH Pod insulin pump affected by external cannula damage
FDA (Devices) · 2026-07-08
- SevereOmnipod Eros Pod Insulin Delivery Failure Risk
FDA (Devices) · 2026-07-08
- SevereOmnipod 5 Insulin Infusion Pods Recalled for External Cannula Damage
FDA (Devices) · 2026-07-08
- SevereOmnipod 5 insulin pods from 49 lots have torn internal cannulas
FDA (Devices) · 2026-04-22
- HighOmnipod 5 app recalled because magnifying a field duplicates typed digits
FDA (Devices) · 2025-10-22
Same hazard · insulin overdose
See all →- HighMedtronic Paradigm insulin pumps may deliver incorrect doses during flight
FDA (Devices) · 2025-03-12
- SevereOmnipod 5 insulin pump app bolus calculator decimal point error
FDA (Devices) · 2024-01-10
- HighMedtronic MiniMed Insulin Pumps Recalled for Cybersecurity Vulnerability
FDA (Devices) · 2022-11-16
- HighMiniMed 670G Insulin Pump Software Flaw Risks Insulin Overdose
FDA (Devices) · 2021-06-09
- HighMiniMed 780G Insulin Pump Software Fault Risks Overdose
FDA (Devices) · 2021-06-09
Similar recalls
- SevereOmnipod 5 insulin pump app bolus calculator decimal point error
FDA (Devices) · 2024-01-10
- HighOmnipod 5 App Software Error Prevented Control on Android Phones
FDA (Devices) · 2023-09-27
- HighOmnipod 5 app recalled because magnifying a field duplicates typed digits
FDA (Devices) · 2025-10-22
- SevereInsulet Recalls OmniPod Insulin Devices Due to Deployment Failures
FDA (Devices) · 2015-09-16
- SevereOmnipod 5 Insulin Infusion Pods Recalled for External Cannula Damage
FDA (Devices) · 2026-07-08
- HighInsulet Recalls Omnipod DASH Insulin Systems Due to Bolus Calculation Error
FDA (Devices) · 2020-06-10