CSL Behring Recalls Siemens N Latex CDT Kit for Negative Bias
CSL Behring GmbH is recalling Siemens N Latex CDT Kits due to negative bias in carbohydrate-deficient transferrin measurements observed after 15 months.
The agency’s classification
FDA (Devices): Class II
Class II is a situation in which use of or exposure to a violative product may cause temporary or medically reversible adverse health consequences, or where the probability of serious adverse health consequences is remote.
This is the agency’s own classification, taken from its notice. The severity score below is our assessment, not the agency’s. How recall classes work
What this means for you
Our assessment, not the agency’s: severity 2/5 (Moderate).
Minor labeling or packaging issue, or a voluntary precautionary recall. Risk of harm is low but real — follow the manufacturer's instructions.
Our severity reasoning: The recall is classified as FDA Class II. The hazard involves potential diagnostic inaccuracy (negative bias) rather than direct physical injury or a high-risk pathogen, fitting the criteria for a moderate severity recall involving a medical device with potential for adverse health consequences through misdiagnosis.
Plain-English summary
CSL Behring GmbH is recalling the Siemens N Latex CDT Kit, an in vitro diagnostic (IVD) product used for the quantitative determination of carbohydrate-deficient transferrin (CDT) in human serum. The recall affects 297 packages in the United States and 1,247 packages outside the United States. The affected US lot number is 49822, and the expiration date is April 11, 2020.
The recall was initiated because shelf life monitoring observed negative bias in the CDT measurement results when compared to the HPLC method. This bias was observed to occur after 15 months of storage. The product was distributed in the US states of Indiana, Kansas, Missouri, Ohio, and Virginia, as well as in numerous foreign countries including Australia, Canada, China, Germany, and Japan.
Healthcare facilities and laboratories should stop using the affected lots of the Siemens N Latex CDT Kit. Users should contact CSL Behring GmbH for further instructions on returning or disposing of the product. No specific instructions for individual consumers are provided as this is a professional in vitro diagnostic product.
The recalled product
- Product
- Siemens N Latex CDT Kit- IVD quantitative determination of carbohydrate-deficient transferrin (CDT) in human serum. Siemens Material Number (SMN): 10445997
- Manufacturer
- CSL Behring GmbH
- Hazard
Is your product affected?
Check your packaging against the codes below. If any of them match, the product is part of this recall.
Lot numbers (4)
- US Lot number: 49822 UDI: 842768018534 OUS: Lot numbers: 49729
- 49849
- 49985
- 50084 Exp. Date: 2020-04-11
Distribution
Related recalls
Same manufacturer · CSL Behring GmbH
See all →- HighCSL Behring Recalls N Latex CDT Kits Due to Measurement Bias
FDA (Devices) · 2018-09-05
- ModerateSiemens N/T Protein Control SL Recalled for Inaccurate Ceruloplasmin Values
FDA (Devices) · 2017-12-06
- ModerateSiemens N Protein Standard SL Recalled for Inaccurate Ceruloplasmin Calibration
FDA (Devices) · 2017-12-06
- ModerateCSL Behring Recalls Dimension Vista Homocysteine Reagent Cartridges
FDA (Devices) · 2016-07-27
Same hazard · diagnostic inaccuracy
See all →- High
- HighAtellica CH Enzymatic Hemoglobin A1c assay well-to-well bias
FDA (Devices) · 2026-07-15
- HighTBS iNsight bone density software calculation variability on Hologic machines
FDA (Devices) · 2025-03-12
- HighBD PCR Cartridges for MAX System signal drift in select lots
FDA (Devices) · 2025-01-29
- HighHPV-16 Diagnostic Probe Recalled for Suboptimal DNA Hybridization
FDA (Devices) · 2024-12-04
Similar recalls
- HighCSL Behring Recalls N Latex CDT Kits Due to Measurement Bias
FDA (Devices) · 2018-09-05
- ModerateSiemens N Protein Standard SL Recalled for Inaccurate Ceruloplasmin Calibration
FDA (Devices) · 2017-12-06
- ModerateSiemens ADVIA Centaur Vitamin D Assay Recalled for Negative Bias
FDA (Devices) · 2013-04-24
- ModerateSiemens Ferritin Reagent Cartridge Recalled for Negative Bias
FDA (Devices) · 2019-08-14
- LowSiemens N Latex IgM Reagent Recalled for Proficiency Testing Failures
FDA (Devices) · 2015-03-25
- HighSiemens ADVIA Centaur Progesterone Kit Recalled for High Bias in Results
FDA (Devices) · 2015-03-11