Maquet Custom Tubing Sets Recalled for Sterility Assurance Concerns
Maquet Cardiovascular is recalling custom tubing sets for extracorporeal circulation because a design flaw may prevent proper sterilization, potentially compromising sterility.
The agency’s classification
FDA (Devices): Class II
Class II is a situation in which use of or exposure to a violative product may cause temporary or medically reversible adverse health consequences, or where the probability of serious adverse health consequences is remote.
This is the agency’s own classification, taken from its notice. The severity score below is our assessment, not the agency’s. How recall classes work
What this means for you
Our assessment, not the agency’s: severity 3/5 (High).
Real risk of harm even if no illness or injury has been reported yet. Stop using the product and follow the agency's guidance.
Our severity reasoning: The recall involves a medical device where sterility cannot be assured, creating a risk of infection or harm, but the source text does not report any specific illnesses or injuries, fitting the 'High' severity criteria for risk-of-harm products without reported injury.
Plain-English summary
Maquet Cardiovascular, LLC is recalling Custom Tubing Sets for Extracorporeal Circulation (ECC) under FDA recall number Z-1645-2019. The affected product is identified by the description "Maquet Getinge-BEQ-TOP 250011/4 1/4 Less than 20 kg up" and material code 701053906R01. The recall covers specific batch numbers: 3000068802, 3000071398, 3000074190, and 3000076543.
The recall was initiated because the closed stopcock and non-vented cap configuration may prevent the flow of humidity and ethylene oxide gas into the fluid path during sterilization. This design issue means the Sterility Assurance Level (SAL) of the stopcock ports cannot be assured, potentially leaving the device non-sterile.
The products were distributed nationwide. Healthcare facilities and users should stop using the affected tubing sets and contact Maquet Cardiovascular for further instructions or replacement.
The recalled product
- Product
- Maquet Getinge-BEQ-TOP 250011/4 1/4 Less than 20 kg up Material: 701053906R01
- Manufacturer
- Maquet Cardiovascular, LLC
Is your product affected?
Check your packaging against the codes below. If any of them match, the product is part of this recall.
Lot numbers (1)
- Batch Numbers: 3000068802 3000071398 3000074190 3000076543
Distribution
Distributed nationwide across the United States.
Part of a larger recall action
This product is one of 183 recalled by Maquet Cardiovascular, LLC under a single FDA (Devices) recall action. The agency files each product separately, so the size of the action is not visible from this page alone.
See all 183 products in this recall →Related recalls
Same manufacturer · Maquet Cardiovascular, LLC
See all →- HighVasoview Hemopro 2 Endoscopic Vessel Harvesting System Recall
FDA (Devices) · 2026-07-22
- HighVessel harvesting systems recalled after out-of-box failures in harvesting tool jaws
FDA (Devices) · 2025-10-01
- HighHeartstring III seal systems recalled over failures to load, deploy and seal
FDA (Devices) · 2025-09-24
- HighHeartstring III 4.3 mm seal systems recalled over loading and deployment failures
FDA (Devices) · 2025-09-24
- HighHeartstring III 3.8 mm seal systems recalled over three identified failure modes
FDA (Devices) · 2025-09-24
Same hazard · sterility failure
See all →- SevereAvKare Carboxymethylcellulose Sodium Eye Drops Recalled for Sterility Concerns
FDA (Drugs) · 2026-09-30
- SevereAvKare Carboxymethylcellulose Sodium Eye Drops Recalled for Sterility Concerns
FDA (Drugs) · 2026-09-30
- SevereROHTO DIGI-EYE Eye Drops Recalled Over Sterility Assurance Gap
FDA (Drugs) · 2026-07-29
- HighROHTO DRY-AID Eye Drops Recalled Due to Sterility Concerns
FDA (Drugs) · 2026-07-29
- SevereOne-Piece Introducer Tray kits recalled for sterility failure
FDA (Devices) · 2026-07-22
Similar recalls
- HighMaquet CARDIOHELP System extracorporeal support device components recalled for sterility
FDA (Devices) · 2023-03-08
- HighMaquet Custom Tubing Packs Recalled for Potential Endotoxin Exposure
FDA (Devices) · 2016-04-20
- HighDatascope Recalls Pediatric Cardiopulmonary Bypass Tubing Kits Due to Valve Defect
FDA (Devices) · 2018-06-20
- HighMaquet Pediatric Arterial Cannula Recalled for Incorrectly Sized Connectors
FDA (Devices) · 2013-07-03
- HighMaquet Pediatric Arterial Cannula Recall for Limited Flexibility
FDA (Devices) · 2012-08-22
- HighMaquet Rotaflow Centrifugal Pump Tubing Recalled for Endotoxin Risk
FDA (Devices) · 2016-04-20