Medtronic Endo GIA Stapler Reloads Recalled for Missing Sled Component
Medtronic is recalling Endo GIA 30mm stapler reloads because a missing sled component may cause staple failure, leading to bleeding or leakage.
The agency’s classification
FDA (Devices): Class II
Class II is a situation in which use of or exposure to a violative product may cause temporary or medically reversible adverse health consequences, or where the probability of serious adverse health consequences is remote.
This is the agency’s own classification, taken from its notice. The severity score below is our assessment, not the agency’s. How recall classes work
What this means for you
Our assessment, not the agency’s: severity 4/5 (Severe).
The hazard poses a serious risk of illness, injury, or structural failure. Stop using the product immediately and contact the manufacturer for a refund or repair.
Our severity reasoning: The recall involves a medical device defect that can lead to significant injury (bleeding or leakage of luminal contents) during surgery, meeting the criteria for a Severe rating.
Plain-English summary
Medtronic is recalling 171,271 units of the Endo GIA 30mm Gray Curved Tip Articulating Vascular Reload with Tri-Staple Technology (Item Code EGIA30AV). The recall covers units distributed worldwide and in the US nationwide.
The recall was initiated because the device may be missing a sled component, which is responsible for staple deployment. If the sled is missing, the device may fail to form a staple line when tissue is divided during surgery. This failure can lead to bleeding or leakage of luminal contents.
Healthcare providers and facilities should check their inventory for the specific lot number N7G0757KX and GTINs 10884521186095 (Inner Carton) and 20884521186092 (Outer Carton). Users should stop using the affected devices and contact Medtronic for instructions on return or replacement.
The recalled product
- Product
- Endo GIA" 30mm Gray Curved Tip Articulating Vascular Reload with Tri-Staple" Technology( Item Code EGIA30AV) Product Usage: The Endo GIA universal and Endo GIA universal XL staplers place two, triple-staggered rows of titanium staples and simultaneously divides the tissue…
- Manufacturer
- COVIDIEN MEDTRONIC
- Affected units
- 171,271
Distribution
Distributed nationwide across the United States.
Part of a larger recall action
This product is one of 40 recalled by COVIDIEN MEDTRONIC under a single FDA (Devices) recall action. The agency files each product separately, so the size of the action is not visible from this page alone.
See all 40 products in this recall →Related recalls
Same manufacturer · COVIDIEN MEDTRONIC
See all →- SevereMedtronic Endo GIA Vascular Reloads Recalled for Missing Sled Component
FDA (Devices) · 2018-07-25
- SevereMedtronic Endo GIA Reloads Recalled for Missing Sled Component
FDA (Devices) · 2018-07-25
- SevereMedtronic Tri-Staple 2.0 Reloads Recalled for Missing Sled Component
FDA (Devices) · 2018-07-25
- SevereMedtronic Endo GIA Stapler Recall Due to Missing Sled Component
FDA (Devices) · 2018-07-25
- SevereEndo GIA 60 mm Articulating Vascular/Medium Loading Unit Recalled for Missing Sled
FDA (Devices) · 2018-07-25
Same hazard · bleeding
See all →- HighHollister Recalls Onli Hydrophilic Urethral Catheters Due to Manufacturing Defect
FDA (Devices) · 2026-09-02
- SevereCannon II Plus Hemodialysis Catheter Recall Due to Sheath Introducer Defect
FDA (Devices) · 2026-05-20
- SevereArrow-Clark VectorFlow Hemodialysis Catheters with Defective Sheath Introducer
FDA (Devices) · 2026-05-20
- SevereNextStep Hemodialysis Catheters with Defective Sheath Introducer
FDA (Devices) · 2026-05-20
- SevereArrow Edge hemodialysis kits contain a sheath introducer that may not split
FDA (Devices) · 2026-05-20
Similar recalls
- SevereMedtronic Endo GIA Stapler Recall Due to Missing Sled Component
FDA (Devices) · 2018-07-25
- SevereMedtronic Endo GIA Vascular Reloads Recalled for Missing Sled Component
FDA (Devices) · 2018-07-25
- SevereMedtronic Endo GIA Reloads Recalled for Missing Sled Component
FDA (Devices) · 2018-07-25
- SevereEndo GIA 45 mm Articulating Medium/Thick Reload Recalled for Missing Sled Component
FDA (Devices) · 2018-07-25
- SevereEndo GIA 60 mm Articulating Vascular/Medium Loading Unit Recalled for Missing Sled
FDA (Devices) · 2018-07-25
- SevereMedtronic Tri-Staple 2.0 Curved Tip Reloads Recalled for Missing Sled Component
FDA (Devices) · 2018-07-25