The Recall Desk
ModerateFDA (Devices)·Z-1910-2013·Announced 2013-08-14

BioMerieux Vitek 2 Software Version 6.01 Recalled for Interface Errors

BioMerieux is recalling Vitek 2 software version 6.01 because specific antibiotic codes are not updated properly when interfaced with Laboratory Information Systems.

What this means for you

Minor labeling or packaging issue, or a voluntary precautionary recall. Risk of harm is low but real — follow the manufacturer's instructions.

Our severity reasoning: The recall is classified as FDA Class II, but the source text does not report any illnesses, injuries, or hospitalizations, fitting the criteria for a moderate severity score.

Plain-English summary

BioMerieux Inc is recalling 12,384 kits of Vitek 2 Systems Software version 6.01. The software is used with the VITEK 2 automated system for the identification and antimicrobial susceptibility testing of bacteria and yeast.

The recall was initiated because a product issue was identified for customers using software version 6.01 who are interfaced with a Laboratory Information System (LIS). During the update of software from version 5.xx to 6.01, the Bidirectional Computer Interface (BCI) antibiotic codes are not updated properly for Cefovecin, Prulifloxacin, and Ceftaroline.

The affected products were distributed worldwide, including the United States and numerous international countries. Users of the affected software should contact BioMerieux Inc for further instructions regarding the recall.

The recalled product

Product
bioMerieux Vitek 2 Systems Software version 6.01 software, used with the VITEK 2 automated system. Kit, UPDT V2S 6.01 DE. Part number: 414539. The product is used with the VITEK 2 automated system which consists of instruments, software and reagent cards designed for the iden
Manufacturer
Biomerieux Inc
Affected units
12,384

Is your product affected?

Check your packaging against the codes below. If any of them match, the product is part of this recall.

Lot numbers (1)

  • version 6.01

Distribution

Distributed nationwide across the United States.

Part of a larger recall action

This product is one of 15 recalled by Biomerieux Inc under a single FDA (Devices) recall action. The agency files each product separately, so the size of the action is not visible from this page alone.

See all 15 products in this recall →