Anspach Effort eMax 2 Plus Console Recalled for Low Motor RPM
The Anspach Effort, Inc. is recalling eMax 2 Plus dental consoles because the motor handpiece may operate at lower than expected speeds in free-running conditions.
What this means for you
Minor labeling or packaging issue, or a voluntary precautionary recall. Risk of harm is low but real — follow the manufacturer's instructions.
Our severity reasoning: The recall is classified as FDA Class II, and the defect involves a performance deviation (lower RPM in free-running state) that does not affect the loaded working range or result in reported injuries, fitting the criteria for a moderate severity precautionary recall.
Plain-English summary
The Anspach Effort, Inc. is recalling 61 units of the eMax 2 Plus dental console (Reference SC2100) due to a performance defect. The recall involves specific system consoles, including the SC2100, SC2101, and SC2102 models.
The defect involves the Motor Handpiece potentially operating at less than 80,000 RPM (approximately 72,000 RPM) in the free-running state when using the optional Hand Control. This condition occurs only when the Hand Control is fully depressed and there is no load on the cutting bur. The issue does not affect the power output of the handpiece in the loaded working range, which typically operates between 50,000 and 70,000 RPM.
The affected units were distributed worldwide, including nationwide in the United States and in Australia, Egypt, Italy, Japan, Switzerland, Taiwan, and Turkey. Users should contact the recalling firm for further instructions regarding the recall.
The recalled product
- Product
- ***REF SC2100***eMax 2 Plus Console***SN G 10307653002*** Product Usage: The RPM of the Motor Handpiece may be less than 80,000 RPM (approximately 72,000 RPM) in the free running state when using the optional Hand Control. The low Motor Handpiece RPM would only be experienced
- Manufacturer
- The Anspach Effort, Inc.
- Hazard
- Affected units
- 61
Is your product affected?
Check your packaging against the codes below. If any of them match, the product is part of this recall.
Lot numbers (6)
- SC2100 System Console Emax 2 Plus
- SC2101 System Console
- Single Port with irrigation
- SC2102 System Console
- Single Port
- without Irrigation.
Distribution
Part of a larger recall action
This product is one of 3 recalled by The Anspach Effort, Inc. under a single FDA (Devices) recall action. The agency files each product separately, so the size of the action is not visible from this page alone.
See all 3 products in this recall →Related recalls
Same manufacturer · The Anspach Effort, Inc.
See all →- HighAdult Craniotome Bearings May Detach During Surgical Use
FDA (Devices) · 2023-01-18
- HighCraniotome Attachment Ball Bearings May Separate During Surgery
FDA (Devices) · 2023-01-18
- HighLarge Craniotome Attachment Ball Bearings Risk Loosening During Surgery
FDA (Devices) · 2023-01-18
- HighAnspach Adult Rotating Craniotome Attachment ball bearings may detach during surgery
FDA (Devices) · 2023-01-18
- HighCraniotome Attachment Ball Bearings May Detach During Surgery
FDA (Devices) · 2023-01-18
Same hazard · performance defect
See all →- ModerateBD GasPak EZ CO2 Pouch System gas generating sachets underperforming
FDA (Devices) · 2026-05-06
- ModerateBD GasPak EZ Campy Pouch System Recall for Below-Specification Gas Production
FDA (Devices) · 2026-05-06
- ModeratePhilips C5-2 Ultrasound Transducers Refurbished Beyond Useful Life
FDA (Devices) · 2025-07-09
- ModerateThermo SCIENTIFIC Sensititre Mueller-Hinton Broth diagnostic device recalled
FDA (Devices) · 2025-06-18
- ModerateID NOW COVID-19 2.0 Point-of-Care Test Device Recall
FDA (Devices) · 2025-04-23