Boston Scientific Recalls iLab Polaris Systems Due to Data Archiving Bug
Boston Scientific is recalling 150 iLab Polaris and Ultrasound systems in Europe because a software bug prevents specific data from being archived without notifying the user.
What this means for you
Minor labeling or packaging issue, or a voluntary precautionary recall. Risk of harm is low but real — follow the manufacturer's instructions.
Our severity reasoning: The recall is classified as FDA Class II, but the hazard is a software data integrity issue (loss of diagnostic data) rather than a direct physical injury or life-threatening malfunction. No injuries or illnesses are reported, fitting the criteria for a moderate severity precautionary recall.
Plain-English summary
Boston Scientific Corporation is recalling 150 units of the iLab Polaris Multi-Modality Guidance System and iLab Ultrasound Imaging System. The affected units were upgraded with a specific version of 2.10 software (disk batch 629774) that contains a bug affecting data storage.
The defect causes the system to fail to archive Diastolic Hyperemia-Free Ratio (DFR) data if the user selects Multi-Frame UltraSound or SC Image Storage formats. When this occurs, the system does not notify the user of the failure, meaning the data is lost without warning. The issue only allows data to be archived correctly if the True Color Secondary Capture format is used.
The recall affects units distributed to Japan, Belgium, Denmark, France, Great Britain, and Spain. There are no U.S. consignees for this recall. Users should check their software version and serial numbers against the provided list to determine if their system is affected.
The recalled product
- Product
- iLab Polaris Multi-Modality Guidance System; iLab Ultrasound Imaging System
- Manufacturer
- Boston Scientific Corporation
- Hazard
- Affected units
- 150
Is your product affected?
Check your packaging against the codes below. If any of them match, the product is part of this recall.
Lot numbers (20)
- Polaris 2.10 software version (disk batch 629774)
- iLab Polaris Systems affected product listing (serial numbers): 200204
- 200209
- 200217
- 200308
- 200310
- 200448
- 200450
- 200456
- 200457
- 200458
- 200461
- 200480
- 200481
- 200103
- 200145
- 200044
- 200118
- 200360
- 200498
Distribution
Distribution scope not specified by the agency.
Related recalls
Same manufacturer · Boston Scientific Corporation
See all →- HighBoston Scientific Recalls Suture Cinch Endoscopic Tissue Approximation Devices
FDA (Devices) · 2026-08-26
- HighBoston Scientific ENROUTE Transcarotid Neuroprotection System Plus Recall
FDA (Devices) · 2026-08-26
- HighBoston Scientific Recalls OverStitch Suture Cinch Devices for Deployment Failure
FDA (Devices) · 2026-08-26
- HighBoston Scientific ENROUTE Transcarotid Neuroprotection System Recall for Tip Separation
FDA (Devices) · 2026-08-26
- SevereRadial Jaw 4 Jumbo Biopsy Forceps Packaging Sterile Barrier Recall
FDA (Devices) · 2026-08-12
Same hazard · data loss
See all →- SeverePhilips Allura X-Ray Systems Recalled Over Unauthorized Hard Disk Drives
FDA (Devices) · 2026-08-12
- HighPhilips Fluoroscopy System Recall Over Hard Disk Drive Issue
FDA (Devices) · 2026-08-12
- HighRayCare 2024A SP4 Oncology System Beam Set Delivery Note Data Loss
FDA (Devices) · 2026-07-08
- HighRayCare 2024A SP1 Oncology Information System Software Recall
FDA (Devices) · 2026-07-08
- HighMint Lesion Software Data Loss During Server Connection Interruption
FDA (Devices) · 2026-05-13