Medtronic Recalls Endo GIA Stapler Loading Units Due to Missing Sled Component
Medtronic is recalling 171,271 Endo GIA Auto Suture Universal Articulating Loading Units because a missing sled component may prevent staple formation, risking bleeding or leakage.
The agency’s classification
FDA (Devices): Class II
Class II is a situation in which use of or exposure to a violative product may cause temporary or medically reversible adverse health consequences, or where the probability of serious adverse health consequences is remote.
This is the agency’s own classification, taken from its notice. The severity score below is our assessment, not the agency’s. How recall classes work
What this means for you
Our assessment, not the agency’s: severity 4/5 (Severe).
The hazard poses a serious risk of illness, injury, or structural failure. Stop using the product immediately and contact the manufacturer for a refund or repair.
Our severity reasoning: The recall is classified as FDA Class II. The hazard involves a structural defect in a surgical device that can lead to significant injury (bleeding or leakage of luminal contents) during a procedure, meeting the criteria for a Severe rating.
Plain-English summary
Medtronic is recalling 171,271 units of the Endo GIA Auto Suture Universal Articulating Loading Unit (30mm-2.0mm, Item Code 030450). The affected lots are N7L0306KX, N7M0221KX, N8B0724KX, N8C0958KX, and N8C1303KRX. The product was distributed worldwide and in the US nationwide.
The recall was initiated because the device may be missing a sled component, which is responsible for staple deployment. If the sled is missing, the device may fail to form a staple line when tissue is divided. This failure can lead to bleeding or leakage of luminal contents during surgical procedures.
Healthcare providers and facilities should check their inventory for the specific item code and lot numbers listed above. If the affected units are found, they should be removed from service and returned to Medtronic according to the recall instructions.
The recalled product
- Product
- Endo GIA" Auto Suture" Universal Articulating Loading Unit 30mm- 2.0mm ( Item Code 030450) Product Usage: The Endo GIA universal and Endo GIA universal XL staplers place two, triple-staggered rows of titanium staples and simultaneously divides the tissue between the two…
- Manufacturer
- COVIDIEN MEDTRONIC
- Affected units
- 171,271
Is your product affected?
Check your packaging against the codes below. If any of them match, the product is part of this recall.
Lot numbers (4)
- Lots: N7L0306KX
- N7M0221KX
- N8B0724KX
- N8C0958KX
Distribution
Distributed nationwide across the United States.
Part of a larger recall action
This product is one of 40 recalled by COVIDIEN MEDTRONIC under a single FDA (Devices) recall action. The agency files each product separately, so the size of the action is not visible from this page alone.
See all 40 products in this recall →Related recalls
Same manufacturer · COVIDIEN MEDTRONIC
See all →- SevereMedtronic Endo GIA Vascular Reloads Recalled for Missing Sled Component
FDA (Devices) · 2018-07-25
- SevereMedtronic Endo GIA Reloads Recalled for Missing Sled Component
FDA (Devices) · 2018-07-25
- SevereMedtronic Tri-Staple 2.0 Reloads Recalled for Missing Sled Component
FDA (Devices) · 2018-07-25
- SevereMedtronic Endo GIA Stapler Recall Due to Missing Sled Component
FDA (Devices) · 2018-07-25
- SevereEndo GIA 60 mm Articulating Vascular/Medium Loading Unit Recalled for Missing Sled
FDA (Devices) · 2018-07-25
Same hazard · bleeding
See all →- HighHollister Recalls Onli Hydrophilic Urethral Catheters Due to Manufacturing Defect
FDA (Devices) · 2026-09-02
- SevereCannon II Plus Hemodialysis Catheter Recall Due to Sheath Introducer Defect
FDA (Devices) · 2026-05-20
- SevereArrow-Clark VectorFlow Hemodialysis Catheters with Defective Sheath Introducer
FDA (Devices) · 2026-05-20
- SevereNextStep Hemodialysis Catheters with Defective Sheath Introducer
FDA (Devices) · 2026-05-20
- SevereArrow Edge hemodialysis kits contain a sheath introducer that may not split
FDA (Devices) · 2026-05-20
Similar recalls
- SevereMedtronic Endo GIA Stapler Recall Due to Missing Sled Component
FDA (Devices) · 2018-07-25
- SevereMedtronic Endo GIA Reloads Recalled for Missing Sled Component
FDA (Devices) · 2018-07-25
- SevereMedtronic Endo GIA Vascular Reloads Recalled for Missing Sled Component
FDA (Devices) · 2018-07-25
- HighEndo GIA Stapler Recall Due to Incomplete Staple Formation
FDA (Devices) · 2020-10-07
- HighEndo GIA Stapler Loading Units Recalled for Staple Formation Defects
FDA (Devices) · 2020-10-07
- HighEndo GIA Auto Suture Universal Articulating Loading Unit Recall
FDA (Devices) · 2020-10-07