Beckman Coulter Recalls Triglyceride Assay Due to Dipyrone Interference
Beckman Coulter is recalling Triglyceride OSR6x118 assays because Dipyrone may cause false low results.
What this means for you
Minor labeling or packaging issue, or a voluntary precautionary recall. Risk of harm is low but real — follow the manufacturer's instructions.
Our severity reasoning: The recall is classified as FDA Class II. The hazard involves potential diagnostic errors (false low results) due to chemical interference, which is a significant quality issue but does not involve direct physical injury or acute toxicity, fitting the Moderate category for labeling/quality defects without reported harm.
Plain-English summary
Beckman Coulter Inc. is recalling the Triglyceride OSR6x118 assay, a product intended for the quantitative determination of triglyceride in human serum and plasma on Beckman Coulter AU analyzers. The recall affects 79,009 units total, including 21,215 units in the United States.
The recall was initiated because Beckman Coulter identified that Dipyrone (Metamizole) may cause erroneous, false low results in the Triglyceride OSR6x118 assay. The source text notes that this interference issue also affects other assays, including Cholesterol OSR6X16, Uric Acid OSR6X98, Lactate OSR6X93, and Lipase OSR6X30.
The product was distributed worldwide, including the United States (nationwide including Puerto Rico) and numerous other countries. The recall covers all lots of the affected product. Users should stop using the product and contact Beckman Coulter for further instructions.
The recalled product
- Product
- Triglyceride, Catalog No. OSR6x118 Product Usage:Triglyceride OSR6x118 is intended as an Enzymatic colour test for the quantitative determination of triglyceride in human serum and plasma on Beckman Coulter AU analyzers.
- Manufacturer
- Beckman Coulter Inc.
Is your product affected?
Check your packaging against the codes below. If any of them match, the product is part of this recall.
Lot numbers (1)
- All lots
Distribution
Part of a larger recall action
This product is one of 5 recalled by Beckman Coulter Inc. under a single FDA (Devices) recall action. The agency files each product separately, so the size of the action is not visible from this page alone.
See all 5 products in this recall →Related recalls
Same manufacturer · Beckman Coulter Inc.
See all →- HighBeckman Coulter DxC 500 AU Analyzer Software Defect Recall
FDA (Devices) · 2026-08-26
- HighBeckman Coulter DxC 500i Analyzer Modules Recalled for Software Defect
FDA (Devices) · 2026-08-26
- High
- SevereBeckman Coulter MicroScan Panel Update DVD Kits Recalled
FDA (Devices) · 2026-08-12
- HighBeckman Coulter Access Total T4 Calibrator Recalled for Inaccurate Results
FDA (Devices) · 2026-04-29
Same hazard · diagnostic error
See all →- HighSiemens Immulite 2000 Syphilis Screen Recall for Negative Bias
FDA (Devices) · 2026-08-26
- HighSiemens Immulite 2000 Syphilis Screen Recalled for Negative Bias
FDA (Devices) · 2026-08-26
- HighGEM Premier 5000 Blood Analyzer Recalled Over Bar Code Reader Patient ID Risk
FDA (Devices) · 2026-08-05
- HighPhilips CT Scanners Recalled Due to Bolus Tracking Trigger Issue
FDA (Devices) · 2026-08-05
- HighGEM Premier 5000 Blood Analyzer Recalled Over Bar Code Reader Patient ID Risk
FDA (Devices) · 2026-08-05