Maquet Getinge Custom Tubing Sets Recalled for Sterility Assurance Concerns
Maquet Cardiovascular is recalling specific custom tubing sets for extracorporeal circulation because a design flaw may prevent proper sterilization of stopcock ports.
The agency’s classification
FDA (Devices): Class II
Class II is a situation in which use of or exposure to a violative product may cause temporary or medically reversible adverse health consequences, or where the probability of serious adverse health consequences is remote.
This is the agency’s own classification, taken from its notice. The severity score below is our assessment, not the agency’s. How recall classes work
What this means for you
Our assessment, not the agency’s: severity 3/5 (High).
Real risk of harm even if no illness or injury has been reported yet. Stop using the product and follow the agency's guidance.
Our severity reasoning: The recall involves a risk-of-harm product (sterility failure in a medical device) where injury has not yet been reported, fitting the 'High' severity criteria.
Plain-English summary
Maquet Cardiovascular, LLC is recalling specific lots of Maquet Getinge Custom Tubing Sets for Extracorporeal Circulation (ECC). The affected products are the BEQ-TOP 15801 1/2 inch ECC Pack (Material 701055490R01, Batch 3000050886) and the BEQ-TOP 15801 1/4 in ECC Pack (Material 701055490R02, Batches 3000066508 and 3000081906). These products were distributed nationwide.
The recall was initiated because the closed stopcock and non-vented cap configuration may prevent the flow of humidity and ethylene oxide gas into the fluid path during sterilization. As a result, the Sterility Assurance Level (SAL) of the stopcock ports cannot be assured, potentially compromising the sterility of the device.
Healthcare facilities and users should stop using the affected tubing sets immediately and contact Maquet Cardiovascular for instructions on return or replacement. The source text does not report any specific illnesses or injuries associated with this recall.
The recalled product
- Product
- Maquet Getinge- (1) Material: 701055490R01 BEQ-TOP 15801 ¿ inch ECC Pack (2) Material :701055490R02 BEQ-TOP 15801 1/4 in ECC Pack
- Manufacturer
- Maquet Cardiovascular, LLC
Is your product affected?
Check your packaging against the codes below. If any of them match, the product is part of this recall.
Lot numbers (2)
- Batch Numbers: (1) 3000050886 (2) 3000066508
- 3000081906
Distribution
Distributed nationwide across the United States.
Part of a larger recall action
This product is one of 183 recalled by Maquet Cardiovascular, LLC under a single FDA (Devices) recall action. The agency files each product separately, so the size of the action is not visible from this page alone.
See all 183 products in this recall →Related recalls
Same manufacturer · Maquet Cardiovascular, LLC
See all →- HighVasoview Hemopro 2 Endoscopic Vessel Harvesting System Recall
FDA (Devices) · 2026-07-22
- HighVessel harvesting systems recalled after out-of-box failures in harvesting tool jaws
FDA (Devices) · 2025-10-01
- HighHeartstring III seal systems recalled over failures to load, deploy and seal
FDA (Devices) · 2025-09-24
- HighHeartstring III 4.3 mm seal systems recalled over loading and deployment failures
FDA (Devices) · 2025-09-24
- HighHeartstring III 3.8 mm seal systems recalled over three identified failure modes
FDA (Devices) · 2025-09-24
Same hazard · sterility failure
See all →- SevereAvKare Carboxymethylcellulose Sodium Eye Drops Recalled for Sterility Concerns
FDA (Drugs) · 2026-09-30
- SevereAvKare Carboxymethylcellulose Sodium Eye Drops Recalled for Sterility Concerns
FDA (Drugs) · 2026-09-30
- SevereROHTO DIGI-EYE Eye Drops Recalled Over Sterility Assurance Gap
FDA (Drugs) · 2026-07-29
- HighROHTO DRY-AID Eye Drops Recalled Due to Sterility Concerns
FDA (Drugs) · 2026-07-29
- SevereOne-Piece Introducer Tray kits recalled for sterility failure
FDA (Devices) · 2026-07-22
Similar recalls
- HighMaquet Custom Tubing Packs Recalled for Potential Endotoxin Exposure
FDA (Devices) · 2016-04-20
- HighMaquet CARDIOHELP System extracorporeal support device components recalled for sterility
FDA (Devices) · 2023-03-08
- HighMedtronic Perfusion Tubing Packs Recalled for Insufficient or Incomplete Welds
FDA (Devices) · 2021-06-23
- HighMaquet Rotaflow Centrifugal Pump Tubing Recalled for Endotoxin Risk
FDA (Devices) · 2016-04-20
- SevereDatascope Recalls BEQ-TOP 24500 Adult ECC Custom Tubing Kits
FDA (Devices) · 2018-05-16
- LowDatascope Recalls Maquet Cardiopulmonary Bypass Custom Tubing Kits
FDA (Devices) · 2018-06-13