Beckman Coulter Lipase Reagent Recalled for Potential False Low Results
Beckman Coulter is recalling Lipase OSR6x30 reagent because Dipyrone may cause erroneous false low results in lipase assays.
What this means for you
Minor labeling or packaging issue, or a voluntary precautionary recall. Risk of harm is low but real — follow the manufacturer's instructions.
Our severity reasoning: The recall is classified as FDA Class II. The hazard involves erroneous test results (false lows) rather than direct physical injury or death, fitting the criteria for moderate severity where the risk is primarily diagnostic error.
Plain-English summary
Beckman Coulter Inc. is recalling Lipase, Catalog No. OSR6x30, a system reagent intended for the quantitative determination of Lipase activity in human serum on Beckman Coulter AU analyzers. The recall involves 32,926 units total, with 6,041 units distributed in the United States.
The recall was initiated because Beckman Coulter identified that Dipyrone (Metamizole) may cause erroneous, false low results in the Lipase OSR6X30 assay. This issue is part of a broader identification affecting several other assays, including Cholesterol, Triglyceride, Uric Acid, and Lactate, but this specific recall notice pertains to the Lipase reagent.
The product was distributed worldwide, including the United States, Puerto Rico, and numerous international countries. All lots are affected. Healthcare facilities and laboratories using this reagent should stop using the product and contact Beckman Coulter for instructions on return or replacement.
The recalled product
- Product
- Lipase, Catalog No. OSR6x30 Product Usage: Lipase OSR6x30 is intended as a System reagent for the quantitative determination of Lipase activity in human serum on Beckman Coulter AU analyzers.
- Manufacturer
- Beckman Coulter Inc.
Is your product affected?
Check your packaging against the codes below. If any of them match, the product is part of this recall.
Lot numbers (1)
- All lots
Distribution
Part of a larger recall action
This product is one of 5 recalled by Beckman Coulter Inc. under a single FDA (Devices) recall action. The agency files each product separately, so the size of the action is not visible from this page alone.
See all 5 products in this recall →Related recalls
Same manufacturer · Beckman Coulter Inc.
See all →- HighBeckman Coulter DxC 500 AU Analyzer Software Defect Recall
FDA (Devices) · 2026-08-26
- HighBeckman Coulter DxC 500i Analyzer Modules Recalled for Software Defect
FDA (Devices) · 2026-08-26
- High
- SevereBeckman Coulter MicroScan Panel Update DVD Kits Recalled
FDA (Devices) · 2026-08-12
- HighBeckman Coulter Access Total T4 Calibrator Recalled for Inaccurate Results
FDA (Devices) · 2026-04-29
Same hazard · false low results
See all →- HighVITROS Chemistry CRP Diagnostic Slides May Underreport Elevated Results
FDA (Devices) · 2024-12-04
- ModerateBeckman Coulter Recalls Lactate Reagent Due to Interference from Specific Drugs
FDA (Devices) · 2017-05-24
- ModerateBeckman Coulter Cholesterol Reagent Recalled for Potential Drug Interference
FDA (Devices) · 2017-05-24
- ModerateBeckman Coulter Recalls HDL Cholesterol Reagent Due to Drug Interference
FDA (Devices) · 2017-05-24
- ModerateBeckman Coulter Recalls Uric Acid Reagent Due to Drug Interference
FDA (Devices) · 2017-05-24